GPG CL800 Manuale utente

CL800
Free Chlorine & pH Monitor
OPERATION GUIDE

CONTENTS
1.
INTRODUCTION…..…………………………………..….……..
2
2.
FEATURES AND TECHNICAL SPECIFICATIONS……….
…
2
2.1
Features..…………………………..……………...………..
2
2.2
Technical S ecifications……...…………….…………......
2
3.
INSTALLATION…….…………………………..………..………
3
3.1
Dimension………………...……...…………………..……..
4
3.2
Connection.…………….……..………..…….....….………
4
4.
SETTING AND OPERATION…..…………...…….……………
6
4.1
User Interface and Descri tion………………….........…..
6
4.2
Parameter Setting and O eration.……..…..….……...…..
7
4.2.1
Password…………………………………………………….
9
4.2.2
Main Setting Menu Directory………………………………
10
4.2.3
Set Parameter……………...……………………………….
10
4.2.4
Set Alarm………….……………...…………………………
14
4.2.5
Set Current Out ut………...…..……………..…………….
17
4.2.6
Calibration……………….…………..……..……………….
18
4.2.7
Save and Read Data………….………..…………………..
23
4.2.8
Restore Factory Setting.…………………..……………….
26
5.
SENSOR…….……………………………….…..……………….
27
6.
WARRANTY……...………………………………………………
30
7.
STANDARD CONFIGURATION…..…………………..……….
30
8.
OPTIONAL CONFIGURATION…………..………...…….…….
30
1

CL800
Intelligent On line Free Chlorine & pH Monitor
1. INTRODUCTION
The CL800 is a micro rocessor controlled Free Chlorine and H
measurement instrument. The unit utilizes a multifunction LCD to dis lay
readings and rovide feedback to the user. It is available with different
o tion to rovide fully configurable control, alarm and feedback with u
to four relays and two 4-20mA current out ut sources. It can be used to
detect the Free Chlorine and H value of the solution in chemical,
etroleum, rubber, a er, tobacco, alcohol, sugar, food, mining,
smelting, iron and steel, ower generation, harmaceuticals, water
treatment and other industries.
2. FEATURES AND TECHNICAL SPECIFICATIONS
2.1 Features
(1) Large back-lit multifunction LCD dis lay
(2) Wall or Panel mounting is available
(3) Man-machine interface is friendly
(4) Measured rocess Free Chlorine, H, tem erature
(5) 0 ~ 100℃ automatic/manual tem erature com ensation for H
(6) Manual or Automatic buffer adjustment
(7) Restore factory setting function is available
(8) ROM inside, measurements and time can be rogrammed store
(9) Galvanic se aration between in uts and out uts and su ly voltage
(10) Different in ut for excellent noise rejection
(11) High and low rogrammable alarm, 250V/10A relay out ut
2.2 Technical Specifications
(1) Ranges of measurement:CL: 0~2 m、0~5 m、0~10 m
PH: -1.00~15.00 H
2

Tem : 0~100℃
(2) Accuracy:±0.5 %F.S m / ±0.2 %F.S PH / ±0.2℃
(3) Linearity:±0.5% of range
(4) Re eatability:±0.5% of range
(5) Tem erature com ensation ty e:Auto / manual 0℃ to 100℃(for
H)
(6) Alarm Out ut:Four relays out uts (250V/10A),full range with
hysteresis adjustable
(7) Current out ut:Two DC 4~20mA,O to-isolated out uts,( 750Ω Max.
load)
(8) Ambient O erating tem erature:-10~+55℃
(9) Humidity:≤95%
(10) Power su ly:AC100~240V,50~60Hz
3. INSTALLATION
De ending on the installation osition, use the attached mounting
bracket and fastened it horizontally or vertically on the rear of the
instrument with screws, then fixed on the wall or anel as figure 1.
3
Figure 1 : installation diagram

3.1 Dimensions
3.2 Connection
Remove the four bolts on the front of the instrument, o en the case
gently, you can see a row of terminals shown in Figure 3, lease wiring
follow the directions.
Connection terminals wiring directions:
1. PH+, H sensor central line (+)
2. PH-, H sensor shield (-)
4
Figure 2 : overall dimensions panel-mounting
Figure 3 : connection terminals diagram

