Gram VK Series Manuale utente

USER’S MANUAL
EN
SERIE
VK

CONTENTS
1. Introduction ........................................................................................................ 3
2. Precautions ......................................................................................................... 3
3. Before Using the Product .................................................................................. 3
4. Product Introduction .......................................................................................... 4
5. Analog Section ................................................................................................... 6
6. Setting mode ....................................................................................................... 7
7. Dual Channel ...................................................................................................... 8
8. Operation .......................................................................................................... 12
8-2-1 Tare .............................................................................................................................. 12
8-2-2Preset Tare ................................................................................................................... 12
8-2-3 Auto-tare ...................................................................................................................... 13
8-3-1Key in a known unit value ............................................................................................. 14
8-3-2 Sample counting .......................................................................................................... 14
8-3-3 ACAI ............................................................................................................................ 14
8-4-1 Choose Accumulation Mode ........................................................................................ 15
8-4-2 Set the Max Account of accumulation ......................................................................... 16
8-4-3 Manual Accumulation .................................................................................................. 16
8-4-4 Accumulation display ................................................................................................... 17
8-4-5 Print while display of accumulation .............................................................................. 17
8-4-6 Delete accumulation .................................................................................................... 17
8-5-1 Set a Print form ............................................................................................................ 18
8-5-2 Set Print mode ............................................................................................................. 21
8-5-3 Select the Printer Model .............................................................................................. 21
8-6-1 Set Serial Number Mode ............................................................................................. 22
8-6-2 Set Max Serial Number ............................................................................................... 22
8-6-3 Change the Current Serial Number ............................................................................. 23
8-7-1 Set the high limit of quantity ........................................................................................ 23
8-7-2 Set the low limit of quantity .......................................................................................... 23
8-7-3 Enable / Disable the quantity check function .......................................................... 24
8-8-1Set the high limit of weight ............................................................................................ 24
8-8-2Set the low limit of weight ............................................................................................. 25
8-8-3Enable / Disable the weight check function .................................................................. 25
8-9-1Saving ID method 1 ...................................................................................................... 25
8-9-2Saving ID method 2 ...................................................................................................... 26
8-9-3Call up ID ...................................................................................................................... 26
9. Option ................................................................................................................ 29
10. Relay Module Diagram ................................................................................... 34

1. Introduction
Thank you for deciding to purchase a JCAII counting scale. This goods has the excellent
performance and splendid properties under severe quality management .It is recommended to
read this manual in full before using it for good function application.
2. Precautions
◎ Place the scale on a flat and stable surface. See 3-3 Leveling the Scale for details.
◎ Verify that the input voltage and the plug type matches the local AC power supply.
◎ Warm up the scale for 15 minutes before using it the first time.
◎ Keep the scale away from EMI noise, strong wind and vibration, which might cause incorrect
reading.
◎ Avoid sudden temperature changes (suitable operating temperature is between -5℃~ 40℃.)
◎ Disconnect the power supply when cleaning the scale.
◎ Do not immerse the scale in water or other liquids.
◎ Service should be performed by authorized personnel only.
3. Before Using the Product
3-1 Unpacking and Checking
Open the package and check the instrument for transport damage. Immediately inform your
dealer if you have complaints or if parts are missing. The package should contain:
● Scale body ● User manual
● Stainless steel pan ● Wire (power cord)
3-2 Installing Components
Cover the stainless steel pan on the top housing properly before using the scale.
Important:
Please find the delivery protection screw in the bottom of the scale. Loose it according to
the instruction sticker next to it
Before use the scale Loose the screw counter-clockwise to the end
Before move the scale Tighten the screw clockwise to the end
3-3 Leveling the Scale
To compensate for small irregularities or inclinations at the location, the scale can be leveled. The
scale is equipped with a level indicator at the front panel .Adjust the adjusting feet until the air
bubble in the indicator is centered as shown.
Note: The scale should be leveled each time its location is changed.

4. Product Introduction
4-1 Display
1) Weight Window: displays weight of object on weighing pan, or accumulated weight value.
2) Unit Weight Window: displays the average piece weight value, or number of weighments .
3) Total Count Window: displays the calculated number of pieces on the weighing pan, or
accumulated piece count value.
4) Tare or Preset Tare indication.
5) Center of Zero Indication, The zeroing range is ±2﹪of weighing capacity.
6)Symbol “ ” points at “PRESET TARE” when Tare value entered via numeric keys.
7) Under the sample counting mode, Symbol “ ” points at “SMPL” when calculated
unit weight is lower than 4/5 of scale division or the quantity of sample is less than 10 pieces.
8) “Accumulation” function is enabled under weighing mode.
9) Under entering a known unit weight mode, Symbol “ ” points at “UW” when entered unit
weight is lower than 4/5 of scale division. Unit weight is too small for ensuring accurate quantity
calculations.
10)Low Battery Indicator: When “ ” appears, the battery power requires recharging.
11)CHANGE Charge Lamp
Red--- battery is charging
Green---battery is fully charged

ACAI Symbol “ ” points at “ACAI” when Automatic counting accuracy improvement is
functioning.
STABLE Stable indication, Symbol “ ” appears at the top of “STABLE” when scale is in
stable condition.
NET Gross weight minus Tare. Symbol “ ” points at “NET” when manual Tare or preset are
actions are done.
4-2 Keypad
·Enters specific values for tare, unit weight and other data entries.
·Key can be also as a space key when edit label.
·Decimal point
·Initiates function setting
·Sets the weight to zero.
·Clears the indicated input values.
·Displays gross and net weight by turns
·Inputs unit weight and run its counting function

