Antec Leyden DECADE II Manuale utente

Edition 5, 2005
171.7034
DECADE II
Electrochemical Detector
Installation of Dual Cell Control
upgrade kit


INI DECADE II DCC upgrade kit install guide, edition 5 1
INI
Symbols
The following symbol are used on the rear panel and oven compartment of
the DECADE II:
Consult the manual for further safety instructions
Frame or chassis ground terminal
The following pictograms are used in the DECADE II manual:
Caution
Caution, risk of electric shock or other electrical hazard (high
voltage)

2
Safety practices
The following safety practices are intended to insure safe operation of the
equipment.
Electrical hazards
The removal of protective panels on the instrument can result in exposure to
potentially dangerous voltages. Therefore, disconnect the instrument from all
power sources before disassembly. Untrained personnel should not open the
instrument.
Replace blown fuses with fuses of proper type and rating as stipulated on the
rear panel and specified in the installation section of this manual. The fuse
holder is integrated in the mains connector. Ensure that the instrument is
never put in operation with fuses of a different type. This could cause fire.
USE ONLY WITH 250 V FUSES
EMPLOYER UNIQUEMENT
A
VEC DES FUSIBLES DE 250V
Connect the detector to a grounded AC power source, line voltage 100 – 240
VAC. The instrument should be connected to a protective earth via a ground
socket. The power source should exhibit minimal power transients and
fluctuations. Replace faulty or frayed power cords.

INI DECADE II DCC upgrade kit install guide, edition 5 3
Place the detector on a flat and smooth surface. Do not block the fan located
at the bottom of the detector. Blocking the fan will impair the cooling capability
of the power supply.
Before starting the replacement of the EPROMS please read the following
safety instructions carefully:
Take precautions against electrostatic discharge during
installation/removal of boards, EPROM’s or other electrical
components at all time to prevent damage of the circuit boards.

4
Spare parts and service availability
Manufacturer provides operational spare parts of the instrument and current
accessories for a period of five years after shipment of the final production run
of the instrument. Spare parts will be available after this five years period on
an ‘as available’ basis.
Manufacturer provides a variety of services to support her customers after
warranty expiration. Repair service can be provided on a time and material
basis. Contact your local supplier for servicing. Technical support and training
can be provided by qualified chemists on both contractual or as-needed
basis.
Copyright ©2003. All rights reserved. Contents of this publication may not be reproduced in any
form or by any means (including electronic storage and retrieval or translation into a foreign
language) without prior agreement and written consent from the copyright of the owner.
The information contained in this document is subject to change without notice. The software and
the information provided herein is believed to be reliable. However,
Antec Leyden shall not be
liable for errors contained herein or for incidental or consequential damages in connection with the
furnishing, performance, or use of this software or this manual. All use of the software shall be
entirely at the user’s own risk.

INI DECADE II DCC upgrade kit install guide, edition 5 5
Table of contents
Symbols 1
Safety practices 2
Spare parts and service availability 4
Dual Cell Control upgrade kit 7
Important notice 7
Installation guide 8
Unpacking 8
Contents of kits 8
Tools 10
Installation procedure 11
1. Upload of firmware version 3.xx 12
2. Removing the rear panel 13
3. Installation of 2nd internal cell cable 15
4. Identification of cables (labels) 15
5. Removing sensor board 17
6. Replacement of boot EPROMS 19
7. Reconfiguration of connector 7 (heater) 20
8. Installation of the DECADE II TC-board 22
9. Installation of power cable for 2nd sensor board 25
10. Configuration of dip switches 26
11. Placement of I/O label 27
12. Installation of sensor boards 28
13. Installation of cell and column clamps 28
14. Placement of cell identification labels 29
15. Upload of firmware version 3.xx 30
16. Check & calibration of DECADE II DCC 30
Appendix A - Installation of 2nd internal cell cable 35

6

CHAPTER 1 DECADE II DCC upgrade kit install guide, edition 5 7
CHAPTER 1
Dual Cell Control upgrade kit
Congratulations on your purchase of the DECADE II Dual Cell Control
upgrade kit, p/n 171.0216, 171.0218 or 171.0220. This DCC kit enables you
to upgrade your DECADE II to an electrochemical workstation with two flow
cells allowing serial or independent parallel operation.
Important notice
Your DECADE II is factory installed with a LM35 temperature sensor mounted
in the heater compartment. The LM35 sensor has a cable with three wires
(red/white/ black) and a 6-pin Molex KK connector (white) . This cable is
originally not connected to any board and is fixed into the grey cable clamp on
the left side panel of the DECADE II inside the housing.
Only a few DECADE II detectors on the market do not have the LM35 sensor
preinstalled (first series s/n, < 171.00025).
If your DECADE II has a s/n < 17100025 DO NOT procede with the
installation of the DCC kit, but please contact the manufacturer. In that
case the DECADE II should be upgraded at the factory.
There are three versions of the DECADE II DCC kit available. The particular
kit required for the upgrade of your detector is depending on the serial
number of the DECADE II.
P/n Description
171.0216 DECADE II DCC upgrade kit, s/n 92 -
171.0218 DECADE II DCC upgrade kit, s/n 72 - 91
171.0220 DECADE II DCC upgrade kit, s/n 25 - 71
Depending on the upgrade kit purchased some of the installation steps
can be skipped. Please follow further instructions in the description of
the installation procedure on page 11 of this manual.

8 Installation guide
CHAPTER 2
Installation guide
Unpacking
Inspect the transport box for possible damage as it arrives. Immediately
inform the transport company in case of damage, otherwise she may not
accept any responsibility. Keep the transport box as it is designed for
optimum protection during transport and it may be needed again. Carefully
unpack the kit and inspect it for completeness and for possible damage.
Contact your supplier in case of damage or if not all marked items on the
checklist are included.
Prior to shipment, the parts of your upgrade kit have been inspected and
tested to ensure the best possible performance.
Contents of kits
Below the contents of the available upgrade kits are listed:
1. p/n 171.0216, DECADE II DCC upgrade kit s/n 92 -,
consists of the following parts:
P/n Description Qty
171.0624 DECADE II sensor board 1
171.0208 DECADE II DCC mounting kit 1
171.0224 DECADE II DCC accessory kit 1
Software:
171.9007 DECADE II Dialogue, demo version 1
Documentation:
171.0020 DECADE II service manual 1
171.7032 DECADE II FW upgrade user guide 1
171.7034 DECADE II DCC upgrade kit install guide 1
Altri manuali per DECADE II
1
Indice
Altri manuali Antec Leyden Sensore di sicurezza