Hardy Process Solutions BSX Series Manuale utente

HARDY HIBSX BENCH SCALES
Installation & Operations Manual
Hardy Process Solutions
Corporate Headquarters
10075 Mesa Rim Road
San Diego, CA 92121
Phone: (858) 278-2900
Web-Site: www.hardysolutions.com
Hardy Process Solutions, #0400-HIBSXMA-0522A
Copyright © 2022 Hardy Process Solutions. All Rights Reserved. Printed in the U.S.A.

2
CAUTION:
UNPACK WITH CARE
WHEN UNPACKING, DO NOT DISCARD THE PACKING CASE OR ANY
PACKING MATERIAL, UNTIL THE CONTENTS OF THE PACKING CASE
ARE INSPECTED AND CAREFULLY COMPARED WITH THE SHIPPING
DOCUMENTS.
IF ANYTHING IS UNSATISFACTORY, PLEASE NOTIFY HARDY
PROCESS SOLUTIONS IMMEDIATELY BY CALLING, FAXING OR E-
MAILING TO:
Hardy Service Center
Hardy Process Solutions
10075 Mesa Rim Road
San Diego, California 92121
Phone: (800) 821-5831 or (858) 278-2900
FAX:(858) 278-6700
Web Address: www.hardysolutions.com
TO RETURN DEFECTIVE OR DAMAGED PRODUCT(S) CALL HARDY
TECHNICAL SUPPORT FOR A HARDY SERVICE TICKET NUMBER
(HST#). YOUR COMPANY NAME, ADDRESS, TELEPHONE NUMBER,
SERIAL NUMBER OF THE UNIT AND A BRIEF DESCRIPTION OF THE
PROBLEM SHOULD BE READY WHEN CALLING. FOR ALL
NON-WARRANTY REPAIRS AN RMA NUMBER AND A PURCHASE
ORDER OR CREDIT CARD ISREQUIRED.
IN CASE OF DAMAGE DUE TO SHIPPING, NOTIFY THE DELIVERING
CARRIER IMMEDIATELY FOR AN INSPECTION.

3
Table of Contents
Contents
HARDY HIBSX BENCH SCALES ..................................................................1
Table of Contents...........................................................................................3
CHAPTER 1 - OVERVIEW ............................................................................3
The Top Cover ...........................................................................................4
BSX Models ...............................................................................................5
CHAPTER 2 - SPECIFICATIONS..................................................................6
CHAPTER 3 - INSTALLATION ......................................................................9
Adjusting the Mechanical OverloadStops ..................................................9
Maximum Deflection at Capacity:...............................................................9
HISP6 - 0.17mm ........................................................................................9
HISPA22 - 0.4mm ......................................................................................9
CHAPTER 4 - CALIBRATION........................................................................5
About Calibration of the Hardy Bench Scale ..............................................5
Test Weight Calibration (Hard Cal) ............................................................5
CHAPTER 5- TROUBLESHOOTING.............................................................6
Service and Repair (All
Models)...................................................................7
CHAPTER 1 - OVERVIEW
Introduction Congratulations on your purchase of a Hardy
Bench Scale. Behind your purchase is over
100 years of quality, reliability and technical
innovation.
The Hardy Bench Scales are designed for table-
top or floor applications requiring high accuracy
and reliability. The covers are all stainless steel
for durability. The bases have no bearings, spirit
levels, or moving parts that can be damaged or
wear out. The active element of the scale is a
precise and accurate strain gauge load sensor.

4
Description All Hardy BSX Bench Scales come standard in
12”x12”, sizes, and range in capacity. Custom
sizes and capacities are availablefrom Hardy
upon request.
The basic Hardy Bench Scale major components:
The Top Cover
Top and
Bottom
Plates
1.
Stainless Steel Top Cover.
2.
Top and Bottom Plates
3.
Load Sensor
4.
Four Leveling feet.
The top cover is a single piece constructed of
304 Stainless Steel. The cover provides a
washdown weighing surface and offers protection.
The top and bottom plates come equipped with
four leveling feet, a load sensor, a shock load
and corner overload protection.

5
Scale Interface Directly attach the scale’s cable to any weighing
instrument. If interfacing with a Hardy instrument,
you can take fulladvantage of WAVERSAVER®
and C2®Electronic Calibration.
NOTE: WAVERSAVER® and C2® are
registered trademarks of Hardy Process
Solutions.
BSX Models
SS-SS
The SS-SS version of the BSX series includes a
stainless IP69K load cell, providing the maximum
amount of washdown capability and protection for
the harshest sanitation environments.
The frame is precision cut from a single piece of
.250 in 304 grade stainless plate.
SS-AL
The SS-AL version of the BSX series includes a
aluminum IP67 load cell with the lowest profile
available and washdown capability.
The frame is precision cut from a single piece of
.250 in 304 grade stainless plate.

