Selectable parameters: 1 (default value), 2, 5, 10, 20, 50
F1.4 Calibration Unit
Selectable parameters: 0 ---- kg (default value)
1 ---- lb
F1.5 Gravitational Acceleration
Selectable parameters: 9.7000069.99999. Default value = 9.79455.
F1.6 Null-point Calibration
Keeping empty the scale
Remove the weights on the weighing platform to
guarantee the scale is in the empty state. Press the [Print]
key and the meter will display [ ]. The displayed
digits will reduce slowly until the meter displays
[]. In the end it will display [ ] for one
second, which indicates the end of null-point
calibration.
F1.7 Load-point Calibration
Loading weights
Load weights on the weighing platform to ensure that 10% of full-scale value ≤weight of weights
≤full-scale value, and then press the [Print] key to start the
next step.
Entering the same weight value as that of the loaded weights.
Entering the same weight value as that of the loaded weights,
please press the [Print] key after the scale becomes stable, and
the meter will display [ ]. After that, the displayed
digits will reduce slowly until the meter displays [ ]. In
the end it will display [ ] for one second, which
indicates the end of null-point calibration.
F1.8 Automatic Null Tracking
Selectable parameters: OFF, 1 d, 2 d, 3 d (default value)
F1.9 Automatic Reset Range at Startup
Selectable parameters: OFF, 2 %, 10 %, 20 % (default value)
F1.10 Button Reset Range
Selectable parameters: OFF, 2 %, 10 % (default value), 20 %
F1.11 Digital Filter
Selectable parameters: 0 ---- Mild Filtering