Denver Instrument TP Series Manuale utente

TP Series
Electronic Precision Balances/ Elektronische
Präzisionswaagen / Balances électroniques
de précision / Balanzas electrónicas de precisión
Operating Instructions / Betriebsanleitung /
Mode d’emploi / Instrucciones de funcionamiento

2
English – page 3
Deutsch – Seite 31
Français – page 59
Español – página 87

Contents
Warnings and Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Basic Weighing Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Calibration/Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Application Programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Net-Total Formulation/Second Tare Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Counting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Weighing in Percent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Weigh Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Toggling between Weight Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Configuring the Balance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Setting the Parameters (Menu Codes) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Balance Operating Menu (Overview) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
ISO/GLP-compliant Printout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Data Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Care and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Instructions for Recycling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
CMarking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3
Contents

4
Safety
§To prevent damage to the equip-
ment, please read these operating
instructions carefully before using
your balance.
!Do not use this balance in a haz-
ardous area/location.
!Make absolutely sure to unplug
the balance from AC power before
you connect or disconnect a
peripheral device.
!Exposure to excessive electromag-
netic disturbance can cause the
readout value to change. Once the
disturbance has ceased, the instru-
ment can be used again in accor-
dance with its intended use.
Setting up the Balance
!Warning when using pre-wired
RS-232 connecting cables: The pin
assignments in RS-232 cables pur-
chased from other manufacturers
may be incompatible with DENVER
balances. Be sure to check the pin
assignment against the chart on
page 21 before connecting the
cable, and disconnect any lines
that do not match.
– Connect only DENVER accessories
and options, as these are optimally
designed for use with your DENVER
balance. Do not try to solve any
problems on your own. The opera-
tor shall be responsible for any
modifications to DENVER equip-
ment and for any connections of
cables or equipment not supplied
by DENVER and must check and, if
necessary, correct these modifica-
tions and connections. On request,
DENVER will provide information
on the minimum operating specifi-
cations (in accordance with the
standards for defined immunity to
interference).
– Do not open the balance housing.
If the seal is broken, this will result
in forfeiture of all claims under the
manufacturer’s warranty.
Warnings and Safety Precautions

5
Getting Started
Storage and Shipping Conditions
– Do not expose the balance to
extreme temperatures, blows,
shocks, vibration or moisture.
Unpacking the Balance
§After unpacking the balance,
check it immediately for any visible
damage
$If you see any sign of damage,
proceed as directed in the chapter
entitled “Care and Maintenance,”
under the section on “Safety
Inspection”
$Save the box and all parts of the
packaging until you have success-
fully installed your balance in case
you need to return it. Before pack-
ing your balance, unplug all con-
nected cables to prevent damage.
Equipment Supplied
– Balance
– Weighing pan
– Pan support (only on models with
a round weighing pan)
– AC adapter, plug type
Additionally supplied with TP-214
models:
– Shield ring
– Shield plate
– Dust cover
Additionally supplied with TP-303,
TP-303 I models:
– Glass draft shield with cover
Installation Instructions
When choosing a location to set
up your balance, observe the
following:
– Avoid placing the balance in close
proximity to a heater or otherwise
exposing the balance to heat or
direct sunlight
– Protect the balance from drafts
that come from open windows or
doors
– Avoid exposing the balance to
extreme vibrations during
weighing
– Do not expose the balance
to extreme moisture over long
periods
Conditioning the Balance
Moisture in the air can condense on
the surfaces of a cold balance when-
ever it is brought into a substantially
warmer place. If you transfer the bal-
ance to a warmer area, make sure to
condition it for about 2 hours at room
temperature, leaving it unplugged
from AC power.