6
CHAPTER 2 - SPECIFICATIONS
Hardy HIBSX Bench Scale Specifications
Platter Construction
304 Stainless Steel (all models)
Size: 12" x 12"
Deck Height
SS-SS - 3.4"
SS-AL - 2.7"
Capacity Range
5 lbs to 200 lbs depending on Model
Rated Output
Overload: 150% R.O.
Corner loading: 100% R.O.
Cable Length
SS-SS - 10ft
SS-AL - 6.5ft
Base Construction
Stainless Steel
Temperature
Operating: -10 to +40 degrees celsius
Storage: -20 to +65 degrees celsius
Warranty
Two year limited warranty
Approvals
FM IS Hazardous Class I, Div 1 Load Cell
Refer to cell certifications for details
Shipping Weight
12" x 12": 22 lbs (11 kg)

7
Load Sensors
Specifications HISP6 (SS-SS) HISPA22 (SS-AL)
Metallurgy 17-4PH Stainless Steel 2024 Aluminum
Rated Output (ES) 2±0.10mV/V 2±10% mV/V
Non-Linearity <±0.0166 % R.O. <±0.0166 % R.O.
Hysteresis <±0.0166 % R.O. <±0.0166 % R.O.
Zero Balance <±5.0 % R.O. <±5.0 % R.O.
Combined Error <±0.02 % R.O. <±0.02 % R.O.
Creep @ 30 Min. <±0.0166 % R.O. <±0.0166 % R.O.
Temp Effect Output <±0.0112 % R.O./C <±0.0233 % R.O./C
Temp Effect
Sensitivity
<±0.010 % R.O./C <±0.010 % R.O./C
Input Resistance 1100 ± 50 ohm 413 ± 20 ohm
Output Resistance 960 ± 50 ohm 350 ± 25 ohm
Insulation
Resistance
≥5000 Mohm ≥5000 Mohm
Excitation 5 - 15 vdc 5 - 15 vdc
Safe Load Limit 200 % Emax 150 % Emax
Safe Side Load 100 % Emax 100 % Emax
Hazardous FM IS, Class1, Div 1 N/A
Approvals IP68/IP69K IP67
Warranty Two years Two Years

8
Standard Sizes & Capacities
PartNumber
C2Load
Cell
Capacity
(kg/lb)
Frame&
Platter
LoadCell
Material
HIBSX‐SS‐SS‐5KGSP6‐105/10StainlessStainless
HIBSX‐SS‐SS‐15KGSP6‐2015/30StainlessStainless
HIBSX‐SS‐SS‐40KGSP6‐5045/100StainlessStainless
HIBSX‐SS‐SS‐75KGSP6‐10075/165StainlessStainless
HIBSX‐SS‐AL‐5KGSPA22‐105/10StainlessAluminum
HIBSX‐SS‐AL‐15KGSPA22‐2015/30StainlessAluminum
HIBSX‐SS‐AL‐25KGSPA22‐3025/50StainlessAluminum
HIBSX‐SS‐AL‐35KGSPA22‐4035/75StainlessAluminum
HIBSX‐SS‐AL‐45KGSPA22‐5045/100StainlessAluminum

Chapter 3 - Installation
9
CHAPTER 3 - INSTALLATION
Precautions Do not store or operate the scale out of its
specified temperature range.
Do not store other equipment on the scale when
it is not in use or in storage.
Do not allow a build-up of debris on, around or
under the scale.
Do not set the scale in water or allow water to
settle around the scale. Provide Proper drainage.
Do not allow static or other electrical discharges
through the scale.
Do not try to repair the scale. Contact Hardy
Process Solutions Technical Support Department
for all repairs.
Unpacking Hardy Bench scales are shipped fully assembled
and only require adjustment of the overload
stops and adjustment of the leveling feet. If there
is any damage to the shipping container or to the
scale, save all the packaging material and file a
report with the shipper.
Adjusting the
Mechanical Overload
Stops
There is a bolt in each corner of the frame that is
fully engaged for shipping protection. Each
should be adjusted to allow the load cell to
deflect to its maximum deflection distance, but
no further.
Maximum Deflection at Capacity:
HISP6 - 0.17mm
HISPA22 - 0.4mm

11
Leveling the Scale Adjust the level of the scale by using the leveling
bubble and the adjustable feet.
Scale Connections The scale has a cable exiting the rear. Wire to the
preferred instrument by referring to the instrument
and cell wiring diagram.
Make sure the cable is not pinched and is clear of the
feet. The Bench Scale requires a recommended DC
or AC excitation voltage of 5 volts with a maximum of
15 volts. Wire the Scale cableto an instrument or
indicator following the color code label found on the
cable.
WARNING USE CAUTION WHEN WIRING. MAKE SURE
TO DISCONNECT ALL POWER TO THE
SCALE. FAILURE TO REMOVE POWER CAN
RESULT IN SCALE OR INSTRUMENT
DAM
AGE, DEGRADATION OF
PERFORMANCE OR
PERSONAL INJURY
NOTE: Drawings and wiring diagrams are available at
www.hardysolutions.com
Questo manuale è adatto per i seguenti modelli
2
Indice
Altri manuali Hardy Process Solutions Scala

Hardy Process Solutions
Hardy Process Solutions C Series Scheda tecnica

Hardy Process Solutions
Hardy Process Solutions HIFS-3030-01-PS-T Scheda tecnica

Hardy Process Solutions
Hardy Process Solutions 400 Series Scheda tecnica

Hardy Process Solutions
Hardy Process Solutions Enviro Series Manuale utente

Hardy Process Solutions
Hardy Process Solutions HI 1769-WS Scheda tecnica

Hardy Process Solutions
Hardy Process Solutions BW Series Manuale utente