7
SETUP ACCESS: To access instrument configuration, calibration or to enable op-
tions, depress the “ZERO” key for five seconds.
The Audit Trail counters (“Pxxxx” and “Cxxxx”) are displayed first followed by access
code request (“AC?”). The initial factory setting is “0000” which is entered with four
numeric “0”’s and enter (“ENT”). If no entry is made, instrument returns to operate
mode. The “Master” access code is “1042”.
The access code can be changed to any four digit combination during setup exit
when display again shows “AC?” or skip with ENT to continue with no change,
The “CHECK” key provides the software version “V X.XX” followed by the display
test routine. Use the “ENT” key to advance to the keyboard test and to exit tests.
After entry, use the “TARE RECALL” key to select a main menu; Configuration
(“SEL.CFG”), Calibration (“SEL.CAL” or “SEL.CL1”), or Options (“SEL.OPX”)
and “START/STOP” to back step.
The “GROSS/NET” key enters selected menu and is used to step through sub
categories. Individual parameter selection is made with the “TARE RECALL” key,
which then steps through the parameter choices (“ZERO” key back steps within the
menu).
The “ENT” key is used at any point to “back” up from categories to menus and to
“save?” and “AC?” and exit.
During the exit steps, if changes were made, the display is “save?” with alternate
“no”. To save changes, use the “UNITS” key to select “yes” and “ENT” to exit.
Calibration numeric entries are entered directly followed with the “ENT” key.
Front panel access is inhibited if conventional “sealing” is applied by setting jumper
J1-1 in the B position. The board mounted “CAL” button is then used for access.