Simplicity®Cart
2 Installation,OperationandCareManual
Section I Introduction
This manual contains installation, operation and care
instructions and user service information for the
Simplicity®Cart Delivery Unit.
The Simplicity Cart is available in two standard con-
gurations: (1) Duo (doctor's and assistant's) instru-
mentation and (2) doctor's only instrumentation.
The Simplicity Cart Delivery Unit is intended to be
used by trained professional dental care personnel
as an interface device to connect the dental oper-
atory hand instruments to the appropriate supply
utility such as air, water, vacuum, drain and electrical.
It functions as a system management device that
provides a method of operating the hand instruments
from a single control input device.
The Simplicity cart, situated on wheels, positions
the handpieces for the optimum presentation to the
operator.
The Unit is designed to provide trouble-free service
when installed, operated and maintained according to
the procedures set forth in this manual.
To ensure correct installation, carefully read all
the procedure instructions contained in this
manual paying close attention to all notices,
notes, cautions and warnings.
Before proceeding, please review the photo below to
become familiar with the basic components of the Unit.
After the Simplicity Cart is installed, please review
the operation procedures and care guidelines with the
doctor's staff. Then leave this manual in the doc-
tor's ofce for future reference.
Utility Service
Compartment
Clean
Water
System
Doctor's
Controls
BASIC COMPONENTS
Work
Surface
Specications
Power Supply - 100-240V, 50-60 Hz., as applicable
Air Pressure - 551.6 pKa (80 PSI)
(at regulator in USC)
Water Pressure - 275.8 pKa (40 PSI)
(at regulator in USC)
Clean Water System -
Reservoir Capacity - 1.5 L
CMU Shipping Weight
Cart with transformer - 61.2 kg (135 lbs.)
Cart without transformer - 47.6 kg (105 lbs.)
Maximum Load for Cart Top - 0.453 kg (10 lbs.)
Recommended Environmental Conditions
– Environmental conditions for transport and storage:
- Relative humidity range within 0% to 95%
- Transport/storage temp. within -29ºC (-20ºF) to
74ºC (165ºF)
– Environmental conditions for operation:
- Conditioned Air
-
Atmospheric pressure range within 500 to 1050 kPa
- Operation temp. within 15ºC (59ºF) to 27ºC (80ºF)
Air and Water Inlet Temperature Range:
5ºC (41ºF) to 27ºC (80ºF)
Medical-GeneralMedicalEquipment
Certiedastoelectricalshock,re
andmechanicalhazardsonlyin
accordancewithUL60601-1,
CAN/CSA-C22.2No.601.1,
CAN/CSA-C22.2No.60601-1(2008)&
ANSI/AAMIES60601-1(2005).
Classications
0459
– Type of protection against electric shock: Class 1
Equipment
– Degree of protection against electric shock: Type B
Applied Parts. The handpiece is considered an applied part.
– Degree of protection against the ingress of water:
Ordinary
–Equipment not suitable for use in the presence of a
ammable anesthetic mixture with air or with oxygen
or nitrous oxide.
– Mode of operation: Continuous
These specications apply to model SP-A (doctor's
only instrumentation) and model SP-B (doctor's and
assistant's instrumentation).
Temperatures not within the air and water inlet
termperature range could cause damage to the unit.
CAUTION