cleanAIR CLF 475 Manuale utente

Clean Air Techniek B.V.
User manual
English Version 5.0
Series CLF
Laminar crossflow-cabinet
Types:
CLF 475
CLF 575
CLF 675
Offers protection for product
Before using the cabinet, please read this
manual carefully. The required instructions
have to be carried out first.
Clean Air Techniek B.V.
Kuipersweg 37
3449 JA Woerden
The Netherlands
Phone: +31 (0) 348 41 11 14
Fax: +31 (0) 348 42 26 84
Internet: www.cleanair.eu

UM-CLF-XX-EN05.doc
2
Preface
Clean Air provides a complete range Laminar Airflow Cabinets for handling of non-hazardous
materials, which require a clean particle free environment. The cabinets offer the highest product
protection. The horizontal airflow prevents contaminated outside air to enter the workstation's
work area, ensuring product protection.
Its application is in the field of pharmaceutical industry, biotechnology, pharmacies, hospital and
university laboratories as well as tissue culture and electronic laboratories.
The CLF is a high-quality product constructed with high-grade components and materials, with
new techniques, such as:
Microprocessor control with a LCD display for the interface with the user;
Automatic up speeding fan; when there is increasing resistance in the filter the necessary
airspeeds are maintained;
Integrated drain on the backside of the working surface to prevent spilled liquids to leak in
to the HEPA filter;
White laminated wooden or stainless steel (304) working surface;
All internal areas of the working space are in line with the effective HEPA filter area, to
obtain a laminar flow;
Side windows of safety glass for a good overview.
Furthermore the CLF is a service friendly cabinet, produced in The Netherlands.
We thank you for buying this cabinet and wish you good times working with this cabinet.
Clean Air Techniek B.V.
Version table
Vers. 5.0 25-11-09 MPE Document structure updated
Version: Date: From serial
number:
By: Check: Change:
©Copyright 2009
All rights reserved. Nothing from this edition may be copied, or used in an automated database, or made public, either electronic,
mechanical, by photocopies, recordings, or in any manner, without preceding written permission of Clean Air Techniek B.V. This also
counts for the accompanying figures and/or plans.
Clean Air Techniek B.V. is entitled to change parts on each desired moment, without preceding or direct notification of the customer.
The content of this user manual can be changed without preceding warning as well. For information concerning maintenance activities
or repairs which are not mentioned in the user manual, please contact the service organization.
This user manual has been put together with all possible care, but Clean Air Techniek B.V. cannot take the responsibility for possible
mistakes in this user manual or for the consequences of it.

CLF Laminar crossflow-cabinet Clean Air Techniek B.V.
UM-CLF-XX-EN05.doc
3
Index
1Introduction _______________________________________________ 6
1.1
CLF__________________________________________________________________6
2Safety regulations ___________________________________________ 7
2.1
Use in conformity with regulations _________________________________________7
2.2
Modifications and changes to the cabinet____________________________________7
2.3
Authorized persons _____________________________________________________7
2.4
General safety aspects __________________________________________________8
2.5
Explanation user safety symbols ___________________________________________9
3Cabinet __________________________________________________ 10
3.1
Assembly-instruction ___________________________________________________10
3.2
Transport instructions __________________________________________________11
3.3
Dismantling & Disposal _________________________________________________11
4Product description _________________________________________ 12
4.1
Type plate ___________________________________________________________12
4.2
Front panel __________________________________________________________12
4.3
Worktop _____________________________________________________________13
4.4
Plenum-construction ___________________________________________________13
4.5
Crossflow filter________________________________________________________13
4.6
Pre-filter _____________________________________________________________13
4.7
Support frame (optional)________________________________________________13
4.8
Main power connection _________________________________________________14
4.9
Electronics ___________________________________________________________14
4.10
Operating panel _______________________________________________________15
4.11
FL-lighting ___________________________________________________________15
4.12
Fan _________________________________________________________________15
4.13
Optional features ______________________________________________________15
5Control___________________________________________________ 18
5.1
Using the cabinet______________________________________________________18
5.2
Display interface ______________________________________________________18
5.3
Starting up the cabinet _________________________________________________19
5.4
Locking functions cabinet _______________________________________________20
5.5
Power save mode _____________________________________________________21
5.6
Turn off _____________________________________________________________21
5.7
Control UV (optional)___________________________________________________22
5.8
Control sockets (Optional)_______________________________________________22
5.9
Menu functions _______________________________________________________23
5.10
General Information ___________________________________________________23
5.11
Information status cabinet ______________________________________________32
5.12
Settings _____________________________________________________________35
6Alarms ___________________________________________________ 38

