Lutron Electronics MCH-383SD Manuale utente

SD card real time datalogger
CO2/HUMIDITY/TEMP.
DATA RECORDER
Model : MCH-383SD
Your purchase of this CO2/HUMIDITY/TEMP. DATA
RECORDER marks a step forward for you into the field of
precision measurement. Although this DATA RECORDER is a
complex and delicate instrument, its durable structure
developed. Please read the following instructions carefully and
always keep this manual within easy reach.
OPERATION MANUAL

TABLE OF CONTENTS
1. FEATURES.........................................................................1
2. SPECIFICATIONS...............................................................2
3. FRONT PANEL DESCRIPTION.............................................
.
5
3-1 Display..............................................................................5
3-2 Lo
gg
er button, Enter button................................................5
3-3 button, Time button.......................................................▲5
3-4 button..........................................................................
.
▼5
3-5 SET button........................................................................ 5
3-6 Humidity/Temp. sensor......................................................
.
5
3-7 Han
g
in
g
holes ( main meter ).............................................
.
5
3-8 Stand................................................................................ 5
3-9 Battery cover/Battery compartment..................................... 5
3-10 Screw of the battery cover................................................ 5
3-11 Reset button....................................................................5
3-12 RS-232 output terminal.....................................................5
3-13 SD card socket................................................................. 5
3-14 DC 9V power adapter input Socket....................................
.
5
3-15 CO2 probe input socket....................................................
.
5
3-16 CO2 probe/sensor............................................................
.
5
3-17 Plu
g
of CO2 probe............................................................ 5
3-18 Han
g
in
g
hole ( CO2 probe )..............................................
.
5
3-19 Main meter han
g
in
g
unit ( with sticker ).............................5
3-20 CO2 han
g
in
g
unit ( with sticker )....................................... 5
4. MEASURING PROCEDURE..................................................
.
6
5. DATALOGGER....................................................................6
5-1 Preparation before execute datalo
gg
er function.................... 6
5-2 Datalo
gg
er........................................................................
.
8
5-3 Check time information.......................................................9
5-4 SD Card Data structure.......................................................9
6. Savin
g
data from the SD card to the computer....................
.
10
7. ADVANCED SETTING.........................................................
.
11
7-1 SD memory card Format.....................................................12
7-2 Set clock time....................................................................12
7-3 Set samplin
g
time..............................................................
.
13
7-4 Set beeper sound ON/OFF..................................................
.
13
7-5 Set SD card Decimal character............................................
.
14
7-6 Select the Temp. unit to or .........................................℃℉ 15
7-7 Set RS232 data output ON/OFF...........................................15
7-8 Set CO2 hei
g
ht ( meter ) compensation value.......................16
7-9 Set CO2 hei
g
ht ( feet ) compensation value.........................
.
16
8. POWER SUPPLY from DC ADAPTER....................................
.
16
9. CLOCK BACKUP BATTERY REPLACEMENT............................17
10. SYSTEM RESET................................................................17
11. RS232 PC serial interface..................................................18
12. ANNEX............................................................................
.
20
13. PATENT...........................................................................20

1. FEATURES
* Monitor with real time data logger, save the measuring
data along the time information ( year, month, date,
minute, second ) into the SD memory card and can be
down load to the Excel, extra software is no need. user
can make the further data or graphic analysis by
themselves.
* Show CO2 ( Carbon dioxide ), Humidity and
Temperature value in the same LCD.
*NDIR method principal for CO2 ( Carbon dioxide )
measurement, available for long term operation. high
repeatability and high accuracy.
* 0.1 %RH resolution for the humidity reading,
0.1 degree resolution for the Temp. reading.
*Used the precision capacitance type humidity sensor,
professional and high accuracy.
* SD card capacity : 1 GB to 16 GB.
* Large LCD display, easy readout.
* Microcomputer circuit, high reliability.
* Low power consumption and long battery life when
use battery power.
* Power supply from AC.DC 9V adapter in.
* RS232/USB computer interface.
*Patented
1

