erlab Captair Flow Smart Manuale utente

Safer
Simpler
User manual
Captair Flow Smart
Mobile ductless filtering clean air enclosure

Captair Flow Smart
Mobile ductless ltering clean air enclosure
2
Contents
General...........................................................................................................................................................................................................................................................3
Safety notices........................................................................................................................................................................................................................................3
Product registration ........................................................................................................................................................................................................................3
Start-up..........................................................................................................................................................................................................................................................4
Calibrating the electronic anemometer
Default settings
Description of the control module
Description of the alarms
eGuard App..............................................................................................................................................................................................................................................6
Connectivity principle
Options of connection
Accessing the Administrator interface
Administrator interface description
Fan setpoints per filtration column type................................................................................................................................................................15
Replacing the filters ........................................................................................................................................................................................................................16
Recommendations for storing and using the filters................................................................................................................................20
Replacement frequency for filtration breakthrough sensor (Molecode option).....................................................21
Admissible weighs on worksurfaces .........................................................................................................................................................................21
Cleaning and maintenance....................................................................................................................................................................................................22

3
General
Choosing Captair Flow Smart ductless ltering clean-air enclosures is choosing an effective and responsible pro-
tection for your handlings.
Erlab’s 50 years of expertise in the ltration systems for laboratories ensures the highest protection for handlings requiring an
ultra clean environment. The new Captair Flow Smart ductless ltering clean air enclosure uses an innovative and straight-
forward mode of communication called Smart technology. This powerful interface uses light to intuitively and effortlessly
communicate with users and leave them free to focus all their attention on the main task: their application.
Captair Flow Smart ductless ltering clean-air enclosures are designed to provide an ultra-clean and dust free workstation.
It features a high efciency system with HEPA H14 (or ULPA U16) ltration technology to achieve an ISO class 5* work envi-
ronment according to the ISO 14644-1 standard.
The system’s connectivity allows for real-time safety alerts and individual device usage reports to be sent via the e-Guard
App.
*Considering a room pollution of up to ISO 9
Safety notices
Proper use, monitoring and maintenance of the device by users will assure its effectiveness. Your laboratory may also be-
net from ergonomic, economic and ecological advantages provided by the Captair Flow Smart throughout its life cycle.
• The equipment provided is not intended to be used in an explosive atmosphere.
• The lters delivered with this device must be removed from their packaging and positioned correctly.
• Erlab recommends that the electronic anemometer is calibrated at least once a year.
• New lters must be stored in their packaging, kept in a dry location. (see recommendations for storing and using the
lters)
Product registration
Take full advantage of the device’s connectivity to enhance your safety
Get up to 10 years warranty on your connected Erlab unit
Register your product online: the registration of the product will automatically give you one extra year of warranty (in addition
to the warranty mentioned in the Erlab’ general terms and conditions of sale).
Connect your unit: Once the device is connected to the Internet and congured to exchange usage data, the warranty is
extended for up to 10 years. Warranty will be successively renewed at each lters replacement and for the life time indicated
on the Valipass® and/or or at the end of lter usage time.
In order to benet from Erlab extension of warranty offer, the following conditions shall be respected:
Warranty applicability is subject to the respect of the Erlab’ general terms and conditions of sale and following requirements:
• The registration and/or the connection of the product shall be performed within the twelve months from the purchase
date;
• Filters replacement must be performed following Valiquest® service recommendations or at the end of lter usage
time; The lter’s serial number, used as an identication key, validates this condition, regardless of your device’s
supplier (and/or the replacement lter’s supplier for the following years);
• The device’s replacement lters must be manufactured by Erlab, as must all other spare parts.
Consumables such as lters and saturation sensors are not covered under warranty.

Captair Flow Smart
Mobile ductless ltering clean air enclosure
4
Start-up
Calibrating the electronic anemometer
Be sure the sash is completely installed before starting the calibration.
1. Press and hold the mute button located on the front of the control panel while simultaneously switching on the main
power button located behind the control panel (Off = O, On=I)
2. Release the mute button. A beep indicates that the unit has started the calibration mode. The white bar above the
front of the control panel will pulse for 5 minutes and 30 seconds. Stick with just the instructions for the anemometer.
3. Once the pulsing light stops, the anemometer is calibrated and the unit is ready to be turned on.
4. To operate, push the power button located on the front control panel.
Good practices :
• Avoid any air turbulence in the area during calibration
• Anemometer should be calibrated every year and every time a modication is done on the ltration column
Having carefully followed the steps described in the installation guide, your Captair Flow Smart ductless ltering clean air
enclosure is now ready to use.
Default settings
The power switch is located at the back of the control panel.
Note: we recommend never turning off the main device power switch after the machine as
been started for the rst time.
The button on the control panel turns on the fan and the lights.
The green indicator light and LED light system should come on.
We also recommend verifying the operating parameters before each new use.
To modify settings, please access the administrator interface.

