LSI LASTEM SVSKA2001 Manuale utente

LSI LASTEM SRL
Page 1 / 18
Environmental monitoring solutions
SVSKA2001
Data logger reprogramming kit
User manual

Data logger reprogramming kit –User manual
LSI LASTEM SRL
IST_03929
Page 2 / 18
Revisions list
Issue
Date
Description of changes
Origin
04/09/2020
1
17/09/2020
Change “Skip Flash Erase”option on pages 13 and 14
2
11/10/2021
Replaced pen drive and related references
3
20/07/2022
Replaced ST-Link utility with STM32 Cube Programmer; added unlock commands; made
minor changes
About this manual
The information contained in this manual may be changed without prior notification. No part of this manual
may be reproduced, neither electronically nor mechanically, under any circumstance, without the prior writ-
ten permission of LSI LASTEM.
LSI LASTEM reserves the right to carry out changes to this product without timely updating of this document.
Copyright 2020-2022 LSI LASTEM. All rights reserved.

Data logger reprogramming kit –User manual
LSI LASTEM SRL
IST_03929
Page 3 / 18
Table of contents
1 Introduction............................................................................................................................................... 4
2 Preparing the data logger for the connection........................................................................................... 5
3 Installing the programmer software and drivers on PC ............................................................................ 6
3.1 Getting started .................................................................................................................................. 6
3.1.1 System requirements ................................................................................................................ 6
3.1.2 Installing the STM32 Cube Programmer ................................................................................... 6
3.1.3 Installing ST-LINK, ST-LINK/V2, ST-LINK/V2-1 USB driver signed for Windows7, Windows8,
Windows10................................................................................................................................................ 9
3.2 Connection ST-LINK, ST-LINK/V2, ST-LINK/V2-1, ST-LINK/V3 to USB port ...................................... 10
3.3 Upgrade the firmware..................................................................................................................... 10
4 Connection to the data logger................................................................................................................. 12
5 Reprogramming data loggers .................................................................................................................. 13
5.1 Firmware upload ............................................................................................................................. 13
5.2 Programming bootloader ................................................................................................................ 15
6 How to unlock LSI LASTEM data loggers in case of locking..................................................................... 17
7 SVSKA2001 programming kit disconnection ........................................................................................... 18

Data logger reprogramming kit –User manual
LSI LASTEM SRL
IST_03929
Page 4 / 18
1Introduction
This manual explains how to install and use the SVSKA2001 kit for reprogramming the Alpha-Log and Pluvi-
One data loggers. Before proceeding with the use of this kit, try the LSI.UpdateDeployer software (see
IST_05055 manual).
The kit can also be used to unlock the data loggers in case of lock.
The USB pen drive contains:
•ST-LINK/V2 software and drivers
•STM32 Cube Programmer software
•firmware of LSI LASTEM data loggers
•this manual (IST_03929 Data logger reprogramming kit –User manual)
The procedure consists of:
➢installing the programming software and the ST-LINK/V2 programmer drivers on the PC
➢connecting the ST-LINK/V2 programmer to the PC and to the data logger
➢sending the firmware to the data logger or sending it the unlock commands in case of lock.

Data logger reprogramming kit –User manual
LSI LASTEM SRL
IST_03929
Page 5 / 18
2Preparing the data logger for the connection
The reprogramming or unlocking of the data logger takes place by means of the ST-LINK programmer. To
connect the programmer, it is necessary to remove the electronic boards of the data logger as described
below.
CAUTION! Before proceeding use an antistatic device (e.g. an antistatic wrist strap) to reduce, damp-
ens, inhibits electrostatic discharge; the buildup or discharge of static electricity, can damage elec-
trical components.
1. Remove the two caps and then unscrew the two fixing screws.
2. Remove terminal 1÷13 and 30÷32 from the terminal board. Then on the right side of the terminal
board, apply light pressure downwards and at the same time push towards the inside of the data
logger until the electronic boards and the display come out completely.