2. SPECIFICATIONS
2-1 General Specifications
Circuit Custom one-chip of microprocessor LSI
circuit.
Display LCD size : 60 mm x 50 mm
Measurement CO2 ( Carbon dioxide )
Humidity and Temperature
Unit CO2 ppm
Humidity %RH
Temp. ,℃℉
CO2 Sensor NDIR
Structure
Response Time
CO2 :
for CO2 < 2 min. typically.
measurement
* Reach the 63% reading value
* Depend the environment air
circulation.
Temperature Automatic temp. compensation for
Compensation CO2 measurement.
Datalogger 5/10/30/60/120/300/600 seconds
Sampling Time or Auto.
*
Default sampling time is 60 seconds.
*
The " Auto " sampling means when
the measuring value is changed ( > ±
1 %RH or > ± 1 or ± 50 ppm )℃
will save the data one time only.
Data error no. 0.1 % no. of total saved data typically.≦
Memory Card SD memory card. 1 GB to 16 GB.
Data Output RS 232/USB PC computer interface.
* Connect the optional RS232 cable
UPCB-02 will get the RS232 plug.
* Connect the optional USB cable
USB-01 will get the USB plug.
2

Advanced * SD memory card Format
setting * Set clock time ( Year/Month/Date, Hour/Minute/
Second )
* Set sampling time
* Set beeper sound ON/OFF
* Set SD card Decimal character
* Select the Temp. unit to or℃℉
* Set RS232 data output ON/OFF
* Set CO2 height ( meter ) compensation value
* Set CO2 height ( feet ) compensation value
Update Time Approx. 1 second if measuring data
of Display is changed.
Operating 0 to 50 .℃
Temperature
Operating Less than 90% R.H.
Humidity
Power Supply
*
DC 9V adapter input. ( AC/DC power
* Meter &
adapter is included ).
probe
Power Supply
*
A
Alkaline or heavy duty DC 1.5 V battery
* Clock backup
( UM4, AAA ) x 6 PCs, or equivalent.
battery
Weight Meter 240 g/0.53 LB.
CO2 probe 158 g/0.35 LB.
Dimension Meter 132 x 80 x 32 mm
( 5.2 x 3.1 x 1.3 inch )
CO2 probe 132 x 38 x 32 mm
( 5.2 x 1.5 x 1.3 inch )
3

Accessories * CO2 probe...................................
.
1 PC
Included * AC to DC 9V adapter.....................1 PC
* Main meter hanging unit ( with sticker )
...................................................1 PC
* CO2 hanging unit ( with sticker )...1 PC
* Instruction manual.......................
.
1 PC
Optional SD Card ( 2 GB ), SD-2GB.
Accessories USB cable, USB-01.
RS232 cable, UPCB-02.
Data Acquisition software, SW-U801-WIN.
2-2 Electrical Specifications (23±5 )℃
CO2 ( Carbon dioxide )
Range 0 to 4,000 ppm
Resolution 1 ppm
Accuracy ± 40 ppm
* 1,000 ppm.≦
± 5%ofreading
* > 1,000 ppm 3,000 ppm.≦
± 250 ppm typically
* > 3,000 ppm, reference only
Humidity
Range 10 % to 90 % R.H.
Resolution 0.1 % R.H.
Accuracy 70% RH :≧± ( 4 % reading + 1% RH).
< 70% RH : ± 4 %RH.
Temperature
Range 0 to 50 ,32 to 122 .℃℃℉ ℉
Resolution 0.1 degree
Accuracy - ± 0.8 ,℃℃- ± 1.5 .℉℉
@ Above specification tests under the environment RF Field Strength
less than 3 V/M & frequency less than 30 MHz only.
4

