BEAMEX PGPH Manuale utente

INSTRUCTION MANUAL
HIGH PRESSURE PNEUMATIC PUMP
MODEL PGPH


Dear user,
We have made every effort to ensure the accuracy of the contents
of this manual. Should any errors be detected, we would greatly ap-
preciate to receive suggestions to improve the quality of the contents
of this manual.
For more detailed technical data about the Instruction manual for
Beamex®PGPH High Pressure Pneumatic Pump, please contact the
manufacturer.
© Copyright 2013, 2015
BEAMEX OY AB
Ristisuonraitti 10
FIN-68600 Pietarsaari
FINLAND
Tel +358 - 10 - 5505000
Fax +358 - 10 - 5505404
Internet: http://www.beamex.com
8804400 / PGPHuEng / Version 1.1


CONTENTS
DESCRIPTION ............................................................................................. 1
Standard Accessories.........................................................................................1
Optional Accessories ..........................................................................................1
Connections and Parts........................................................................................2
Connection Diagram ...........................................................................................3
OPERATION.................................................................................................3
Keep in Mind.......................................................................................................4
TROUBLESHOOTING/MAINTENANCE .....................................................5
Seal Replacement...............................................................................................6
Cleaning the Pump..............................................................................................6
Leak Tests...........................................................................................................6
Leak Test for Positive Pressure..................................................................6
Leak Test for Negative Pressure (Vacuum)................................................6
SPECIFICATIONS........................................................................................7
The Pump Unit....................................................................................................7
Pressure Measurement Hoses (Optional Accessories).......................................8
Connections................................................................................................8
WARNINGS AND CAUTIONS .....................................................................9

Hose/connector
set for connect-
ing a calibrator's
high pressure
module to
PGPH
Hose/connector
set for connect-
ing the instru-
ment to be cali-
brated to PGPH
Connections when calibrating a transmitter using an internal pressure module.
Communication
cable.
Hose/connector
set for connect-
ing the instru-
ment to be cali-
brated to
PGPH.
Connections when calibrating a transmitter using an external pressure module.

1
DESCRIPTION
The PGPH High Pressure Pneumatic Pump is designed to manually generate up to
140 bar (2000 psi) gauge pressure or up to -0.95 bar (-14 psi) vacuum for quick and
accurate calibration of pressure gauges, transducers and other pressure measure-
ment instruments.
Standard Accessories
The standard accessories are as follows:
One G ¼’’ (male) plug for Pressure connectors
A set of seals (o-rings), type: NBR70 for Pressure connectors (size 6 × 2).
Guarantee leaflet
This Instruction Manual
Optional Accessories
The optional accessories are as follows:
ITEM
ORDER CODE
Carrying case
8003315
Service seal kit
8003180
Tube of grease for the Fine Adjust stem
0005180
400 bar hose/connector set for connecting the instru-
ment to be calibrated to PGPH
8003380
400 bar hose/connector set for connecting a calibra-
tor's high pressure module to PGPH (*
8003365
*)Not needed when used high pressure module is an External Pressure Module (EXT).

2
Connections and Parts
Pump lever
Shut-off valve. Isolates the pump form the measurement circuit.
Release valve. Clockwise to close, anticlockwise to open.
Pressure connector, G ¼", for instrument / calibrator. Hand tightened.
Vent connection/hole
Fine Adjust handwheel. Rotating clockwise increases pressure.
Pressure connector, G ¼", for instrument / calibrator. Hand tightened.
Equivalent with Pressure connector
Pressure/Vacuum selector. Pull out for pressure and push in for vacuum.
Do not change when pressurized!
Selector lock
Pump body

3
Connection Diagram
Atmosphere
Pump
Shut-off valve
Release valve
Calibrator
(reference)
Fine Adjust
Instrument to be calibrated (DUT)
Pressure/Vacuum selector
OPERATION
Preparation
1. Select Pressure Mode .
2. Open Shut-off valve (counterclockwise) and Release valve.
3. Rotate Fine Adjust to mid position
4. Close Release valve
5. Connect instrument to be calibrated (Device Under Test, DUT) to Pressure con-
nector and the calibrator (pressure module, reference) to Pressure connector
. The location of the instrument and the calibrator may also be interchanged.
Pressure connectors and are equivalent. See also Leak Tests on page 6.
Calibration
6. Pump pressure/vacuum .
7. Close Shut-off valve (clockwise). See also notes further on.
8. Adjust pressure by rotating the Fine Adjust handwheel.
9. Repeat stages 6 to 8 until instrument's maximum pressure / deepest vacuum is
reached
When maximum pressure / deepest vacuum is reached continue as follows:
10.Gently open Release valve to release pressure/vacuum and close when ap-
proaching next calibration point.
11.Adjust pressure by rotating the Fine Adjust handwheel.
12.Repeat stages 10 and 11 until atmospheric pressure (or the instrument's meas-
urement span's pressure closest to atmospheric pressure) is reached.
End by opening Shut-off and Release valves. If you want to do several repeats,
restart from stage 3.

4
Keep in Mind
Immediately after a pressure change, the pressure may change slightly due to the
stretching/shrinking of the pressure measurement hose. Thermodynamic effects may
also cause pressure variation. In that case, adjust the pressure back to required value
using the Fine Adjust handwheel.
Do not exceed the max. operating pressure of the pump and the hose. Also observe
the pressure limit set by environmental conditions.
Ensure that there always is a reliable pressure indicator connected to the pump when
the pump is used.
If there is a strong counterforce while operating the pump lever, but no pressure
change is indicated, stop pumping and locate the fault. Always keep a reliable indi-
cator connected to the measurement system.
If the pump fails to indicate a pressure change after considerable pumping action of
the handle, check to assure that the connections between the pump, the hose end
and the attached instrument(s) are tight and retry pumping. Also check to assure that
all the unused output connectors are properly plugged.
If a pressure increase still cannot be obtained, it is possible that one or more of the
seals in the Pressure Pump is leaking and needs to be replaced.
Do not continue pumping if the functionality of the pump is not normal. Locate
the fault before you continue using the pump.
A full list of warnings is at the end of this manual.
Indice
Altri manuali BEAMEX Pompa dell'acqua
Manuali Pompa dell'acqua popolari di altre marche

Sykes AmeriPumps
Sykes AmeriPumps GP100M Guida alla risoluzione dei problemi

DUROMAX
DUROMAX XP WX Series Manuale utente

BRINKMANN PUMPS
BRINKMANN PUMPS SBF550 Manuale utente

Franklin Electric
Franklin Electric IPS Manuale utente

Xylem
Xylem e-1532 Series Manuale utente

Milton Roy
Milton Roy PRIMEROYAL Manuale utente