Page 8/ 12
4. Installation
Before beginning the installation, always:
Inspect the packaging for damage –When the instrument is received, inspect the item carefully to check
any transit damage. In the event of damage, always report the damage to the shipping carrier and your
local representative immediately.
Included in the package:
1. Stainless Steel Perforated shelf –2 Pcs
2. Power Cord –1 Pcs
3. Main Power Adapter 12V DC(100 ~ 230V, 50 ~60Hz) –1Pcs
4. 12V DC Car Jack –1 Pcs
5. Stainless Steel Water Container for humidity- 1P
6. CO₂Gas Tubing with in-line CO₂Gas filter –1Pcs
7. Operating Manual- 1Pcs
Above parts are packed inside incubator. When received, open the door of incubator and remove all
parts to check confirm receipt.
Cleaning before use
Before conducting cell culture in this mini incubator, It is recommended to clean up entire chamber and
shelves and water container by using soft cloth with at least 70% Ethanol mixed of 30% distilled water.
Do not make hot air or H₂O₂or UV decontamination in this plastic housing incubator.
Installation Procedure:
1. Place the incubator at the desired location.
Always avoid placing the incubator in:
- An area near equipment generating heat or cold air to incubator.
- Direct sunlight
- An uneven surface or table.
- A place with heavy vibration
- A place with little air ventilation space behind the incubator.
2. Place the shelves and water container as shown in chapter “CONFIGURATION”. If desired, place
the stainless steel water container at bottom of chamber toward the back of incubator (in order to get it
to be close to circulation fan). This increases the humidity.
Distilled water is recommended in order to avoid contamination and corrosion.
If possible, use warm (~37℃) distilled water for immediate humidification.