DentalEZ EverLight Manuale utente

User Manual
EverLight®
Post Mount

1
User Manual
Table of Contents
Section I Introduction
Specications .......................................................................................................................... 2
Classications......................................................................................................................2-3
LightHeadRangeofMotion............................................................................................... 3
Dimensions ............................................................................................................................. 4
Packaging ................................................................................................................................. 4
Section II Installation
PostMountwithServiceConsole..................................................................................5-6
PostMountforMagellanCMU .......................................................................................... 6
Section III Operation
LightSwitch............................................................................................................................. 7
TouchPads .............................................................................................................................. 7
PositioningLight..................................................................................................................... 8
EliminatingAdjustableArmDrift........................................................................................ 9
Section IV Care
Cleaning&Disinfecting ...................................................................................................... 10
Section V User Service Information
Troubleshooting................................................................................................................... 11
ServiceInstruction .............................................................................................................. 11
DisposalofEquipment........................................................................................................ 11
Section VI Parts List/Diagrams
LightArm............................................................................................................................... 12
LightHead............................................................................................................................. 12
Limited Warranty ........................................................................................................ 13
Wiring Schematic ........................................................................................................ 14
EMC Information .................................................................................................... 15-18

User Manual
2
EverLight® (Post Mount)
Section I Introduction
Classications
Medical-GeneralMedicalEquipment
Certiedastoelectricalshock,re
andmechanicalhazardsonlyinaccor-
dancewithUL60601-1,
CAN/CSA-C22.2No.601.1,
CAN/CSA-C22.2No.60601-1(2008)&
ANSI/AAMIES60601-1(2005).
–Type of protection against electric shock: Class 1
Equipment
–Degree of protection against the ingress of water:
Ordinary
–Equipment not suitable for use in the presence of
a ammable anesthetic mixture with air or with
oxygen or nitrous oxide.
–Mode of operation: Continuous
–Environmental conditions recommended ranges
for:
Transport and Storage:
- Temperature range of - 10º C to + 70º C
- Relative humidity range of 10% to 100%
- Atmospheric pressure range of 500 to 1060 kPa
Normal Use:
- Temperature range of + 10º C to + 40º C
- Relative humidity range of 30% to 75%
- Atmospheric pressure range of 700 to 1060 kPa
Power Supply: Primary: 100-240V AC, 50/60 Hz.
Secondary: 12V DC
Operating Amperage: .4 max. - .16 min. Amp
Focal Length: 68.6 cm (27")
Color Temperature: High: 5800° Kelvin
(Color Matching)
Medium: 4500° Kelvin
(Operative Setting)
Low: 3200° Kelvin
(Composite Setting)
Light Intensity: High: 2790 Footcandles
(30,000 Lux)
Medium: 2325 Footcandles
(25,000 Lux)
Low: 1395 Footcandles
(15,000 Lux)
Light Pattern: 16.5 cm (6-1/2") x 8.9 cm (3-1/2")
Unit Weight: Post Mount: 12 kg (27 lbs.)
TheauthorizedEuropeanrepresentativeis:
DentalEZ(GB)Ltd.
ClevelandWay
HemelHempstead,Hertfordshire
HP27DY,England
Phone:(01442)269301
Attn:Mr.AndyKing
This manual contains installation, operation and care
instructions and user service information for the
EverLight® LED operatory light.
The EverLight operatory light system is intended to
be used by trained professional dental care personnel
to provide shadow free illumination to the oral cavity
for dental practices. The operatory light arm system
allows optimal light positioning, which for a reclined
patient, is normally from the sagittal plane directly
above or slightly behind the patient's oral cavity. The
operator will typically be positioned behind or to one
side of the patient. The arm assembly is counterbal-
anced so that minimum force is required to move
the light head assembly. The operatory light is not
designed to handle any external loads, such as hanging
foreign objects from the light system.
The EverLight operatory light is designed to provide
trouble-free service when installed, operated and
maintained according to the procedures set forth in
this manual.
To ensure correct installation, carefully read all
the instructions contained in this manual
paying close attention to all notes, notices,
cautions and warnings.
After the light is installed, please review the operation
procedures and care guidelines with the doctor's staff.
Then leave this manual in the doctor's ofce for
future reference.
53HN
Specications