3. FRONT PANEL DESCRIPTION
Fig. 1
3-1 Display
3-2 Logger button, Enter button
3-3 button, Time button▲
3-4 button▼
3-5 SET button
3-6 Humidity/Temp. sensor
3-7 Hanging holes ( main meter )
3-8 Stand
3-9 Battery cover/Battery compartment
3-10 Screw of the battery cover
3-11 Reset button
3-12 RS-232 output terminal
3-13 SD card socket
3-14 DC 9V power adapter input socket
3-15 CO2 probe input socket
3-16 CO2 probe/sensor
3-17 Plug of CO2 probe.
3-18 Hanging hole ( CO2 probe )
3-19 Main meter hanging unit ( with sticker )
3-20 CO2 hanging unit ( with sticker )
5

4. MEASURING PROCEDURE
1)Plug the " Output plug " of " AC/DC 9V adapter " to
" DC 9V power input socket ( 3-14, Fig. 1 ), plug the
" Input plug " of the " AC/DC 9V adapter " into a properly
AC outlet ( AC 100 V to 240 V ) following.
2)The " Display " ( 3-1, Fig. 1 ) will show both the
CO2, Humidity and Temperature value at the same
time. Up Display - Humidity value
Middle Displa
y
- Temperature value
Low Display - CO2 value
3)Backup batteries consideration :
It recommend to install the batteries DC 1.5 V battery
( UM4, AAA ) x 6 PCs into the " Battery compartment "
( 3-9, Fig. 1 ) for clock time base backup. If do not
install those batteries, the internal clock may can not run
properly after the meter is disconnect from the " AC/DC
9V adapter " for a certain period.
Remark :
* How to change the Temp. unit ( , ), refer to chapter 7-6,℃℉
page 15.
5. DATALOGGER
5-1 Preparation before execute datalogger function
a. Insert the SD card
Prepare a " SD memory card " ( 1 GB to 16 GB, optional ),
insert the SD card into the " SD card socket " ( 3-13, Fig. 1)
with the correct direction exactly.
* It recommend use memory card's capacity is 4 GB.≦
6

b. SD card Format
If SD card just the first time use into the meter, it
recommend to make the " SD card Format " at first,
please refer chapter 7-1, page 12.
* It recommend strongly, do not use memory cards that have been
formatted by other meter or by other installation ( such as camera...)
Reformat the memory card with your meter.
* If the SD memory card exist the trouble during format by the meter,
use the Computer to reformat again can fix the problem.
c. Time setting
If the meter is used at first time, it should to adjust the
clock time exactly, please refer chapter 7-2, page 12.
d. Decimal format setting
The numerical data structure of SD card is
default used the " . " as the decimal, for
example "20.6" "1000.53" . But in certain
countries ( Europe ...) is used the " , " as the
decimal point, for example " 20, 6 "
"1000,53". Under such situation, it should
change the Decimal character at first, details
of setting the Decimal point, refer to Chapter
7-5, page 14.
e. 3 Information of LCD display
* If the Display show :
It means that the SD card exist the problem
CH- or the SD card memory is full, it should
CArd change SD memory card.
* If the Display show :
It means that the SD card is not plugged
no into the meter.
CArd
7

5-2 Datalogger
a. Start the datalogger
Press the " Logger button ( 3-2, Fig. 1 ) > 2 seconds
continuously, until the Display show the indicator
" DATALOGGER ", release the " Logger Button " ( 3-2, Fig.
1 ), then the measuring data along the time information
will be saved into the memory circuit.
Remark :
*
How to set the sampling time, refer to Chapter 7-3,
page 13.
*
How to set the beeper sound is enable, refer to
Chapter 7-4, page 13.
c.. Finish the Datalogger
During execute the Datalogger function ( Display show the
" Datalogger " indicator ), press the " Logger button " ( 3-2,
Fig. 1 ) > 2 seconds continuously, until the Display
indicator " DATALOGGER " is disappeared, release the
" Logger Button " will finish the Datalogger function.
Before take away the SD card from the
meter, it should execute the
procedures of " Finish the Datalogger ",
otherwise some existing recent save
data may loss.