- Before using RIGHTEST ALPHA Blood Glucose Monitoring System to test your blood glucose,
please read all of the info and conduct all of the tests including the quality control test.
- We recommended you to perform the quality control test regularly to ensure the test results are
accurate.
- The RIGHTEST ALPHA Meter can only be used with RIGHTEST ALPHA Test Strip. Other brands' test
strips should not be used under any circumstances. The use of other brands' strips may give
inaccurate results.
- If RIGHTEST Meter and Test Strips are exposed to temperature environments out of range for the
meter – below 10°C (50°F) or above 40°C (104°F) – please wait at least 30 minutes before testing
again.
- Follow all environmental protection regulations when disposing of batteries, strips and lancets.
- RIGHTEST Blood Glucose Monitoring System has not been validated for use on neonates. DO NOT
use it test for neonates.
- RIGHTEST Blood Glucose Monitoring System is NOT intended for arterial blood testing.
- Avoid contact with dripping or splashing liquids.
- Once patient is suspected as rare disease (e.g. galactosemia), the glucose result should basedon
laboratory test.
54
Caution Caution
- The minimum blood sample size for testing is 0.75 µL :( )
Blood sample size above 3.0 µL might contaminate the test strip port and the meter.
Blood sample size below 0.75 µL may cause an inaccurate result or may prevent a meter reading.
An Er4 reading will be displayed if the sample size is too small. In this case, repeat the test with a
new test strip.
Sample Size Example 0.75 µL 1.0 µL 1.5 µL 2.0 µL 3.0 µL
Important Safety Notes
- All parts of the kit are considered biohazardous and can potentially transmit infectious
diseases, even after following the cleaning and disinfecting procedures. Please refer to the
section "Maintain the products" on page 35.
- Users should wash their hands thoroughly with soap and water before and after handling the
meter, lancing device, or test strips.