3
In accordance with Part 15 of FCC rules, this
equipment was tested and complies with Class
A digital device limits. These limits are designed
to give equipment reasonable protection against
detrimental interference when operated in a
commercial environment.
Medical electrical equipment needs special
precautions regarding electromagnetic
compatibility (EMC) and needs to be installed
according to EMC information. (See EMC
Information Pgs. 15 through 18.)
Mobile radio frequency (RF) communications
equipment can effect medical electrical
equipment.
Section I Introduction
=Caution
=Warning
=Biohazard
=WarningDangerousVoltage
=AlternatingCurrent
=DirectCurrent
=TypeBAppliedPart
=ProtectiveEarth(ground)
=GeneralMandatoryAction
=RefertoManual(followinstructions)
= EuropeanCertication,ifapplicable.
Seenameplate.
=SerialNumber
=ManufactureDate
=Manufacturer
SN
Light Head Range of Motion
NOTICE
145º
145º
180º
150º
Explanation of Symbols and Signs:

User Manual
4
EverLight® (Post Mount)
Section I Introduction
Dimensions
The EverLight components are packaged and shipped
according to the packing list below. Please verify the
carton contents against the packing list.
Post-Mount Packing List
•Light Arm Assembly
•Light Post
• Flanged Bronze Bushing (optional)
• Power Supply Box
•Hardware Package
•Damage Instructions
•Manual
•Warranty Card
NOTE:Be aware that some cartons are shipped as
llers and may be empty (even if optional components
are ordered).
Carefully unpack the light. Inspect all of the items
received to assure there has been no shipping damage.
If damage is discovered, contact the carrier
immediately. Then report and le the appropri-
ate claims with the carrier within 24 hours after
delivery. (See Limited Warranty, Page 13.)
Light Arm
Assembly
Light Post
Post-Mount Arm Assembly Dimensions
15"
17"
7.25"
2"
24" 31.75"
1.701"
Packaging
For any questions about an order, please
contact a DentalEZ Equipment customer service
representative at 1-866-DTE-INFO.
NOTICE
All parts supplied are vital for proper installation.
Do not discard any hardware or components until
installation is complete.
NOTICE

5
Section II Installation
Isolating the light from the supply mains is accomplished
by unplugging the light from its power supply.
NOTICE
Installation by an authorized DentalEZ dealer service
technician is recommended.
NOTICE
Do not modify this equipment without authorization
from the manufacturer.
WARNING
Light Post
Set Screw
Spindle Slot
Light Arm
Assembly
NOTE: Refer to your chair mounted unit manual for
installation of the light pole.
NOTE: The post-mount arm assembly dimen-
sions are illustrated at the top of Page 4.
NOTE: The side covers of the post mount service
console must be off. Refer to the unit manual for
removal.
NOTE: If the chair's seat upholstery was installed,
remove it according to the respective chair's manual.
Light Arm Assembly
1. Place the chair base in its lowest position.
2. Disconnect the chair power.
3. Back out the set screw from the light post.
4. Apply a light coat of lubricant to the spindle.
7. Insert the light arm assembly into the bushing
until it is fully seated in the light post.
8. Using a 5/64 Allen wrench, turn the set screw
in the light post until it passes through the
light arm assembly spindle slot and engages
the light arm assembly. Then loosen the set
screw two revolutions for clearance.
Power Supply Box
1. Inside the post-mounted service console, plug the
operatory light cord into the power supply cord.
Power
Supply
Supplied
Power Cord
5. Feed the low-voltage operatory light cord
through the center of the light post, arm
pivot and console post into the post-mounted
service console.
6. Pull the light cord all the way through the
post-mounted service console until the cord is
not too tight or loose going through the posts.
To avoid the risk of electrical shock, this
equipment must only be connected to a
supply mains with protective earth.
Before proceeding with electrical installation,
all wiring must be in accordance with NEC
and local electrical codes.
WARNING
WARNING
Rating of main circuit breakers should be
20 Amp maximum.
CAUTION
Post Mount with Service Console
2. Plug the supplied power cord into the power supply.

