BM3www.BIO2NET.com
RevisionA
-8-
1. BesurethatAC power supplylineisappropriatetouse.
(AC 100 - 240V)
2.Besurethat thepower sourceistheonesuppliedfromBionet.
(DC 18V, 2.5A)
3.Besurethattheentireconnection cableof thesystemisproperlyandfirmly
fixed.
4.Besurethattheequipmentiscompletelygrounded.(Otherwise,noise
could result.)
5.Theequipmentshould notbeplaced inthevicnityofelectricgenerator,X-
ray,broadcastingapparatustoeliminatetheelectricnoiseduringoperation.
Otherwise, it may causeincorrect result.
NOTE
TheEquipmentshouldbeplacedfarfromgenerator,X-rayequipment, broadcasting
equipment,or transmitting wires, soastopreventtheelectricalnoisesfrombeing
generatedduringtheoperation,When thesedevicesare near theEquipment, itcan
produceinaccuratemeasurements. For BM3,bothindependentcircuitandstable
groundingare essentiallyrequired.Intheeventthatsamepower sourceis shared
withother electronic equipment, it canalsoproduceinaccurateoutput
Warning
Donot usetheSystemfor anymonitoringprocedure on apatientifthemonitor isnot
working properly,or ofit is mechanicallydamaged.Contactthehospitalbiomedical
engineer, or yoursupplier