Baoding Shenchen F Series Manuale utente

Shenchen Precision Pump
Manual of F series

Important Information
Please read the manual carefully before operating the product.
Warning
Tubing may have crack due to wear. It results in the overflow of fluid
from tubing. In that time human body and instruments may be damaged.
So user must check usually and change tubing in time.
Connect directly the power line to the wall socket, and avoid using the
extense electric line.
If the power line or plug had wear and other damage, please hold the plug
to unplug it, not hold the line.
If following situations happened, please turn off the electric power and
unplug the plug, holding the plug and not the line.
1. Fluid splash on the pump.
2. You think the pump need to maintain or amend.
The user’s power socket must have ground wire, and have reliable
grounding

Shenchen Pump F Series
Catalogue
1. F Series Product Introduction....……………………………………………-1-
2. F Series Product Appearance……………………………………………….-1-
3. F Series Keypad Instruction………………………………………...………-2-
4. F series Interface Composition……………………………………...……...-2-
5. External Control Interface Instruction ……………………………………-10-
6. F Series Technical Specification…………………………………………..-14-
7. F Series Function and Features……………………………………………-15-
8. Product Dimension………………………………………………………...-16-
9. Product Maintenance ……………………………………..………………-18-
10. Warranty and After Sales Service………………………...……………….-19-

Shenchen Pump F Series
- 1 -
1. FSeries Product Introduction
F series of peristaltic pump, with 4.3 inch color touch screen control, graphical
interface, animation display working status, with the screen display filling
parameters and system settings, attention to user experience, suitable for
pollution-free, high-precision filling liquid, filling accuracy up to 0.5 %-1%.
2. FSeries Product Appearance
A——Pump Head
B——Drive
C——Power Switch
D——Power Socket
E——External Control Input Port
C
D
E

Shenchen Pump F Series
- 2 -
3. FSeries Keypad Instruction
MAIN
Calibration
System Set
PumpHead
&Tube Suck-Back
Angle Date
&Time
Common
Mode
Add
Mode
Monitoring Disp. Vol. Disp. Time Pause TimeDisp. Vol.
Run/Pause Button:In dispensing mode, after setting the dispensing data,
press run/pause button, the pump start work; press run/pause button again,
pause the current working. After press the run/pause button, except calibration
and monitor button all other buttons on main interface become gray, can not be
used.
In transferring mode, press run/pause button, pump start working, now it do
not have pause function.(In system setting interface, users can select working
mode: dispensing mode or transferring mode. )
Stop Button:Press stop button, stop working. Forbidden buttons can be used
on the main interface.
Full Speed Button: Press this button at the stop state, the filling unit will
running with 400rpm. This button is for washing tube and filling liquid
rapidly.
CW/CCW Button: Press this button, the motor will change running direction.
4. F Series Interface Composition
MAIN
Calibration
System Set
PumpHead
&Tube Suck-Back
Angle Date
&Time
Common
Mode
Add
Mode
Monitoring Disp. Vol. Disp. Time Pause TimeDisp. Vol.

Shenchen Pump F Series
- 3 -
FSeries Operation Interface Instruction:
4.1.Booting Interface:
After power on the system, enter the welcome interface, Click anywhere or wait
for 2.5 seconds, the system will automatically enter the English main interface.
4.2.Main Interface:
Main Interface Composition as below:
2012-3-15 8:45:05 AM
00.00ml /
C
F
KJI
G H
B A D E
Disp. Vol. 10.00 ml
Disp. Time 1.00 s
Pause Time 1.00 s
Copy CNT 0000/0005
Pump Head YZ1515x
Tube Size 25#
Suck-Back 0.00
Work Status
A. Real-time Display Volume: Select the specific mode and display
content in the system setting interface. In transfering mode, display the
speed or flow rate in real time, you can modify the motor speed or flow
rate by directly clicking on A area. In dispensing mode, display
dispensing volume or current total dispensing numbers.
B. Real-time Dynamic Display: Display the dispensing and transferring
state and monitor result
C. Real-time Parameter Display
In dispensing mode: Display dispensing parameters, include dispensing
volume, current dispensing time count down, current pause time count
down, copy numbers.
In transferring mode: Display the current transferring total volume,
and current working time. The total transferring volume maximum is
9999L. The current working time maximum is 9999 hour. Each one of it