Data logger reprogramming kit –User manual
LSI LASTEM SRL
IST_03929
Page 6 / 18
3Installing the programmer software and drivers on PC
The STM32 Cube Programmer software facilitates fast in-system programming of the STM32 microcontrol-
lers during development via the ST-LINK, ST-LINK/V2 and ST-LINK-V3 tools.
Note: The part number of the STM32 Cube Programmer software is “SetupSTM32CubeProgram-
mer_win64.exe".
3.1 Getting started
This section describes the requirements and the procedures to install the STM32 Cube Programmer
(STM32CubeProg).
3.1.1 System requirements
The STM32CubeProg PC configuration requires as a minimum:
•PC with USB port and Intel® Pentium® processor running a 32-bit version of one of the
following Microsoft® operating systems:
oWindows® XP
oWindows® 7
oWindows® 10
•256 Mbytes of RAM
•30 Mbytes of hard disk space available
3.1.2 Installing the STM32 Cube Programmer
Follow these steps and the on-screen instructions to install the STM32 Cube Programmer (Stm32CubeProg):
1. Insert the LSI LASTEM pen drive on the PC.
2. Open the folder “STLINK-V2\en.stm32cubeprg-win64_v2-11-0”.
3. Double-click the executable SetupSTM32CubeProgrammer_win64.exe, to initiate the installation,
and follow the on-screen prompts (from fig. 1to fig. 13) to install the software in the development
environment.
fig. 1 –Preparing to install
fig. 2 - Click Next to continue

Data logger reprogramming kit –User manual
LSI LASTEM SRL
IST_03929
Page 7 / 18
fig. 3 –Click Next to continue
fig. 4 –Click on “I accept the terms…” and then
Next
fig. 5 - Click Next for the default directory
fig. 6 - Click OK to confirm the directory
fig. 7 - Click Next to continue
fig. 8 –Wait the installation progress

Data logger reprogramming kit –User manual
LSI LASTEM SRL
IST_03929
Page 8 / 18
fig. 9 - Click Next to continue
fig. 10 - Click Finish
fig. 11 - Click Next to continue
fig. 12 - Click Next to continue
fig. 13 - Click Done

Data logger reprogramming kit –User manual
LSI LASTEM SRL
IST_03929
Page 9 / 18
3.1.3 Installing ST-LINK, ST-LINK/V2, ST-LINK/V2-1 USB driver signed for Windows7, Win-
dows8, Windows10
This USB driver (STSW-LINK009) is for ST-LINK/V2, ST-LINK/V2-1 and ST-LINK/V3 boards and derivatives
(STM8/STM32 discovery boards, STM8/STM32 evaluation boards and STM32 Nucleo boards). It declares to
the system the USB interfaces possibly provided by the ST-LINK: ST Debug, Virtual COM port and ST Bridge
interfaces.
Attention! The driver must be installed prior to connecting the device, to have a successful enumeration.
Open the folder “STLINK-V2\Driver”of the LSI LASTEM pen drive and double-click the executable:
•dpinst_x86.exe (for 32-bit operating system)
•dpinst_amd64.exe (for 64-bit operating system)
To initiate the installation, follow the on-screen prompts (from fig. 14 to fig. 16) to install the drivers.
fig. 14 - Click Next
fig. 15 - Click Install
fig. 16 –Click Finish

Data logger reprogramming kit –User manual
LSI LASTEM SRL
IST_03929
Page 10 / 18
3.2 Connection ST-LINK, ST-LINK/V2, ST-LINK/V2-1, ST-LINK/V3 to USB port
Connect the USB cable:
•Micro-USB to ST-LINK/V2
•USB type-A to USB port PC
It will turn on red LED on the programmer:
3.3 Upgrade the firmware
1. Open and after few seconds will appear the main window
2. Proceed to upgrade the firmware as described from fig. 17 to fig. 20. The PC must be connected on