EHP KGW 1 Manuale utente

No. 900141 May 2006
Digital Crane Scale Type KGW
User Manual

KGW
1- 2
Contents
Declaration of conformity 3
1. Safety Instructions 4
2. Display and Keypad 5
3. Function Keys 5
4. Display Symbols 6
5. Automatic Power-off 7
6. IR-remote control 7
7. Power supply of electronic device 7
8. Charger/ Charging 7
9. Accessories 7
10. Technical Data 8
11. Parts-Catalog 8
12. Trouble Shooting 10
13. Error Detection 10

KGW
1- 3
Declaration of conformity
Manufacturer EHP-Wägetechnik GmbH
Dieselstrasse 8
D-77815 Bühl (Baden), Germany
declares that the weighing instrument
Digital Crane Scale Type KGW
is conform with following harmonized standards:
EN 45501, EN 50081 part1 according to the provisions of the directive 89/336/EEC
(electromagnetic compatibility, EMC).
EN 60950 according to the provisions of the directive 73/23/EEC (low voltage directive).
This product is marked with the CE sign.
Bühl, May 2006 Stefan Tisch / Technical Manager
This declaration of conformity is suitable to EN 45014.

KGW
1- 4
1. Safety Instructions
Please read this instruction manual carefully before you set digital crane scale into operation
to prevent injuries and protect your digital crane scale against damages.
More satefy instructions and warnings could be found directly in the corresponding chapter.
...before setting into operation
•Use the EC type -approved version of this digital crane scale only in weather protected (roof
protected) environment, so do not expose it directly sun, rain and snow etc.
•Do not operate this digital crane scale outside the nominal temperature range, e.g. in frost
risk environment (less than -10°C) or high heated areas (over +40°C), in order to perform
best accuracy.
•Check if your crane scale is placed absolutely correct and safe into corresponding hook.
•Respect ACCIDENT PREVENTION REGULATIONS.
•Keep this instruction manual.
...in operation
•Pay attention that scale and load are placed correctly.
•Do not use any solvents for cleaning of the crane scale to prevent defects and damages.
...in case of defects
•Repair work is only allowed to be done by qualified personnel.
•See also chapter TROUBLE SHOOTING and ERROR CODES.
IMPORTANT:
Please be careful with seals and marks on EC type-approved crane scales. If these are
damaged, please inform your local office of weights and measures.
Your digital crane scale works by the principle of a high precision strain gauge sensor system
with subsequent signal processing.
The user/operator is responsible for the safety of this weighing instrument, i.e. perform a visual
control of all tension stressed parts (e.g. ring, hook and shackle) at regular intervals.
In case of visual abrasion of these mounting parts, EHP can offer you a complete technical
check according to the German accident prevention regulations.

KGW
1- 5
2. Display and Keypad
3. Description of Function keys
Function Key for Scales Power-On
•Switches on the scale.
•Activates the automatic display segment check.
•After testing without error display is automatically set to “0“.
Please always switch on scale minimum 5 minutes (warm-up time) before
start of weighing. If preload exceeds 20% of crane scales nominal capacity
(initial zero-setting range) then only +/- sign is displayed. Normal operating
status /weighing mode will be reached again by reducing preload under the
20%-limit and power off-/on-switching.
Function Key for Scales Power-Off
•Switches off the scale.
Power off-/on-switching will reset all crane scale errors (reset function).
Function Key for Test- Routine
Activates during approx. 5 seconds the automatic display segment check
and further particulars are displayed.
For approx. 5 seconds display is flashing:
8 8 8 8 8 (Segment Check)
LAH 11.03 (Software version)
- - - 0 1 (Scale No.)
C0 1 (Frequency Channel No.)
H0 1 (IR- Transmitter Channel No.)
Keypad

KGW
1- 6
Function Key to store and to delete Tare Values (TARE- Button)
•(Subtractive) TARE function is activated.
•Current weight value is stored into tare memory.
•Display shows ″0″.
•Net -LED is illuminated.
•Exit by repressing TARE button and GROSS weight is displayed again.
Function Key for Zero-Setting (Zero- Button)
•Sets display to ″0″(semi-automatic zero-setting device).
•Zero-setting range: -1…+3% of nominal capacity.
Zero-setting outside semi-automatic zero-setting range and also in active
tare weighing mode is not possible. Outside of scales 1/4d-range (a quarter
of verification interval) only +/- sign is illuminated.

