Lode Schiller CS200 Manuale dell'operatore

Interface sheet - Schiller CS200 - CPET Treadmill
Service documentation
Lode BV
Zernikepark 16
NL 9747 AN Groningen
The Netherlands
t: +31 50 571 28 11
f: +31 50 571 67 46
@: ask@lode.nl
www.lode.nl
INTERFACE SHEET - SCHILLER CS200 - CPET TREADMILL / VERSION: 1.0 - 2016-01-20

33
33
44
4
4
4
55
5
6
6
77
1212
1616
1818
18
Table of contentsTable of contents
1 - List of symbols used1 - List of symbols used
2 - Validity2 - Validity
3 - Setup instructions - Schiller CS2003 - Setup instructions - Schiller CS200
3.1 - Cable selection
3.2 - Software setup
3.3 - Ergometer setup
4 - Setting up the hardware4 - Setting up the hardware
4.1 - Connection overview
4.2 - Connecting with the usb cable
4.3 - Connecting with the RS232 cable
5 - Treadmill or ergometer with 3.5" touch screen unit5 - Treadmill or ergometer with 3.5" touch screen unit
6 - Treadmill or ergometer with 7" touch screen control unit6 - Treadmill or ergometer with 7" touch screen control unit
7 - Nummerical Control Unit - Treadmill only7 - Nummerical Control Unit - Treadmill only
8 - Treadmill or Bicycle without control unit8 - Treadmill or Bicycle without control unit
8.1 - Device without control unit

Next
Previous
1 - List of symbols used1 - List of symbols used
2 - Validity2 - Validity
This interface form covers
* All OEM treadmills manufactured by Lode B.V. with the part number 939900/01/02/07/08/09/10
starting from serial #939900-20090001;
* All Katana treadmills 9349XX-201X0001
* All Valiant treadmills starting with serial number 932900-20122XXX and 9329XX-20121XXX
* All Valiant 2 treadmill starting with serial number 9389xx-20140001
If your treadmill or ergometer is not covered by this interface form please contact Lode service
service@lode.nl
3/21INTERFACE SHEET - SCHILLER CS200 - CPET TREADMILL / VERSION: 1.0 - 2016-01-20

3 - Setup instructions - Schiller CS2003 - Setup instructions - Schiller CS200
3.1 - Cable selection3.1 - Cable selection
You will need the straight male-male RS-232 cable (part no. 930912) to
establish communication between the Lode ergometer and your stress
system.
Connect the RS232 cable to the Schiller AT-200 stress system. (see
picture)
3.2 - Software setup3.2 - Software setup
Switch on the ECG monitor.
Note: You must be logged in as “administrator” to configure the Schiller CS-200.
Select the “SETTINGS” menu and choose “SYSTEM CONFIGURATION”.
Select “ERGO DEVICES”.
Select Treadmill “TM 400/425 DIGITAL”
Press “OK”.
The system is now ready for communication with the Treadmill
3.3 - Ergometer setup3.3 - Ergometer setup
The following protocol needs to be set in the Lode ergometer to communicate with the external
equipment.
Treadmill ergometer: Trackmaster
The next chapters describe the procedures to setup these protocols depending on the type of
ergometer.
4/21INTERFACE SHEET - SCHILLER CS200 - CPET TREADMILL / VERSION: 1.0 - 2016-01-20

4 - 4 - Setting up the hardwareSetting up the hardware
Lode ergometers have two ways of connecting external devices. The cable selection depends on the
connection possibilities of the external device. The previous chapter tells you which cable you need. This
can be a RS232 cable or a USB cable.
4.1 - 4.1 - Connection overviewConnection overview
Connect the cable to the USB
port if you are using a USB
cable.
Please note that you may have
to install the FTDI driver on the
external equipment. This
procedure is only needed if
windows fails to instal the driver
automatically.
You can find the driver on the
Lode website under the Services
tab.
Locations:
- For the Treadmill the communication ports are Located on the front side near the power switch.
- For the bicycle ergometer the communication ports are located on the bottom of the bicycle. Tilt the
ergometer on it's side to be able to acces these ports.
Communication Module:
Use the single RS232 port to connect the communication module to the external equipment.
Legacy communication card (picture above -> 1 * USB, 2 * RS232)
In case you are using a RS232 cable you must use the RS232 port closest
to the USB connector, on the picture indicated with RS232 (1).
5/21INTERFACE SHEET - SCHILLER CS200 - CPET TREADMILL / VERSION: 1.0 - 2016-01-20

4.2 - Connecting with the usb cable4.2 - Connecting with the usb cable
If there is no other connection available you can also use a USB cable to
connect.
1. You will need an USB A to B cable (Lode part no. 930916) to establish
communication between the Lode ergometer and your stress system or
ergospirometrysystem. See picture left.
2. Connect the cable to the Lode ergometer.
3. Connect the cable to the external device.
4.3 - Connecting with the RS232 cable4.3 - Connecting with the RS232 cable
Connecting with the RS232 cable
1. Depending on the external device you may need to use a straight
RS232 cable (Lode part no. 930911) or a cross linked cable. The cable
selection chapter specifies which cable is needed.
2. Connect the cable to the Lode ergometer.
3. Connect the cable to the external device.
6/21INTERFACE SHEET - SCHILLER CS200 - CPET TREADMILL / VERSION: 1.0 - 2016-01-20

5 - Treadmill or ergometer with 3.5" touch screen unit5 - Treadmill or ergometer with 3.5" touch screen unit
Start the Lode ergometer.
During start up information
about the ergometer, like the
kind of device, the serial
number and the connected
modules is displayed.
After initialisation the terminal
mode is automatically
displayed. Swipe with your
finger over the screen as
indicated with the red arrow in
the picture to go to the main
menu.
7/21INTERFACE SHEET - SCHILLER CS200 - CPET TREADMILL / VERSION: 1.0 - 2016-01-20

Select "Settings" to change
settings of the device.
You can choose your Language
in the "Language" menu. The
menu "Sleep Mode" sets a
screensaver. The menu
"Terminal Settings" lets you set
the communication settings for
external communication. The
menu "Service" is only intended
for service purposes to be
executed by factory trainend
personel.
Select "Terminal Settings" to
change the communication
protocol.
8/21INTERFACE SHEET - SCHILLER CS200 - CPET TREADMILL / VERSION: 1.0 - 2016-01-20

Select the Communication
protocol as desribed in
"Ergometer setup" chapter.
Select "Default" for the
baudrate.
9/21INTERFACE SHEET - SCHILLER CS200 - CPET TREADMILL / VERSION: 1.0 - 2016-01-20

Select "RS232" or "USB"
depending in which cable you
have connected to the Lode
ergometer.
Select the button with the house
to go back to the main menu.
10/21INTERFACE SHEET - SCHILLER CS200 - CPET TREADMILL / VERSION: 1.0 - 2016-01-20
Indice
Altri manuali Lode Apparecchiature di registrazione
Manuali Apparecchiature di registrazione popolari di altre marche

Strymon
Strymon NIGHTSKY Manuale utente

Mitsubishi Electric
Mitsubishi Electric 16CH DIGITAL RECORDER DX-TL5000U Manuale utente

Tews Technologies
Tews Technologies TPMC465 Manuale utente

Honeywell
Honeywell Excel 50 Manuale utente

SeaLevel
SeaLevel COMM+8.LPCI Manuale utente

Arturia
Arturia AUDIOFUSE STUDIO Manuale utente