dacell DN-130L Manuale utente

Ver 4.0
MODEL : DN-130L
DIGITAL INDICATOR
USER MANUAL
DACELL CO.,LTD.
Address : 681-1 Cheoksan-Ri, Nami-Myeon, Cheongweon-Gun,
Chung-buk
TEL :82-43-260-2242
FAX : 82-43-260-2245
http://www.dacell.com

1
Contents
1. Sepcial Features 2
1-1. High-speed conversion method 2
1-2. Calibration method 2
1-3. HOLD & PEAK HOLD 2
1-4. Comparison Output 2
1-5. Data Back-up 2
1-6. Watch dog 2
1-7. Option 3
1-8. Power 3
2. Attention 4
2-1. Attention for installation 4
2-2. Attention for use 4
3. Specification 5
4. Front Panel 6
5. Rear Panel 7
6. Wiring Diagram 9
7. Components & Function 10
7-1. Flow Chart 10
7-2. How to use Hold mode 11
7-3. How to use comparison output function 12
8. Setting mode 13
8-1. Types of Setting mode & Set-up 13
8-2. Function mode (Function Set-up) 14
8-3. Digital calibration (Calibration by sensor output value) 21
8-4. Actual load calibration (Real-weight Calibration) 22
8-5. SPAN constant calibration (Calibration by Shunt CAL value) 23
8-6. Data Back-up & Restore 24
8-7. Lock Set-up 25
9. Product Inspection 26
10. OPTION (OP-02, OP-03)
Option 02 (RS232C) 27
Option 03 (RS485) 29

2
1. Special Features
We are grateful to you for picking up our product. If you experience problems while
using it, you may refer to the user manual or contact the Technical Support Department
of the company.
This product is an indicator which displays Torque and RPM in digital mode and it
has the following features.
1-1. High-speed Conversion Method
24bit High-speed A/D converter that can detect the sensor’s input signal at a rate of
1000times per second is used. For analog output, 16bit D/A converter with 1000times per
second is used so that there is almost no devuatuib betweeb display value and output value.
1-2. Calibration Method
2 types of calibration systems are adopted. Calibration by actual load (standard weight) and
calibration by rated output of sensor (mV/V).
1-3. Hold and Peak Hold
You can choose Peak hold or Sample hold as per your need.
1-4. Comparison Output
You can do set-up the maxium and lower limit value by the keys on the front panel and output
contact point signal of the rear panel.
1-5. Data Back-up
All the set-up values will be memorized on the flash memory so the inputted data can be
saved and no need to do the re-setting even in case of the interruption of the electric power or
power disconnection.
1-6. Watch dog
This function is for automatic reset in case the system is stopped due to the external factors
such as noise.

3
1-7. Option
You can use RS232C and RS485 output as the optional specification.
1-8. Power
AC 85 ~ 265V 50/60Hz Free voltage

4
2. Attention
For the efficient and safe use, please carefully read and be fully aware of the following
details before using this product.
It is strictly forbidden to use this product for any other purpose of use or to attempt to
make any alteration on this product.
2-1. Attention for installation
Please keep it out of wet places.
Do not set it up near vibration & impulse, high temperature and humidity.
Keep it out of the direct rays of the sun. Set it up where there is less dust, and Keep
it out of direct air including salt and ion.
Do not use when there is inflammable gas or heavy machinery, and smog.
Ground earth-terminal ( ).
Make wire separately from power system wiring and noise wiring.
Make sure the use of 4 line sealed cable as a sensor cable. Too long cable leads to
measurement error due to wiring resistance (around 10meters).
2-2. Attention for use
During calibration, Do not input free-load state and real-weight load until it becomes
stable. Pressing Enter Key in unstable condition leads to calibration error.
Do not press any Key in use at one's discretion. Please refer to 7. Setting-up mode for
the function and method of Key.

