5
The absence of an arrow does not necessarily signify
stability. It may mean that the particular reading is undergoing
little to no change. Therefore, slow changes could occur without
being indicated by a trend arrow.
To exit Live Data view, select iew Log Data from the menu.
NOTE:
Smart TrendTM Indicators - identify rising, falling or stable
readings. The probe compares each measurement with recent
values to provide the current direction of change.
Logging Mode
In Logging Mode, the PosiTector DPM L probe will
automatically record datasets into memory at user selected time
intervals. To enable, select Start Logging from the PosiTector
App or PosiTector Advanced gage menu, then enter the desired
logging measurement interval (between 1 minute and 8 hours)
and select OK. All previously logged datasets are deleted from
the probe’s memory.
The PosiTector DPM L immediately records the first dataset and
records subsequent datasets at the user-specified interval. The
PosiTector DPM L does not need to remain connected to the
PosiTector App or PosiTector Advanced gage body to continue
logging datasets. A maximum of 10,000 datasets (Ta, Ts, Td, Ts-
Td, RH, Tw) can be stored in PosiTector DPM L
memory. All datasets are date and time stamped.
The Logging Mode icon will appear on the
display alongside the set interval. Measurements
on the display are only updated when a
dataset is stored. To view current
environmental
conditions, select
iew Live Data from
the menu. In View
Live Data mode, the
PosiTector DPM L
probe will continue
recording. To return
to Log view, select
iew Log Data.