5
Description of the controller
2
3
41
1. Switch on fan and lights in
workstation
2. Silence the alarm (Mute key)
3. Smart-Light: pulses when an
alarm occurs
4. Alarms pictograms - to indi-
cate which alarm is active
4
Smart Technology communicates simply about performances and lter ef-
ciency of your product via a LED band of light called Smart-Light. This light
signature casts a steady LED glow when unit is operating properly. If normal
operation is disrupted, the LED signature simply pulses, drawing the atten-
tion of the operator only when necessary.
Description of the alarms
Note:
When using the Mute key to silence the alarm, please note the alarm can be triggered again if the problem has not been xed.
Resetting alarms via the Mute key will consequently modify usage settings.
Please access the administrator interface to verify user settings.
Alarm
type
Light
signal Problem Details Silence
the alarm Reset the alarm
Air speed
1 beep
5 seconds
apart
Pulses Low air
speed
The air speed
value is
<0.3 m/s
Press Mute
key
Please check the front sash is
correctly closed
Check particulate lter.
If the lter is not clogged up,
perform a new calibration of
the anemometer (with door
opened)
Fan
2 beeps
5 seconds
apart Pulses
Fan fault
The rotation speed
(RPM) is +/-
10% of
the fan setpoint. Press Mute
key
Please contact Erlab or your
usual maintenance contact.
Fan
Unserviceable
The rotation
speed (RPM)
is < 700 RPM
Filtration
3 beeps
5 seconds
apart
Pulses
Filter breakthrough
(Molecode S/A/F
option)
The Molecode detection
value is > the sensitivity
setting for a period of 40s. Press Mute
key
Please contact Erlab or your
usual maintenance contact.
Replace lter
The lter(s) has/ have
reached the end of their
service life/lives
Filter breakthrough sensor replacement (Molecode)

Captair Flow Smart
Mobile ductless ltering clean air enclosure
6
Alarm
type
Light
signal Problem Details Silence
the alarm Reset the alarm
4 beeps
5 seconds
apart
Pulses
Filter breakthrough
sensor replacement
(Molecode S/A/F
option)
The sensor has reached
the end of its service life.
Please contact Erlab or your
usual maintenance contact.
Reset network settings
Forgot network settings? Follow these steps to reset the IP address:
• The main power switch is in the “On” position (located on the left of the controller behind the panel)
• The workstation should not be running (no lights or fans are “On”),
• Hold in the mute button for «10» seconds
• Wait until you hear «click», «click»
• Release the mute button
• Reboot unit:
1. Turn the power switch to the “Off” position (located on the left side of the controller).
2. Turn the power switch to the “On”.
• The Primary IP address to use is 192.168.0.200.
Note: In some cases, you may need to use the secondary IP address 192.169.0.200.
eGuard App
Welcome to a safer connected world
The connectivity of Erlab Smart devices allows you to monitor all your safety settings remotely.
After registering your product on line, download eGuard App and:
• Stay in touch wherever you are
• Receive safety alerts
• Access your statistics usage
• Make the most of an exclusive warranty program
Connectivity principle
Ecosystem designed for simpler use and safer protection
Embedded
service
Embedded
service
Embedded
service
Embedded
service
Fume hood storage cabinet Air Filtration System Air quality sensor
Internet network and/or local area
network (LAN - for eGuard PC only)
Database
Monitoring
the product
Monitoring &
controlling
the product
eGuard App
Mobile
PC
Ethernet-TCP-IP

7
3 methods of
connection to
the Smart Portal
Embedded
service Mobile app PC app
Conditions
of use
Direct connection on PC
with data cable (RJ45)
Web connection
(via 3G/4G)
Web and/or local
connection
Hardware requirements 1 PC + 1 cable 1 Apple or Android
Smartphone
1 PC connected to
Internet or local network
Parameters Monitoring + Controlling Monitoring Monitoring + Controlling
Data access One unit Multiple units Multiple units
Historical data access
Historical data download
Alerts
Monitor multiple devices
Allow multiple user accounts
Automated status report
Download available on
www.erlab.com
Options to connect
Embedded service
RJ45
Mobile or computer app
and/or
Internet
network
Flash the QR code to
access usage datas of
the device.

Captair Flow Smart
Mobile ductless ltering clean air enclosure
8
Accessing the administrator interface
To monitor the parameters and modify the settings of the unit.
In order to connect:
• Use a computer equipped with an Ethernet port (to plug the RJ45 cable)
• WIFI of the computer must be switched off
• Web browser (Internet Explorer, Edge, Chrome, Mozilla Firefox, Safari, …) must be installed on the computer
Note: RJ45 cable used to plug the unit to the computer is provided.
1Direct connection on computer
• Check that main switch of the device is ON.
• Take RJ45 cable (black) is already connec-
ted to the unit and placed behind the back
of the control panel.
2 Open your web browser, type the following IP address 192.168.0.200 into the address bar and validate
OK
You are connected to the embedded software
You enter the « Status » page and you can have
access to the « Settings » using the following
credentials:
Login : erlab / Password : smart
Please go to page 12
Page is not accessible
Computer network parameters are not allowing
the access to the embedded software.
Apply the following procedure

9
Page is not accessible
3Modify computer network parameters
Right click
Left click
Network
parameters
4Access to the Network and sharing center
Left click

Captair Flow Smart
Mobile ductless ltering clean air enclosure
10
5Access to the network connection
Right click
Left click to access properties
Note : This conrms
your WIFI is
disconnected!
6Enter compatible network parameters as indicated below
Left click
Left click