CLF Laminar crossflow-cabinet Clean Air Techniek B.V.
UM-CLF-XX-EN05.doc
4
6.1
Flow alarms __________________________________________________________38
6.2
Other alarms _________________________________________________________39
7Cleaning__________________________________________________ 41
7.1
Cleaning the installation ________________________________________________41
7.2
Cleaning the UV-light (optional) __________________________________________41
8Small maintenance and testing ________________________________ 42
8.1
Periodic maintenance __________________________________________________42
8.2
Replacement of the pre-filter ____________________________________________42
8.3
Replacing the fluorescent-light (FL light) ___________________________________43
8.4
Service contact _______________________________________________________44
8.5
Service contract _______________________________________________________44
8.6
Testing ______________________________________________________________44
9Technical specifications ______________________________________ 45
9.1
Physical surrounding ___________________________________________________45
9.2
General specifications __________________________________________________45
10 Trouble shooting ___________________________________________ 48
Appendix I: Support frame CLF ____________________________________ 49
Appendix II: Replacement of the HEPA-filters ________________________ 50
1.
When does a HEPA filter need to be replaced? ______________________________50
2.
replacing the HEPA-filter ________________________________________________50
Appendix III: Efficiency test of the HEPA-filters _______________________ 52
Requirements ______________________________________________________________52
Appendix IV Measuring method air velocities _________________________ 53
Air velocities crossflow _______________________________________________________53
Appendix V: Lay-out boards ______________________________________ 54
Appendix VI: Alarm-test procedure_________________________________ 55
Appendix VII: Stickers __________________________________________ 57
Appendix VIII: Information installed options _________________________ 58
Installed options microprocessor _______________________________________________58
Appendix IX: User code__________________________________________ 61
Activating the user code _____________________________________________________61
Changing the user code ______________________________________________________62
Appendix X: Required specifications ________________________________ 63
Overview of the required specifications__________________________________________63
Appendix XI: Recommended spare parts list _________________________ 64
Appendix XII: TUV Low Pressure Mercury Lamps (Optional) _____________ 65
Appendix XIII: Statement of the agreement for machines _______________ 66
Appendix XIV: Memo____________________________________________ 67

CLF Laminar crossflow-cabinet Clean Air Techniek B.V.
UM-CLF-XX-EN05.doc
5
Index________________________________________________________ 68

CLF Laminar crossflow-cabinet Clean Air Techniek B.V.
UM-CLF-XX-EN05.doc
6
1Introduction
1.1
CLF
The laminar crossflow-cabinet CLF has as purpose that the working surface is kept clear of
particles by means of a clean laminar crossflow of air. By applying a HEPA-filter (High Efficiency
Particulate Air) it is possible to work in this cabinet under conditions which are sterile, dust-free, or
aseptic. The cabinet, therefore, offers a product protection.
Short description of the CLF
•At the top of the cabinet the inflow air is drawn
through a pre-filter by a fan.
•This fan blows the air in the correct proportion to
the crossflow HEPA-filter. Because the crossflow
air passed through a HEPA-filter (High Efficiency
Particulate Air), the working surface is free of
particles.
•The airflow in the workspace is crosswards, thus
preventing background air to flow into the class A
workspace.
•The air flow leaves the cabinet by the work
opening.
Figure 1: Sectional view CLF