3. CL+, Free Chlorine sensor central line (+) <Red>
4. CL-, Free Chlorine sensor shield (-) <Black>
5. T1, Pt 100 tem erature sensor(1)
6. T2, Pt 100 tem erature sensor(2)
7. CLI-, 4~20mA current out ut for Free Chlorine (-)
8. CLI+, 4~20mA current out ut for Free Chlorine (+)
9. PHI-, 4~20mA current out ut for H (-)
10. PHI+, 4~20mA current out ut for H (+)
11. CLL1, Free Chlorine low alarm relay out ut (N/O, normally o en)
12. CLL2, Free Chlorine low alarm relay out ut (Common)
13. CLH1, Free Chlorine high alarm relay out ut (N/O, normally o en)
14. CLH2, Free Chlorine high alarm relay out ut(Common)
15. PHL1, H low alarm relay out ut (N/O, normally o en)
16. PHL2, H low alarm relay out ut (Common)
17. PHH1, H high alarm relay out ut (N/O, normally o en)
18. PHH2, H high alarm relay out ut(Common)
19. QX1, cleaning relay out ut (N/O, normally o en) <O tional>
20. QX2, cleaning relay out ut (Common) <O tional>
21. Power su ly terminal: Connect AC100 ~ 240V
22. Power su ly terminal: Connect the ower su ly hase
★
CAUTION ! :
The s ecified erformance of the CL800 is entirely de endent on correct
installation. For this reason, the installer should thoroughly read the
instructions before attem ting to make any electrical connections to the
unit.
5

4. SETTING AND OPERATION
4.1 User Interface and Description
Front anel descri tion:
(1) CL alarm light,Under the conditions of setting Free Chlorine high or low
alarm, when the measured Free Chlorine value of the solution is
higher than the value of High alarm or lower than the value of Low
alarm, the CL alarm light will be turned on.
(CL high alarm light when only CL measuring)
(2) PH alarm light,Under the conditions of setting H high or low alarm,
when the measured H value of the solution is higher than the value
of High alarm or lower than the value of Low alarm, the PH alarm light
will be turned on.
(CL low alarm light when only CL measuring)
(3) UP key. In the setting state, using the UP key the user can cycle
through the front menu. To adjust a value, the UP key is used to
6
Figure 4: front panel diagram

Figure 5 measurement state 1
increment the digit.
(4) MENU switch. Press this key to enter or exit the setting state.
(5) DOWN key,In the setting state, using the DOWN key the user can
cycle through the next menu. To adjust a value, the DOWN key is
used to select a digit.
(6) ENTER key is the enter button to confirm enter the menu and store
the setting arameters.
(7) Meas indicator light, the light will be turned on when entering the
measuring state.
(8) SET indicator light, the light will be turned on when entering the
setting state.
(9) LCD digital Monitor ,dis layed the measured values ( Free
Chlorine、H、℃),and can also be dis layed rom t function,
arameter values and error codes in interactive. (When measure
Free Chlorine only, H will be dis layed “none”, and tem erature will
be dis layed “25 ”℃)
4.2 Parameter Setting and Operation
When the instrument is owered security, after a brief self-test rogram,
Meas indicator lights and screen dis lay the Free Chlorine value, H
value, tem erature and time simultaneously as Figure 5. Indicates that the
instrument is working in the measurement state.
7

Instrument has two kinds of o erating states: "Measuring" state and
"Setting" state. In the measuring state, there are two kinds of dis lay:
“only Free Chlorine” and “Free Chlorine & H” dis lay status. Generally,
when the instrument is owered on, it enter into the measuring state. Use
the MENU key to switch the "Measuring" state or "Setting" state. Press
MENU key in the measuring state, it will enter the setting state, and the
SET indicator light will be turned on.
At this oint ress the UP or DOWN
keys to select various setting functions, ress the MENU key to return to
measuring state.
If there is an error code “ER01” dis lay on the osition of measurement
reading as Figure 6, indicated the Free Chlorine value is lower than
measuring range. (In the case electrode unused for more than two hours,
it need activation for period of time, during which the meter readings will
e displayed “ER01”)
If there is an error code “ER02” dis layed, indicated the Free Chlorine
value is higher than measuring range.
If there is an error code “ER03” dis layed on the osition of the
tem erature, indicated that there is no tem erature in ut signal.
If there is an error code “NONE” dis layed in front of H, indicated that
there is no H in ut signal.
Figure 6 measurement state 2 (error code)
8

Instrument has the indication of cleaning function (o tional) on the screen.
When the cleaning function is o ened, there is a “●” mark on the screen
as Figure 7. When cleaning function sto ed, cleaning mark disa ear.
Figure 7 measurement state 3 (cleaning state)
4.2.1 Password
The CL800 Monitor is locked to revent unauthorised access to the
configuration menu’s by using a assword. If you want to enter the
setting menu, you must enter correct assword.
The user assword is a 4 digit number that can be set to any value the
user chooses. When ress the MENU key, the screen will dis lay:
When su lied new, the user assword is set to 2223, we recommend that
this is changed when the instrument is installed. Use UP and DOWN
to
modify it, then ress ENTER to enter the main setting menu.
*Note: The factory default assword is 2223, if the assword has been
forgotten lease contact Gold oint technical su ort.
9