5
3. 사양
Available sensor
Strain Gauge sensor (Bridge 120, 350, 700)
Upper limit Display : -19999 ~ +99999
A/D Converter: 24bit, 500 times/sec
D/A Converter : 16bit
Temperature Characteristic (Amp Characteristic)
Zero : 0.5μV/℃
Span : 50ppm/℃
Front Panel
Measured value display: 7 segment 5 Digit, Character Height 14mm
Status display : Red LED 6 pcs
Key Switch : 4 pcs
Output
Comparison output: Upper limit(HI), Lower limit(LO)
Contact Capacity - AC 125V 0.5A/ DC 24V 1A
Analog Output : DC 0~10V (4~20mA available by internal operation)
• Temperature Range: -10℃∼60℃, less than 80% RH (no dewing.)
External dimension : 97 ×97 ×141 mm (W ×H ×D)
Panel cutting size : 92.5 ×92.5 mm
Temperature Range : About 650g
Power Supply : AC 85~265V 50/60Hz free voltage
Option : RS232C (OP-02), RS485 (OP-03)

6
4. Front Panel
①HI, LO Indication LED
: Hi or LO LED will be lighted when measured value exceeds the setting value.
②HOLD indication LED
: This LED will be lighted when the measured value is on Hold.
③Measured value Indication
: It indicates the measured value and each setting value..
④
Measruing Mode : Once this Key is pressed for more than 1 second, the current
measuring value will be Zero (0) and the Analog output will be 0V (4mA) as well.
SET-UP Mode : Once this is pressed on the Function Set-up Mode, you will
return to the measuring mode.
⑤
Measuring Mode : Once this Key is pressed, the lower limit setting value will be
displayed and this value also can be changed.
SET-UP Mode : The location of row for the number flickering can be moved.
⑥
Measuring Mode : Once this Key is pressed, the upper limit setting value will be
displayed and the value can be changed.
SET-UP Mode : The flickering number will be increased by 1 and 1.

7
⑦
Measuring Mode : Once this Key is pressed, HOLD will be selected and once this
key is pressed again, HOLD will be cancelled.
SET-UP Mode : Save each set-up value.

8
5. Rear Panel
Please check the location of terminal and its use. While the projected button is being
pressed, please insert the cable into the lower hole completely. As soon as you release
the button, the connection will be completed. At the point, please slightly pull the calbe
and check whether the cable is come off or not.
(The most suitable calbe is Φ0.5~1. Linking cable must be brazing or used with 1
terminal)
①AC_L, N : Main Power Supply Wiring Terminal
②: Ground Terminal (as an independent ground connection.)
③E1+ : CH1 Sensor Supply Voltage + Connection Terminal(FED)
④E1- : CH1 Sensor Supply Voltage - Connection Terminal(White)
⑤S1+ : CH1 Sensor Ouput Signal + Connection Terminal(Black)
⑥S1- : CH1 Sensor Output Signal –Connection Terminal(Green)
⑦GND : SHIELD connection terminal of CH1 sensor(Tare)
⑧E2+ : CH2 Sensor Supply Voltage + Connection Terminal(Red)
⑨E2- : CH2 Sensor Supply Voltage - Connection Terminal(White)
⑩S2+ : CH2 Sensor Ouput Signal + Connection Terminal(Black)
⑪S2- : CH2 Sensor Output Signal –Connection Terminal(Green)
⑫GND : SHIELD connection terminal of CH2 sensor(Tare)

9
⑬OUT1, 2 : Analog (DC 0 ~10V/DC 4 ~ 20mA) Output Terminal
⑭GND : External Input Common Terminal
⑮IN1 : External HOLD Input Terminal (CH1)
⑯IN2 : External ZERO Input Terminal (CH1)
○
17 IN3 : External HOLD Input Terminal (CH2)
○
18 IN4 : External ZERO Input Terminal (CH2)
○
19 COM : RELAY Output Common Terminal
○
20 RY1 : CH1 Lower Limit Output Terminal
○
21 RY2 : CH1 Upper Limit Output Terminal
○
22 RY3 : CH2 Lower Limit Output Terminal
○
23 RY4 : CH2 Upper Limit Output Terminal
◀Attention▶
1. When wiring, please pull out the power plug. (Main power supply)
2. Please use the thick calbe for ground terminal ( ) so that you can avoid any trouble
due to the impulse voltage or surge. If possible, please keep it as an independent
ground.
(This must be used in the area with heavy noise. If the ground is made with other
device, it will be affected by noise.)
3. Please check the function of the terminal before you do the wiring so that you can
prevent any malfunction in advance.
4. Manufacture will be relieved of its responsibility for any damge or injury due to the
disassembly and alteration made without any agreement. Also, no A/S will be made.