6
Setting up the Balance
Balances with an Analytical Draft Shield
§Place the components listed below inside
the chamber in the order given:
– Shield plate
– Shield ring
– Pan support
– Weighing pan
Balances with a Glass Draft Shield
§Place the components listed below inside
the chamber in the order given:
– Draft shield base – place it on the balance so
that the edge for fitting the glass draft shield
faces upwards and turn it until it is firmly in
place
– Pan support
– Weighing pan
– Glass draft shield
– Draft shield cover – place it on the balance
so that the edge faces downwards
Balances with a Round Weighing Pan
§Place the components listed below inside
the chamber in the order given:
– Pan support
– Weighing pan
Balances with a Rectangular Weighing Pan
§Place the weighing pan on the balance

7
Connecting the Balance to AC Power/
Safety Precautions
Use only original Denver AC adapters
§Insert the right-angle plug into the jack
§The AC adapter rated to Class 2 can
be plugged into any wall outlet without
requiring any additional safety precautions
The ground is connected to the balance
housing, which can be additionally grounded
for operation.
Using a Non-Rechargeable/Rechargeable Battery:
for models TP-6101, TP-3101, TP-12
$A non-rechargeable or rechargeable battery
is not included with the equipment supplied
!Use only a commercially available non-
rechargeable or rechargeable 9-volt battery
!When using a rechargeable battery, always
use an external charger to recharge the bat-
tery
§Lay the balance/scale on its side
§Open the battery compartment:
Lift the compartment cover
§Insert the 9-volt (rechargeable) battery in
the
compartment
$Make sure to connect the positive and nega-
tive poles correctly
!All used batteries are classified as waste
that requires special handling (not “house-
hold” waste). Dispose of rechargeable bat-
teries in accordance with the applicable spe-
cial waste disposal regulations.
§Close the battery compartment:
Press down on the cover until it clicks into
place

8
Leveling the Balance
Level the balance any time you set it up in a
new location. Use only the 2 front feet of the
balance for leveling.
TP-3002, TP-1502, TP-6101, TP-3101, TP-12:
§Turn the 2 rear feet until they are in position
§Turn the 2 front feet as shown here in the
illustration until the air bubble is centered in
the level indicator
> In most cases, this will require several
adjustment steps
Anti-theft Locking Device
To protect against theft, use the mounting lug
on the rear panel of the balance.
§Secure the balance at the place of installa-
tion, for example with a chain or a lock

9
Operation
Basic Weighing Function
Preparation
§Turn on the balance:
Press [ON/OFF]
§To change configurations: See
the chapter entitled “Configuring
the Balance”
$To tare the balance: Press [TARE]
Additional Functions:
§To turn off the balance:
Press [ON/OFF]
Warmup Time
$To ensure accurate results, the bal-
ance must warm up for 60 minutes
before operation. Only after this
time will the balance have reached
the required operating temperature.
Example
Basic weighing
Step Key (or instruction) Display/Printout
1. Turn on the balance [ON/OFF]
Self-test is performed
2. Place container on balance +52.0 g
(here: 52 g)
3. Tare the balance [TARE] +0.0 g
4. Place sample in container +150.2 g
on balance (here: 150.2 g)

10
Calibration/Adjustment
Available Features
Calibration/adjustment can only be performed when
– there is no load on the balance,
– the balance is tared,
– the internal signal is stable.
If these conditions are not met, an error message is displayed.
Otherwise, the weight required for calibration/adjustment is displayed.
You can use any of the following weight units to calibrate/adjust:
g, kg*, lb (menu code 1. 4. x)
You can block calibration/adjustment of the balance:
– Select menu code 1. 5. 3
External calibration/adjustment
Step Key (or instruction) Display/Printout
1. Tare the balance [TARE] 0.0 g
2. Begin calibration [CAL] +1000.0
Calibration weight is displayed
without weight unit (here: 1000 g)
3. Place the indicated calibration 1000.0
weight on the balance
After calibration, the calibration +1000.0 g
weight is displayed with wt. unit
4. Remove the calibration weight 0.0 g
* = not on models with a readability of 0.1 mg
Questo manuale è adatto per i seguenti modelli
8
Indice
Lingue:
Altri manuali Denver Instrument Scala

Denver Instrument
Denver Instrument XE-100A Manuale utente

Denver Instrument
Denver Instrument TB-124 Manuale utente

Denver Instrument
Denver Instrument DA Series Manuale utente

Denver Instrument
Denver Instrument TB Series Manuale utente

Denver Instrument
Denver Instrument DL series Manuale utente

Denver Instrument
Denver Instrument Summit Manuale utente

Denver Instrument
Denver Instrument MAXX Manuale utente

Denver Instrument
Denver Instrument XP Series Manuale utente

Denver Instrument
Denver Instrument P-114 Manuale utente

Denver Instrument
Denver Instrument ACA-100 Manuale utente