2.6 Recommended Fitting for A9
A87 M36 × 1,5 outside thread and pyramid receiver
A22 Threaded Adapter, M36 × 1,5 and pyramid
A23 Threaded Adapter, M36 × 1,5 and pyramid receiver
3. Indication/ Contraindication
Indication:
•Amputation of lower extremities
Contraindication:
•not known
4. Side effects
There are no known side effects
5. General safety instructions
•This medical product is designed for: single patient, multiple use
•Fitting/service of this medical device is only allowed by a certificated orthopedic
professional
•Improper modification or application to the product is not allowed In case of non-
observance, the function of the product may be impaired, so that product liability
is excluded
•The maximum weight loads are listed under “2 Technical Data” Additional loads
must be taken into account when calculating the weight
•The sliding adapter VA9-Al is intended solely for use during the trial run in the
orthopedic workshop It is absolutely unsuitable for use in actual prostheses
Order No. Material Weight Limit Weight Article No.
A22-Ti Titanium < 150 kg 40 g 4 114 090 99 33 000
Order No. Material Weight Limit Weight Article No.
A23-Ti Titanium < 100 kg 51 g 4 114 090 77 33 000
Order No. Material Weight Limit Weight Article No.
A87 Steel < 150 kg 55 g 4 620 040 00 00 000
6. Installation
6.1 Installation A1 and A2
The Shifting Adapters VA1-Ti and VA2-Ti have a shifting possibility of approx 10 mm in
sagittal direction
6.1.1 Connection to the Ø 30 mm Tube Adapters
Before fitting the tube, first loosen the Hexagon socket head cap screw (M33) with two
turns Be certain that the cut surface is even and the longitudinal axis of the tube is
rectangular Smoothen the area of the tube cut on the in- and outside Insert the tube
into the clamp adapter until the tube reaches the bottom of the clamp Fix the tube in
the clamp adapter by tightening the clamp screw with a torque wrench
6.1.2 Shifting of the Shifting Adapter
To shift both ends of the Shifting Adapter VA1-Ti against each other, first loosen the
headless screw in which direction you want to shift Reclamp the adapter by tightening
the opposite headless screw Take care to always tighten all headless screws
6.1.3 Instructions of Use - eadless Screws
By using the adjusting screws M28-12 or M28-16, static corrections are possible at any
time When replacing or removing a module, the previously set position is maintained
if only two adjacent adjusting screws (not opposite) are removed, which are screwed in
at the deepest
If the adjusting screws are too long (damage of the foam cover) or too short
(reduced stability) these have to be replaced with appropriate ones
Use a toque wrench to tighten the headless screws
Tightening torque: 12 Nm
When the prosthesis is completed secure thread with Loctite® 222
6.2 Installation A9
The shifting elements enable/ allow adaptation via 4-hole attachment or adapters with
M36 fine thread
Screw the required adaption to the elements of the shifting adapter
Fix the shifting adapter in neutral position
Tighten the six lateral cylinder head screws
Tightening torque: 5 Nm