Applied DRDL Manuale utente

Operating Instructions
Dual Range Torque Transducer
DRDL
DRDL-n
DRDL-w
English

1
Content
Important Advice..........................................................................................................................2
1. Introduction .............................................................................................................................3
2. Application ...............................................................................................................................3
3. Technical Details.......................................................................................................................4
3.1 Torque Shaft......................................................................................................................4
3.2 Case ...................................................................................................................................4
3.3. Measurement...................................................................................................................4
3.4 Disturbances and their Compensation ..............................................................................5
4. Operating conditions................................................................................................................5
4.1. Environment Temperature...............................................................................................5
4.2 Humidity and Dust.............................................................................................................5
4.3 Chemically Resistance .......................................................................................................5
4.4 Deposition .........................................................................................................................5
5. Mounting..................................................................................................................................6
5.1 Precautions at Assembly ...................................................................................................6
5.2 General Mounting Instructions .........................................................................................6
6. Measuring Chain ......................................................................................................................7
7. Connection ...............................................................................................................................7
7.1 Hints for Connection..........................................................................................................8
7.2 Connectors ........................................................................................................................8
7.3 Pinout of the Connector....................................................................................................8
7.4 Prolongation of Cable........................................................................................................9
7.5 Supply Voltage...................................................................................................................9
8.
Pin Configuration
....................................................................................................................10
8.1 Standard-Measuring Cable AK12.4..................................................................................10
8.2 Robot cable AK12.5
..........................................................................................................11
8.3 Thin cable AK12.2
............................................................................................................12
9. Output....................................................................................................................................13
10. Recalibration ........................................................................................................................13
11. Disposal ................................................................................................................................13
12. Angle and speed measurement at high rotation speeds......................................................14
13. Technical Specifications ....................................................................................................... 16
13.1 Mechanical Dimensions.................................................................................................17
13.2 Mechanical Values and Load Limits...............................................................................18
Applied Measurements Ltd
appmeas.co.uk

2
Important Advice
The torque transducers of type DRDL are suitable for applications in laboratories
(for example in testing equipment) as well as in industrial environment.
The torque transducer is not designed with the usual safety factor (2...20),
favoring high sensitivity instead.
Pay attention to the overload limit.
You have to protect users from danger of being hurt (metal cover etc.)
The torque transducer is not designed for explosion endangered areas.
Warranty is void if opened or disassembled.
The transducer must only be opened by authorized personnel.
Applied Measurements Ltd
appmeas.co.uk

3
1. Introduction
The torque transducers have different shafts according to measurement range.
Torque transducers’ measurement unit is Nm.
2. Application
Torque transducers are able to measure clockwise as well as counter clockwise torque.
With clockwise torque the output is positive. The type label indicates the range of the
transducer.
The torque transducers measure static torque as precise as dynamic torque. Yet you have
to pay attention to the transducer’s signal rise time. It is indicated in the Data sheet.
The torque transducers are maintenance-free.
Handle the torque transducer with care, especially when transporting or mounting.
Because it can be damaged by hard shocks or by dropping to ground. Even a short peak
torque above the allowed overload capacity can damage the measure shaft. In cases
where this can happen you have to take precautions to avoid this.
The absolute maximum ratings regarding mechanical, thermal and electric parameters are
listed in the data sheet and must be observed in design, mounting and operating.
Applied Measurements Ltd
appmeas.co.uk

4
3. Technical Details
3.1 Torque Shaft
The shaft is fixed in the chasing with two ball bearings.
To measure the torque it causes a proportional bending of the shaft (within the elastic
range) and this is measured with applied strain gages. The strain gages are connected as a
wheatstone bridge. Additional angle or speed measurement is optionally available. (See
the data sheet.)
3.2 Case
The housings of the torque transducers are made of high-strength aluminum, the surface
is hard anodized for protection. The torsion shaft has two bearings in the housing via deep
groove ball bearings. The mechanical attachment of the transducer is carried out using a
foot attachment. Evaluation electronics for torque and speed or angle of rotation are
mounted on the housing.
3.3. Measurement
The torque causes the bending of the shaft and the strain gages. The resistance of the
strain gages changes proportional to the change of their length. And this is converted into
an electrical signal that causes a frequency modulation. This modulated signal is
transmitted optically to an electronic circuit in the casing. There it is converted back to a
proportional analog output voltage. This output voltage has a separate ground that is
electrically separated from the transducer’s supply.
For speed and angle measurement the pulses of a code wheel are encoded to a square
wave signal. With option speed measurement the output is 60 pulses per revolution, with
option angle measurement two square wave signals are provided with an output of 360
pulses per revolution. While clockwise rotation the signal of channel A is approximately 90
degrees ahead of channel B.
Applied Measurements Ltd
appmeas.co.uk