Shenchen Pump F Series
- 4 -
reach the limit, both data will back to zero. Also can set the data back to
zero in the system setting.
D. Set Parameter Display: This area display current working state, pump
head, tubing size and suck-back angle
E. Date and Time Display: Display the current date and time, can amend it
in system settings
F. System Setting Button: Press this button enter system setting interface,
include set the pump head and tube size, set the suck-back angle, set
common mode, set current date and time.
G. Flow Calibration Button: Press this button enter calibration interface.
H. Real-time Monitor Button: Press this button enter monitor interface,
you can check the flow rate, motor speed and dispensing efficiency here.
I. Dispensing Volume Button: Press this button, the numerical keyboard
come out, input the dispensing volume here
J. Time Button: Press this button, the numerical keyboard come out, input
the filling time. Click confirm, it will ask whether continue to input the
pause time, choose ‘YES’, you can input the pause time, choose ‘NO’,
back to the main interface
K. Copy Numbers Button: Press this button, the numerical keyboard come
out, input the copy numbers. If input ‘0’, the copy number is unlimited
4.3. Basic Configuration Interface:

Shenchen Pump F Series
- 5 -
On the main interface, click System Setting button, click Configuration button,
enter Configuration interface.
Click Pump Head, Tube Size drop-down menu choose pump head and tube.
Reference flow rate display the minimum and maximum flow rate of the pump
head and tube.
4.4. Mode Setting Interface
In main interface click ‘ System setting’ button, click’ Mode setting’ button, enter
mode setting interface. Users can choose transferring mode or dispensing mode
freely. According with user’s request, in transferring mode, can choose display
motor speed or flow rate; in dispensing mode, can choose display filling volume
or copy numbers. The display mode decide the content display in A area of main
interface, can display current dispense volume or total numbers. Maximum
dispense numbers is 9999, manual reset.
4.5 Back Suction Angle Setting Interface
Back suction angle setting interface as below:

Shenchen Pump F Series
- 6 -
Please enter back suction angle
360.00
Back Suction angle
range: 0~360
Settings can raise the
dispensing accuracy,
Please re-calibrate the
flow rate.
!
CancelOK
Click the System Setting button on the main interface, click Suck-back Angle
button, enter suck-back angle setting interface. Click Set Suck-backAngle button,
the numerical keyboard will come out, input the suck-back angle and click
‘Confirm’ button, all the filling unit will work with this angle. Click ‘Cancel’
button back to the system setting interface.
4.6.Common Mode Interface
Common Mode Interface as below::
Add OK
PumpHead Tube Disp. Vol. Pause Time Suck-Back
YZ1515x
YZ1515x
14#
16#
2
7
2
2
0
0
Disp. Time
1
1
Cancel
ClearDec. 》
《
Click System Setting button on the main interface, click Common Mode button,
enter choose common mode interface.
Add Button: Click this button to add one common mode, in dispensing
mode .click ADD button to pop up and dialog box to ask if to add the curent
setting to common mode.Clike YES to add this mode. It can totally save 60
modes.

Shenchen Pump F Series
- 7 -
Delete Button:Choose one common button, click Delete button, it will ask whether
delete, click ‘YES’, then you can delete this mode.
Clear Button:Click this button, it will ask whether clear all data, click ‘YES’, then
empty all the common mode.
Selecting Button: Choose one of the common mode, click ‘OK’ button, then it will
back to the main interface. The filling parameter is same as the one you chose
from the common mode.
Cancel Button: Click this button back to the system setting interface.
Page Up/Down Button: Can check previous page or next page common mode.
4.7 Communication Settings Interface
The communication setting interface as below;
Click ’System Setting’button in main interface, click ‘Communication setting’
button,enter communication setting interface. Users can select baud rate and
communication interface RS232/RS485. Local address can be any number between
1 and 32.
4.8 Flow Calibration Interface
Flow Calibration Interface as below:
Questo manuale è adatto per i seguenti modelli
6
Indice
Altri manuali Baoding Shenchen Pompa dell'acqua
Manuali Pompa dell'acqua popolari di altre marche

Sykes AmeriPumps
Sykes AmeriPumps GP100M Guida alla risoluzione dei problemi

DUROMAX
DUROMAX XP WX Series Manuale utente

BRINKMANN PUMPS
BRINKMANN PUMPS SBF550 Manuale utente

Franklin Electric
Franklin Electric IPS Manuale utente

Xylem
Xylem e-1532 Series Manuale utente

Milton Roy
Milton Roy PRIMEROYAL Manuale utente