CLF Laminar crossflow-cabinet Clean Air Techniek B.V.
UM-CLF-XX-EN05.doc
7
2Safety regulations
ATTENTION!
Before starting service with the cabinet always ask the responsible person if there is no danger
for the service engineer, labor personnel, the laboratory and the surrounding of the cabinet.
2.1
Use in conformity with regulations
Personnel that is not properly informed about the safety regulation and/or does not meet the
requirements mentioned in
§2.3
Authorized persons
on page 7
, is not allowed to use the cabinet.
Improper use of the cabinet is not allowed.
2.2
Modifications and changes to the cabinet
In order to operate the cabinet safely changes and/or alterations shall only be made by Clean Air
Techniek B.V., or shall be made after consultation with and permission in writing from Clean Air
Techniek B.V. If modifications and changes are made on own initiative, without permission from
Clean Air Techniek B.V., the parts of the cabinet concerned are no longer covered by the warranty.
Clean Air Techniek B.V. cannot be held liable for the consequences of the modifications and the
dangers, which might possibly occur as a result. The modifications mentioned here include the
connection of the cabinet to an exhaust system.
2.3
Authorized persons
Operating the cabinet
The cabinet shall only be used by personnel which:
is familiar with the content, regulations and warnings which are mentioned in the user manual;
is familiar with how to operate the cabinet;
is familiar with the start up procedure, the stop procedure and knows how to react in case of
emergency;
is familiar with the regulations to keep the safety in all circumstances;
is capable to end a disturbance.
Maintenance and repairs general
Maintenance shall only be performed by qualified service engineers, who are sufficiently trained by
Clean Air Techniek B.V. to:
Estimate and avoid the dangers of the cabinet;
Estimate the consequences of their actions.
Before starting maintenance to the cabinet (mechanical and/or electric) disconnect the power
supply. Pay attention to rotating parts, after switching off the cabinet the fans will still rotate
for a while.

CLF Laminar crossflow-cabinet Clean Air Techniek B.V.
UM-CLF-XX-EN05.doc
8
Maintenance and repair of the electrical cabinet
Maintenance shall only be performed by professional electro-technical service engineers, who are
sufficiently trained by Clean Air Techniek B.V. when:
They can estimate and avoid the dangers of the cabinet;
They can estimate the consequences of their actions;
They are sufficiently informed about the operation of the cabinet and the operation of the sub-
systems.
2.4
General safety aspects
1. The cabinet shall never be used unless the fan is switched into work mode and there is no
alarm.
2. When the cabinet is not used, the cabinet should be switched in to power-save mode or
closed/covered, to prevent that dust particles etc. from the room can get into the cabinet.
3. Always try to limit the amount of equipment in the workroom. Each object in the workroom
will disturb the pattern of the airflow and a large number of equipment may lead to serious
disruptions.
4. It’s is important that the cabinets will be kept clean. Even the smallest contamination which
is not visible will be easily forgotten. It is important to clean often the working area,
especially the work top.
5. Be careful with flammable liquids such as alcohol. The airflow is in the direction of the
person, so be very careful, or even better; prevent to use this kind of liquids.
6. The use of hand gloves and over sleeves is recommended.
7. Always use a chair with the correct height.
8. Transport and installation of the cabinet is only allowed for authorized staff. Please contact
the manufacturer or the seller of the products

CLF Laminar crossflow-cabinet Clean Air Techniek B.V.
UM-CLF-XX-EN05.doc
9
2.5
Explanation user safety symbols
GENERAL DANGER
You can harm yourself and others seriously if you
do not follow the procedures carefully
HIGH VOLTAGE
DANGER OF EXPLOSION
DANGER BY INDUSTRIAL VEHICLE

CLF Laminar crossflow-cabinet Clean Air Techniek B.V.
UM-CLF-XX-EN05.doc
10
3Cabinet
GENERAL DANGER
Be careful: high centre of gravity, crossover point.
The cabinet must be installed in a safe way. Clean Air Techniek B.V. supplies an optional support
frame, which is designed for this purpose.
When the cabinet is purchased without a support frame, it is advised to discuss the installation
with the supplier.
3.1
Assembly-instruction
The cabinet has to be placed on the support frame recommended by the supplier. The assembly
needs to be done as specified in instruction underneath.
Assembly support frame (Optional)
Build the support frame (see also
Questo manuale è adatto per i seguenti modelli
2
Indice