1009
If you find the actuator suddenly cannot move
SET2:
If all the cables are firmly connected correctly, the
table still doesn’t work, then open the shielded cable
cover and unplug the actuator cables from the control
box. Keep on pressing the foot-control, and measure
the output voltage of the socket on the control-box
using a multi-meter. If the meter reading is 29 volts, it
means the actuator is defective, and you need to
replace the defective actuator. If the multi-meter does
not read 29 volts DC, the control box or the foot-pedal
may be defective.
blue yellow black red
1.
multi-meter
Disconnect the main power.
Check all the cables and reinsert them if needed.
Remove any weight that may be on the table.
Allow the system to rest for 20 to 30 minutes with the main power unplugged.
Section IV : How to Care for Your Upholstery
Avoid storing your table in extreme cold temperatures as excessive cold will cause cracking.
Avoid storing your table in extreme hot temperatures as excessive heat will cause the upholstery to
stretch and pucker. Be sure to keep your table away from heaters, electrical appliances and direct
sunlight.
Protect your table and accessories with sheets and fleece.
We recommend our three-piece Flannel Sheet Set and Fleece Sheet Set.
We recommend a mild mixture of anti-bacterial dish soap with water in a spray bottle. Spray the
table with the solution and wipe down the table with a clean damp towel.
Use both ends of the table. Occasionally, switch the face cushion/head extention from one end
of the table to the other to avoid uneven use.
Protect your table from general abrasions, cuts and punctures.
5IFVTFPGBMDPIPMXJMMESZPVUUIFWJOZMBOEDBVTFJUUPDSBDL
2. Prolonged exposure to extreme temperatures may cause damage to vinyl
In an effort to ensure your products' quality and to help you maintain the life of
your table, we have tested several preventative steps in order to advise you how
to protect the upholstery on your table and accessories:
Step1: Disconnect the power, and take off the shield cover from the control box using a slotted
screwdriver.
Step 2: Pull out the broken foot-control plug from the control box and insert the plug of new
foot-control.
Step 3: Try the new foot control, if it works, replace shield cover.
Check the outlet to ensure it has power
and the cable is plugged in on both ends.
Check the control box to ensure the cables for actuator,
foot-control is firmly connected to the control box.
blue brown pink
SET1:
*The blue label port in charge of height control
*The brown label port in charge of chest part control
*The pink label port in charge of control paddle
If this doesn’USFTPMWFUIFQSPCMFNUIFOUIFDPOUSPMCPY
NBZCFEFGFDUJWF
If you happen to have another electric table, you may diagnose the problems by
interchanging the actuators, control box and foot pedal to help with the diagnosis.
Please note! The plugs for foot-pedals are interchangeable. This means you may replace the
pedal that doesn’t work with another pedal to help with the diagnosis.
You will have a 2 year warranty for the Motor Actuator and the Control box from the
date of purchase. Please contact the retailer or wholesaler to apply for a new part for
replacement if the table does not work after the above diagnosis.