5
3.4 Disturbances and their Compensation
Avoid bending, axial and radial forces. When you have problems with this, use ETH
clutches.
To connect the transducer to a measurement unit you need a shielded cable.
The transducers are EMC-tested and are complying with EN 55011/03.91 (DIN VDE 0875-
11/07.92) respective prEN 50082-2/1993 (E DIN EN 50082-2/03.94).
4. Operating conditions
4.1. Environment Temperature
For best accuracy you have to meet the environment temperature specification. The
temperature should be constant or slowly changing. The specified temperature errors
apply only for changes less than 5 K/h. Radiation heat or cooling on one side has to be
avoided. Or appropriate precautions have to be taken.
4.2 Humidity and Dust
The torque converter corresponds to IP40 according to DIN 40050
Advice: Don’t let humidity seep into the transducer’s connector!
4.3 Chemically Resistance
The torque transducers are not resistant against chemicals.
4.4 Deposition
Dust, dirt and other particles mustn’t accumulate so that they can get into the ball
bearings and the connector.
Applied Measurements Ltd
appmeas.co.uk

6
5. Mounting
5.1 Precautions at Assembly
•Handle the transducer carefully
•Important Note:
Do not overload the transducer, not even temporarily. If possible, we suggest
connecting the transducer to a display unit before mounting, so you can watch the
torque while mounting.
•Avoid false axial and radial alignment.
•Provide good electrical ground contact to the casing.
5.2 General Mounting Instructions
Don’t mix up the transducer’s drive side and the measuring side, as this causes errors
(especially when accelerated).
When you look at the type label the measuring side is on the left.
The measuring side also can be identified by a deepening on the cover.
Bending, axial and radial forces are causing errors. Keep thermal expanding of the
construction in mind.
Applied Measurements Ltd
appmeas.co.uk

7
6. Measuring Chain
For measuring with the transducer, a whole measuring chain is necessary.
•Torque transducer
•Measuring cable
•Supply and display unit
You need a supply and display unit to supply the transducer with power and to display the
measured torque.
7. Connection
Connection of a torque transducer
Applied Measurements Ltd
appmeas.co.uk

8
7.1 Hints for Connection
Electric and magnetic fields cause interference with the measuring signal. This interference
is mainly caused by power cords, relays or motors installed nearby. Besides these,
interference can be caused by multiple grounding of the measurement chain on more than
one point.
Pay attention to the following:
•Use only shielded cables with low capacitance (like our measuring cables).
•Connect supply voltage correctly (no reverse polarity protection).
•Measuring cables shouldn’t be nearby high voltage or control cables.
•Magnetic radiation from transformers, motors or relays must be avoided.
•Don’t ground transducer and display unit multiple. Connect all devices of the
measurement chain to the same ground.
7.2 Connectors
The transducer is equipped with a 12 pin fitted connector.
7.3 Pinout of the Connector
The pinout of the connectors are on the following pages.
The transducer internally generates a galvanically isolated measurement signal.
The masses must not be bridged directly on the transducer, otherwise - depending on the
cable length to the supply and evaluation device - leads to measurement errors. If
necessary, however, these can be bridged directly on the supply and evaluation device.
The “control activation” is used to test the transducer. This emits its maximum signal of
e.g. + 10V (with right-hand load). The control level is 4.5V up to the supply voltage, the
reference ground point being the supply ground.
Applied Measurements Ltd
appmeas.co.uk
Questo manuale è adatto per i seguenti modelli
2
Indice
Manuali Trasduttore popolari di altre marche

Mianyang Weibo Electronic
Mianyang Weibo Electronic WB Series Manuale utente

ProMinent
ProMinent Dulcometer DMT Manuale utente

MKS
MKS MicroPirani 925 Series Come usare

WIKA
WIKA WU-20 Manuale utente

Alcatel Vacuum Technology
Alcatel Vacuum Technology BARATRON 622A Manuale utente

Camille Bauer
Camille Bauer SIRAX CH-5610 Manuale utente