Choose the adaptor US0029
for low flow rate.
4.
When installing Ex-Proof wire,
it is necessary to tighten it with
wrench. Torque:1.5Nm
5.
Please purchase ema qualified
Ex-proof wire for this Ex-proof product as the requirement.
6.
The housing of product in the pipe should be correctly
connected with the equipotential grounding system.
7.
Do not open when an explosive dust asmosphere is present.
8.
1. To screw the nut smoothly, please add the lubricant to nut ③
and threads. (Fig.4)
Notice: It is disallowed to add lubricant on the probe.
2. Screw a suitable adapter ② to the joint ① . (Fig.4)
3. Insert the sensor to the adapter and then screw the nut ③
The Max screwing torque: 50Nm. (Fig.4)
4. The insertion depth for the probe: The minimum insertion
depth to the pipeline ≧ 12mm. To ensure correct depth, the
user can use ema adapter. (optional order)
34