·Enters the sampling mode and change its scale choice(scale 1 or 2)
·Switches between main scale and remote platform
·
Adds the indicated weight or piece count value into Accumulation
memory.
·Enters label editing mode
·Steps to next parameter
·Deletes accumulation records
·Display year, date, time
·Go back to previous parameter
·Recalls and displays the accumulated data
·Selects/ sets the Hi Value of weight or piece
·change the next content of parameter
·units switch for the pre-tare function
·Selects the Lo value of weight or piece
·Changes to previous content of parameter
·Tare
·Give up any setting and return to normal weighing
·Call up unit weight ID
·Enable / switch on relay output function under HI/LO setting.
·Print
·Disable/ switch off relay output function under HI/LO setting.
5. Analog Section
Input sensitivity
1~4mV/V
Zero adj. Range
20% FS
A/D conversion rate
10 times per second.
Load cell excitation
5V DC
Number of load cells.
up to 4x350 ohm loadcells
Resolution.
1/30,000

6. Setting mode
1. Under the weighing mode, long press key for 3 seconds to enter basic
parameter setting mode.
2. Press or to switch between parameters from P-00 to P-07
3. Press key or to step through all the available setting options.
4. Press key to save the setting and return to weighing mode.
Parameter description
1. P-00 Power (default: off) Set the period of inactivity before the scale automatically turns off.
Options are Off =non power-off, 10, 20,30,40,50, 60, 70, 80 and 90 (minutes).
2. P-01 Backlight (default: auto)
=always off =always on =turn on when the weight is over 20e
3. P-02 Serial number display (default: off)
=never display =will display when price is zero
4. P-03 Beep sound (default: in)
= when check is OK = when check is Hi/Lo ( , = using scale’s buzzer)
- =when check is OK - =when check is Hi/Lo ( - , - =using light tower’s
buzzer)
5. P-04 ACAI (default: off)
Options: off/5/10/15/20/25/30/35/40/45
6. P-05 Printer model (default: normal)
Options: normal / EZ/ BP/SH/ZEBRA (normal as compatible printer)
7. P-06 Rs-232 Baud Rate (default: 9600)
Options: 19200 / 9600 / 4800 / 2400
8. P-07 Data (default: n81)
Options: n81 / o81 / e81 / n71 / o71 / e71

7. Dual Channel
7-1 Diagram of sub channel connector
7-2 Installation Process
1. Turn on the main scale
2. Connect the remote platform to JCAII main scale
7-3 Setup
1) Under the weighing mode, long press key for 3 seconds to enter basic parameter
setting mode.
2) Long press key to enter advance parameter setting mode.
3) Press key or to select parameter A-08.
4)Press key or to step through all the available setting items.

= Using main scale together with the remote platform
= Use the remote platform only
=Use the main scale only
5) Press key to save the setting and return to weighing mode.
7-4 Maximum Capacity & Division Setting & Unit
1) Press and hold while turning on the scale. Do not release till is
displayed in the Total Count window.
2) Use numeric keys together with to input the new capacity, press key to clears
the indicated input values.
3) Press key advance to division setting. Use numeric keys to input the new division.
4) Press key to advance to Maximum Capacity & Division Setting of the remote platform.
5) Repeat step 2-3 to complete the settings.
6) Press key to save the settings and return to weighing mode.
7)Press key to change its unit: kg, lb, g.

7-5 Calibration
= remote platform
= main scale
Note:
l Please calibrate the main scale first.
l The calibration steps of main scale and remote platform are the same. Press key to
switch the calibration for the main scale or remote platform before the zero point calibration.
l You can only do the calibration for the remote platform after you have set the max. capacity
and division of the remote platform.
Here take JCAII-6K as an example.
1. Press key and hold while turning on the scale. Do not release till
is displayed in the Unit Weight window.
3. With no load on the weighing pan, press key to start the zero point calibration and wait
till is displayed in the Unit Weight window. Input the first calibration weight value
by the numeric key (6K model, 1/3 of full load is 2kg). And then put the corresponding weights
on.
3... Press key and wait till is displayed in the Unit Weight window. And then put
the corresponding weights on.
4. Press key and wait till is displayed in the Unit Weight window. And then put
the corresponding weights on.
Indice
Altri manuali Gram Scala

Gram
Gram RZ Series Manuale utente

Gram
Gram SBZ-600 Manuale utente

Gram
Gram AHS 300 Manuale utente

Gram
Gram AHV Series Manuale utente

Gram
Gram C7 Series Manuale utente

Gram
Gram FD 410 SERIES Manuale utente

Gram
Gram EK Series Manuale utente

Gram
Gram QUBIC ONE Manuale utente

Gram
Gram RS Series Manuale utente

Gram
Gram RV-300i Manuale utente

Gram
Gram FV 220 Manuale utente

Gram
Gram SPX 600D Manuale utente

Gram
Gram ACUREX TXM Manuale utente

Gram
Gram S3 SERIES Manuale utente

Gram
Gram SILK ISLAND Manuale utente

Gram
Gram C6 1T Manuale utente

Gram
Gram S3 SERIES Manuale utente

Gram
Gram CM-3 Manuale utente

Gram
Gram SILK Q Manuale utente

Gram
Gram GX Series Manuale utente