KGW
1- 7
4. Description of display-symbols
1 - Display
Display consists of 5-digit, 7-segments-LED
2 - Dwell Control Indication
This sign is illuminated when load on crane scales’ hook is motionless
AND a correct weighing result is displayed.
3 - Symbol for Net Weight
This symbol is displayed if a tare value is set. The displayed weight is a
net weight.
4 -
Sign- Symbol
Every weight value is displayed combined with its (algebraic) sign.
Æ
Overload Indication:
If crane scale is loaded with more than 2e (2x scales verification
interval) above the nominal capacity range (see model type
indication plate), so display will switch ‘dark’ and only sign-symbol is
displayed.
Normal weighing status is reached again when crane scale is
unloaded (inside permissible weighing range). Example: Capacity
10t: 2x verification interval 5kg = 10kg, i.e. display is switched ‚dark’
at a weight of 10010kg. In respect of SATEFY AND HEALTH please
do not overload crane scale.
5 - Symbol for Decimal Point
Is displayed to indicate position of decimal point.
6 - Symbol for Battery Control
Scales’ accumulator is to charge below a battery voltage of 5.4 volts,
displayed by a flashing symbol (empty accumulator). Now you
2
46 53 197 8

KGW
1- 8
can still work with your instrument for the next 30 minutes (e.g. if it is
not possible to pause your weighing process). By then the battery
needs to be charged. At a battery voltage of 5.3 volts the scale is
switched off automatically to avoid a total discharge and consequently
accumulator damage.
7 - Symbol for pre-tare-function
This symbol comes together with the - symbol, when a tare-value
was preset by the remote control. The shown weight is a Net- weight
8 -
Symbol for Multirange-scale
Shows if the scale is in range 1 or 2 (only when the scale is adjusted as
multi-range system)
9 - Receiver (diode) of IR- Transmitter
5. Auto Power-Off
This instrument is equipped with an automatic power-off setting device, which is set default =
deactivated. In scales setup mode you can choose a power-off setting time between 1…99
minutes. To activate this device or changing of setting time, please contact after sales service.

KGW
1- 9
6. IR-remote control
By means of IR-transmitter you can activate further additional functions beside scale keypad
functions.
Change transmitters’ batteries (4pcs. a 1.5V Type AAA, Micro, LR03, AM4, MN2400) when
operating distance is getting insufficient (less than approx. 20m).
1 - Press button to input and delete tare values (TARE- Button)
This key has same function as on scales’ keypad.
2 - Press button to activate the PRINT- Function (PRINT- Button)
Through this scales display value (plus additional data) will be transmitted
via UHF radio to receiver units (see accessories), e.g. a single print out will
be activated on the integrated printer of TELEDATA receiver unit.
3 - Press button for zero-setting (Zero- Button)
This key has same function as on scales’ keypad.
4 - Press button to activate the ADD- Function (Add- Button)
Press this button if several different single weights within one charge should
be collected resp. added. Scales display value (plus additional data) will be
transmitted via radio to receiver units, e.g. a single weight registration within
a total print out will be activated on TELEDATAs integrated printer.
1
3
5
7
9
13
12
2
4
6
8
10
11
14
Picture 3.1: IR-remote control

KGW
1- 10
5 - Non-active, only possible in combination with dual-range scales.
6 - Press button to activate Print Total- Function
Through this scales display value (plus additional data) will transmitted be via UHF
radio to receiver units (see accessories), e.g. a total print out (total of all single weight
values will be activated on the integrated printer of TELEDATA receiver unit (additon
of all single weight registrations transmitted by ADD- Function).
7 - Press button to activate Test Routine (Test- Button)
This key has same function as on scales’ keypad.
8 - Press button for scales’ (remote) power- off
Press this button to switch off the scale by remote control. A re-power-on is possible
by pressing any button of IR- transmitter or by pressing function key on scales’
keypad. This function could be set deactivated by factory or in scale setup.
9 - Non-active
10 - Non-active
11 - Enter- Button
This press button transfers a code (max. 5 digits) which is entered by numeric keys
together with actual weight value via radio / RS232 (see accessories) to receiver
devices.
12 - Shift- Button
Press this button to activate numeric keys. Now you can input any numbers (max. 5
digit code) between 0…9 (blue numbers).
13- Symbol for active numeric keys
If this LED is illuminated then you can select the (blue) numbers of IR- transmitter
numeric block (see also Shift- Button).
14- Transmission- Symbol
Transmitting remote control is displayed by flashing LED.
Note:
If functions: Print, Add and Total are activated by IR-transmitter, their characters are also
displayed on scale. Activating of function keys is now locked during the next approx. 5 seconds in
order to prevent transmitting several commands by mistake.