Me GSV-4 Manuale utente

4 channel measuring amplifier GSV-4
GSV-4BT
GSV-4USB
Operating instructions
Updated: 02.02.20 8
ME-Meßsysteme GmbH
676 Hennigsdorf Fax: +49 3302 89824 69 Web: www.me-systeme.de


Contents
Inhaltsverzeichnis
Straingage measuring amplifier GSV-4USB.......................................................................................4
Description..........................................................................................................................................4
Dimensions.........................................................................................................................................5
Technical data....................................................................................................................................6
Connection assignment....................................................................................................................7
Strain Gage measuring amplifier GSV-4BT...................................................................................... 8
Description........................................................................................................................................ 8
Dimensions....................................................................................................................................... 9
Technical data..................................................................................................................................20
Wiring diagram.................................................................................................................................2
Technical data GSV-4BT M 2.......................................................................................................25
Switch configuration GSV-4BT M 2............................................................................................26
Measurement resolution................................................................................................................28
Order variants GSV-4BT.................................................................................................................29
Programming / configuration............................................................................................................30
Scaling of measured values..........................................................................................................30
Measuring range 2.0 mV/V............................................................................................................30
Measuring range 0.0 mV/V.........................................................................................................30
Measuring range 0.0 to 5 V............................................................................................................3
Measuring range 0.0 to 0 V.........................................................................................................3
Measuring range PT 000...............................................................................................................3
Measuring range K thermocouple cable.....................................................................................32
Commands for configuration........................................................................................................32
List of commands...........................................................................................................................32
Description of commands.............................................................................................................34
Protocol for measured values.......................................................................................................37
Protocol for commands.................................................................................................................37
Protocol for responding to commands.......................................................................................37
Digital IOs..........................................................................................................................................39
Analogue input.................................................................................................................................43
CAN bus.................................................................................................................................................44
Protocol for measured values.......................................................................................................44
Protocol for commands.................................................................................................................44
Protocol for responding to commands.......................................................................................44
Configuring the CAN-ID..................................................................................................................45
Data frequency and filter................................................................................................................46
Analogue filter..................................................................................................................................46
Digital filter........................................................................................................................................46
Annex......................................................................................................................................................48
Connection figures for GSV-4BT SD and GSV-4BT LD.............................................................48
Changelog.........................................................................................................................................52
ME-Meßsysteme GmbH
676 Hennigsdorf Fax: +49 3302 89824 69 Web: www.me-systeme.de 3

Straingage measuring amplifier GSV-4USB
GSV-4USB M 2
Front view sensor connection
GSV-4USB SUB-D37
Front view sensor connection
Description
This 4-channel measuring amplifier for sensors with strain gauges is equipped with a USB
interface. The voltage is supplied via the USB port at the back of the measuring amplifier.
The measuring amplifier can be delivered with an SUB-D37 connection or with 4x M 2 ports.
The measuring amplifier has eight digital inputs and outputs.
On the backside SubD25 socket, strain gauge full-bridges and half-bridges 20 Ohm up to
kOhm as well as PT 000 temperature sensors and 000 Ohm single grid strain gages or
voltages 0 ... 5V can be connected.
The front-end M 2 socket is configured by default for strain gauge full-bridge connections
and for voltage inputs 0 ... 5V and 0 ... 0V.
ME-Meßsysteme GmbH
4 676 Hennigsdorf Fax: +49 3302 89824 69 Web: www.me-systeme.de
•4-channels
•Power supply via USB port
•Inputs for Straingage / 0– 0 V / PT 000
•Measurement ranges 2 mV/V / 0 mV/V
•Straingage quarter / half / full bridges
•8 digital inputs / outputs
•Data rate 0 Hz–500 Hz

Dimensions
Figure 2: Dimensions GSV-4USB M12
ME-Meßsysteme GmbH
676 Hennigsdorf Fax: +49 3302 89824 69 Web: www.me-systeme.de 5
Figure 1: Dimensions GSV-4USB SUB-D37

Technical data
Accuracy class 0.05 %
Inputs
Resolution 6 Bit
Straingage inputs
Full bridge
Half bridge
Quarter bridge
89-5000
89-5000
20 / 350 / 000
Ohm
Ohm
Ohm
Common mode rejection
at 60 Hz common-mode signal 95– 0 dB
Measurement frequencies
Data frequency 0 – 500 Hz
Sampling frequency .92 MHz
Cut-off frequency
analogue
digital
450
Notch filter
Hz
Hz
Outputs
Bridge supply voltage
Current load capacity
2.5
30
Volt
mA
Fixed voltage output
Current load capacity
5
20
V
mA
Switching outputs/inputs
I/O -8
Current load capacity:
TTL level
5 (active High)
5
V
mA
Interface USB . , USB 2.0 compatible
Supply voltage
Nominal range
Isolation voltage
Current consumption
4.5...5.5 via USB port
000
< 200
V DC
Vrms
mA
Temperature range
Nominal temperature range
Storage temperature range
Zero point drift
Sensitivity drift
- 0…+65
-40…+85
< 0.05
< 0.0
°C
°C
%/ 0°C
%/ 0°C
Dimensions
L x W x H 06 x 9 ( 32) x 32 mm x mm x mm
Protection type / Weight
Protection type IP40
Weight GSV-4USB SUB-D37 239 g
Table : Technical data GSV-4USB
ME-Meßsysteme GmbH
6 676 Hennigsdorf Fax: +49 3302 89824 69 Web: www.me-systeme.de

Connection assignment
Connection plan for 37-pin D-sub port
GSV-4USB assignment 37-pin D-sub port (PIN No.)
Channel Channel 2 Channel 3 Channel 4
+USpositive sensor supply 20 2 29
+UDpositive differential input 22 4 3 3
QB 000 quarter bridge extension kOhm 23 5 4 32
HB half bridge extension 24 6 5 33
-UDnegative differential input 25 7 6 34
-USnegative sensor supply 27 9 8 36
UEanalogue input 28 0 9 37
screen
Table 2: Analogue inputs
Connection assignment for GSV-4USB 12
5-pin port M 2x , type 763
Top view:
5-pin Description Colour code for cables
2 -US negative bridge supply white
+US positive bridge supply brown
3 +UD positive differential input blue
4 -UD negative differential input black
5 AUXin input without cable grey
ME-Meßsysteme GmbH
676 Hennigsdorf Fax: +49 3302 89824 69 Web: www.me-systeme.de 7
1
19
37 20
12
4 3
5

Connection of full bridge with SUB-D37 version
The following graphics show the connection of a full bridge to channel through to channel
4.
ME-Meßsysteme GmbH
8 676 Hennigsdorf Fax: +49 3302 89824 69 Web: www.me-systeme.de

Connection of half bridge with SUB-D37 version
The following graphics show the connection of a half bridge to channel through to channel
4.
The bridge extension should be adapted depending on the application.
ME-Meßsysteme GmbH
676 Hennigsdorf Fax: +49 3302 89824 69 Web: www.me-systeme.de 9

Connection of quarter bridge with SUB-D37 version
The following graphics show the connection of a quarter bridge to channel through to
channel 4.
The bridge extension should be adapted depending on the application.
ME-Meßsysteme GmbH
0 676 Hennigsdorf Fax: +49 3302 89824 69 Web: www.me-systeme.de
Altri manuali per GSV-4
1
Questo manuale è adatto per i seguenti modelli
4
Indice
Altri manuali Me Amplificatore