Precautions For Use
3/ 63
2Precautions for Use
In this manual, precautions are classified into warning and caution as follows:
WARNING:
A Warning alerts you to a potential serious outcome, adverse event or safety hazard.
Failure to observe a warning may result in serious injury.
CAUTION:
A Caution alerts you to where special care is necessary for the safe and effective use of
the product. Failure to observe a caution may affect normal use of the product.
WARNING
⚫The infusion pump must be operated by qualified clinical professionals.
⚫The infusion pump cannot be used for blood transfusion.
⚫The pump does not have an upstream occlusion alarm, clamps must be below the
pump at all times.
⚫Prior to use, please check the status of the pump, power cord and other related
accessories to ensure the device can be used normally and safely.
⚫Pay extra attention to kinks of the infusion line when it is used for low-infusion. The
smaller the set infusion rate becomes, the longer it takes from the occurrence of
occlusion to its detection, which may suspend the infusion for a long time.
⚫To avoid the risk of fire or explosion, do not use the infusion pump in a flammable or
oxygen rich environment.
⚫The altitude difference between the pump and heart position of the patient should
not be larger than 100cm (3.3ft). Smaller difference of the altitude will increase the
accuracy of the pressure sensor’s result.
⚫In the event of tube twisting, filter condensation or intubation occlusion during
infusion, the internal pressure of the infusion tube will increase. Once the causes for
occlusion are removed, too much infusion liquid may be infused into the patient.
Therefore, proper actions should be taken. For example, clamp the infusion tube
before removing the occlusion causes.
⚫To guarantee the infusion safety and alarming function, it is recommended to use the
IV sets specified by the manufacturer only.
⚫Only the IV set, tube, infusion needle and other medical parts complying with the