C-1
PRODUCT MANUAL
SCC-GHR-MASTER 011619
Contact us for expert product application assistance.
1.225.644.5255
sales@genielters.com
genielters.com
41041 Black Bayou Road
Gonzales, Louisiana 70737 United States
Leading the industry in Analytically Correct™ Sampling Systems, Components, and Technology.
By applying sound principles of science and a minimalistic design approach, we manufacture
our products to solve your specific sampling needs. We specialize in Analytically Correct
Sample System Design, Sample Conditioning Components, and revolutionary gas and liquid
sampling technology. We are The Sampling Experts™. ISO 9001:2015 certied
Genie®, GHR™, are trademarks or registered trademarks of A+Corporation, LLC. All other referenced trademarks are the property of their respective
owners. © 2012 A+ Corporation. All rights reserved.
INSTALLATION AND OPERATION
Safety Warnings
The images below will be used to alert you to various types of safety messages. Obey all safety messages that follow this symbol to avoid possible injury/death or
equipment damage.
WARNING messages will appear before procedures that could result in serious personal injury or death.
CAUTION messages will appear before procedures that could result in minor or moderate injury.
NOTICE messages will appear before procedures that could result in damage to equipment.
WARNING: Read this manual completely and carefully before installing or using this product. Failure to do so can result in serious injury or death.
Save these instructions for future reference and for the local inspector’s use. Observe all governing codes and ordinances.
General
• Improper selection/installation/maintenance, misuse, or abuse of this equipment can cause death, serious injury, and property damage.
• Do not exceed ANY equipment pressure ratings. Refer to the technical specifications table for maximum pressure rating of the system. Verify the designed
pressure ratings of all equipment in your system (supply lines, fittings, connections, filters, valves, gauges, etc.). Prior to use in a system, a properly-sized relief
device is to be installed, which limits the use of 110% of the MAWP.
• Some fluids, when burning, do not exhibit a visible flame. Use extreme caution when inspecting and/or servicing systems using flammable fluids to avoid death
or serious injury to personnel. Provide a device to warn personnel of these dangerous conditions.
• Not designed for external fire or for use with mixtures of an explosive gas and air/oxygen.
• Do not use the regulator as a shut-off device. A ball valve on the outlet is recommended for this purpose. Our Genie Regulator Accessory Manifold™ is
designed to thread directly into any Genie Probe Regulator™ or Genie Regulator and provides a means to monitor regulator outlet pressure and block flow.
901-GR Electrical
WARNING: Electrical power must be “OFF” before and during installation and maintenance or personal injury may result.
• All appropriate national, state, and electrical codes must be followed.
• Do not remove the temperature controller from the instrument enclosure; the wires may disconnect from the controller terminals.
• To adjust the set temperature of the heated regulator, remove the instrument enclosure cover and use the arrow keys to adjust the set temperature. This
must be done in a non-hazardous location prior to installation. Note: the range of adjustment is 95°F to 300°F.
• When installed in a hazardous location, the instrument enclosure cover should NOT be removed unless power to the regulator is de-energized prior to
maintenance.
• Do not exceed 5 minutes with power without fluid flow or the thermal cut-off will open the circuit.
• A sealing fitting must be installed with access, allowing the dams to be made and the sealing compound to be properly poured.
• All unused conduit openings must be plugged. Plugs must be a minimum of 1/8” thick and engage in a minimum of 5 full threads.
GHR Electrical
WARNING: Electrical power must be “OFF” before and during installation and maintenance or personal injury may result.
• All appropriate national, state, and electrical codes must be followed. In Europe, EN 60079-14 latest applicable edition should be followed.
• The power supply is with the help of a silicone cable, notch and oil resistant. The cable is intended for installation inside a rigid metal conduit, cable tray
or a similar method which provides sufficient mechanical strength to the cable against stress and bending. Ground to be provided at the final location of
installation.
BEFORE YOU START
901-GRGHR