D3998-US Page 4 Rev A 4/10/2015
Moisture Tracker Adjustment Procedure
The Dry Matter results (DM) of a MT (MT) can be aligned with a known reference DM (from Lab, Dehydrator,
Koster, etc.). Individual feedstuff DM readings may not all align the same due to natural variations in feeds,
but should be within acceptable ranges (+/- 2%). If you have questions please contact:
Dick Bonner - Technical Sales Representative
Office: 920-568-6210
Mobile: 608-359-5930
W5527 Hwy 106, Fort Atkinson, WI 53538
Toll Free: 1-800-225-7695, Ext 6210 (USA only)
1. Gather samples of 2 –3 common feedstuffs (corn silage, alfalfa haylage, high moisture
ground shell corn, etc.) that have Calibrations available for the MT. Gather enough material
that you can send some to the lab and save back enough to do future scans with the MT.
2. Divide the total sample of each feedstuff into one portion for testing per your preferred
reference method, a second portion for original measurement with MT, and a third portion
(packaged in an air-tight container (zip-lock bag) and stored in a refrigerator or freezer) for
later MT readings to verify the results after adjustment.
3. Take 2-3 DM readings on each of the feedstuffs using the MT.
4. Compare DM readings with reference DM reading. Compare the reference DM results to
your original MT DM readings.
If the MT DM is within 1-2% of the reference DM: do nothing.
If the MT DM is more than 2% higher than the reference DM: decrease the Calibration
Code.
If the MT DM is more than 2% lower than the reference DM: increase the Calibration
Code.
5. Adjust the Calibration Code:
On the MT, Select Main Menu > Calibrate Sensor > Enter New Calibration Code (3)
Change the Calibration Code by 5. (Ex: changing a code of 545493 to 545488 –a
reduction of 5 counts.) Use the right or left arrow buttons to get to the desired digit
and the up or down arrows to scroll the digit to the desired number. Press the “Enter”
button to save the adjusted code.
6. Recalibrate the MT.Hold the probe on the Calibration Disk and press “2”.
7. Take 2-3 new DM readings on each of the stored feedstuff samples using the MT. Be sure
that the samples are completely thawed out if they had been frozen. If you decreased the
code number, you should see a reduced DM reading for each of the feeds.
8. Compare with reference DM reading.
If the MT DM is within 1-2% of the reference DM, no more adjustment is needed.
If the MT DM is still more than 2% higher or more than 2% lower than the reference
DM, make additional adjustments to the calibration code as in step 5.
9. Record the adjusted Calibration Code. Write down the adjusted 6 digit calibration code and
save it with the disk. Use the adjusted code for future calibrations.