User Manual
6
EverLight® (Post Mount)
Section II Installation
Post Mount for Magellan CMU
3. Route the supplied power cord through the
bottom of the console and into the front of
the chair. (P-clamps provided in bag of supplies.)
5. Roll up any excess wires and neatly place
the power supply and excess wires inside the
console making sure no wires interfere with
closure of the console covers.
6. Replace the side covers on the console.
7. Reconnect the chair power and test the light
switch as described in Section III Operation.
Light Arm Assembly
1. Place the chair in its lowest position.
2. Disconnect the chair power.
3. Apply a light coat of lubricant to the spindle.
4. Slide the bushing (lipped end rst) over the light
cord and onto the spindle.
5. Feed the light cord through the center of the
light pole, pulling the cord out of the bottom
of the pole.
Power Supply Box
1. Plug the supplied power cord into the power
supply. (See photo below.)
2. Plug the light harness from the light arm into
the power supply.
3. Route the supplied power cord harness
through the front of the seat casting and plug
it into the chair harness. (See second photo on this
page.)
4. Roll up any excess wire and neatly place them
inside the seat casting making sure no wires
will interfere with moving parts. (Wire ties
provided in bag of supplies.)
5. Using the supplied Velcro, adhere the power
supply to the magellan manifold cover.
4. Plug the supplied power cord into the chair
harness located inside the chair.
P-Clamp
Console
Supplied
Power Cord
Chair Harness
Supplied
Power Cord
NOTE: Refer to the Magellan Unit Installation
Manual for installation of the light pole.
NOTE: If the chair's seat upholstery was installed,
remove it according to the respective chair's manual.
Magellan Manifold
Cover
Power
Supply
6. Reconnect the chair power and test the light
switch as described in Section III Operation.
NOTE: Replace the seat upholstery if it was removed.
Power
Cord

7
Section III Operation
Designed with three procedural and three intensity
settings EverLight offers a total of nine settings to
assure the correct light in every procedure.
Procedural Settings
•A- Sun - Color Matching - 5800° Kelvin
•B - Sun/Bulb - Operative Setting - 4500° Kelvin
•C - Bulb - Composite Setting - 3200° Kelvin
Low
(15,000 Lux)
Sun/Bulb - 4500°Kelvin
(Operative Setting)
Bulb - 3200°Kelvin
(Composite Setting)
Sun - 5800°Kelvin
(Color Matching)
Procedural
Settings
Intensity
Settings
Intensity Settings
•1 - Low - 15,000 Lux
•2 - Medium - 25,000 Lux
•3 - High - 30,000 Lux
Medium
(25,000 Lux)
High
(30,000 Lux)
The light ON/OFF toggle switch is located on the
elbow section of the light arm assembly.
Toggle
Switch
•Flip the toggle switch to the UP to turn the light ON.
•Flip the toggle switch DOWN to shut the light OFF.
ON
OFF
The EverLight dental light is designed to provide illu-
mination of the oral cavity for use by dental profes-
sionals in the dental operatory.
Light Switch
Touchpads
é
é

User Manual
8
EverLight® (Post Mount)
Section III Operation
Using the handles on either side of the LED head,
position the light to illuminate the oral cavity. (See
Light Head Range Of Motion, Pg. 3.)
Handles
Positioning Light
To take full advantage of the EverLight's intensity
and color correction, position the light between 20"
and 45" away from the oral cavity.
NOTICE

9
Section III Operation
The adjustable arm is factory-adjusted to counter-
balance the weight of the light head. If it becomes
necessary to adjust the mechanism because of added
weight, such as a mirror, perform the following steps:
Tools Required
•5/64 Allen wrench
•9/64 Allen wrench
•Phillips-head screwdriver
•Large at-head (slotted) screwdriver
1. Using a Phillips-head screwdriver, loosen the
#6 screw under the cap at the end of the
adjustable arm nearest the light head.
2. Using a 5/64 Allen wrench, remove the four
Allen-head screws from the underside of the
adjustable arm (two on each end). Then slide off
the top cover along with the end cap.
3. Insert a 9/64 Allen wrench into the two small
holes in the side of the adjustable arm and
completely loosen the friction block.
4. Insert a large at-head screwdriver into the
large slot in the side of the adjustable arm and
tighten the slotted nut.
5. Adjust the friction block to provide one to
two pounds maximum of break-away force.
6. Replace the adjustable arm cover and secure
using the four 5/64 Allen-head screws
previously removed.
7. Replace the end cap and tighten the #6 screw
using a Phillips-head screwdriver.
Adjustable Arm
Top Cover
5/64 Allen-head Screws (2)
End Cap
5/64 Allen-head
Screws (2)
9/64 Allen
Wrench
Tw o
Small
Holes
Friction Block
Adjustable
Arm
Adjustable Arm
Large
Slot
Slotted
Nut
Large Flat-head
Screwdriver
Set
Screw
End Cap
Adjustable
Arm
Eliminating Adjustable Arm Drift