Dickson KT8P Manual V6 52418
NEED TO KNOW
NEED TO KNOWNEED TO KNOW
NEED TO KNOW
The pen movement for the KT8P0 is slower than on the AC powered models in order to onserve
battery power. In 24-hour mode the pen(s) will move every 1 minute and 30 se onds. In 7-day
mode the pen(s) will move every 8 minutes and 30 se onds.
The re order takes temperature readings every 50 se onds. An average of the reading taken
between pen(s) movements is what is drawn on the hart.
TROUBLE SHOOTING
TROUBLE SHOOTINGTROUBLE SHOOTING
TROUBLE SHOOTING
Why isn't the hart keeping time or running slow?
Why isn't the hart keeping time or running slow?Why isn't the hart keeping time or running slow?
Why isn't the hart keeping time or running slow?
•The hart may be hung up or restri ted, possibly aused by a rip on the outer edge of the
hart or the hart hub, or the hart may be aught between the arm and pen(s) arm
platform.
•In orre t hart installed for the sele ted hart speed.
Why did the hart stop tu
Why did the hart stop tuWhy did the hart stop tu
Why did the hart stop turning?
rning?rning?
rning?
•Chart hung up or restri ted, (ripped hart)
•Unit may be "lo ked up", this an be onfirmed by pressing any of the buttons on the
keypad, if the unit is lo ked up there will be no response to button presses and the Chart
Re order may appear to be working, but the readings won't hange, also the hart will
not rotate. Remove power and battery, and then re-power.
Why don't the display and hart mat h?
Why don't the display and hart mat h?Why don't the display and hart mat h?
Why don't the display and hart mat h?
•Dip swit hes set for a spe ifi range, but using hart for another range, or vi e versa.
•Pen(s) not inserted on pen arm(s) all the way
•To adjust pen(s) to mat h hart see “Pen Home Adjust” in the “Need to Know” se tion of
this manual.
Why does the display read PROB?
Why does the display read PROB?Why does the display read PROB?
Why does the display read PROB?
•Che k the sensor pod and KTC probe(s). Make sure it is properly plugged into the unit.
•Is the unit lo ked up? This an be onfirmed by pressing any of the buttons on the
keypad, if the unit is lo ked up there will be no response to button presses and the unit
may appear to be working, but the readings won't hange, also the hart will not rotate.
Remove power and battery, and then re-power.
Why does alibration seem to be off?
Why does alibration seem to be off?Why does alibration seem to be off?
Why does alibration seem to be off?
•What is the toleran e of the unit it's being ompared to?
•It's ok if the unit is within the sum of the two toleran es.
•Has an outside alibration house attempted alibration? It may not have been adjusted
properly.
•Walk through alibration adjustment, found in the manual
Why won't the battery ba k
Why won't the battery ba kWhy won't the battery ba k
Why won't the battery ba k-
--
-up work?
up work?up work?
up work?
•Are there good batteries in the Chart Re order?
•Keep in mind that battery ba kup varies signifi antly depending on temperature, pen
movement and hart rotation speed.
•If there was a momentary power outage (Brown out), the unit may not have had enough
time to re ognize this and swit h to battery mode. (The Chart Re order may lo k up or
shut off in this situation) This situation may also o ur if the unit is plugged into an outlet
that shares the ir uit with other ma hinery that has phase motors or ompressors that
y le periodi ally. As these other ma hines y le, they momentarily have a high urrent
draw, therefore drawing from the unit.