1
I. Introduction
A. InstrumentFunctionOverview
TheMS3500E/3510EDataLoggerisdesignedto provide a continuous record ofcorrosiveactivityin
remotelocationsthatareinfrequentlyvisited, such ascross-countrypipelines,andunmannedproduction
platforms.However,thisunitfindsequalapplicationinlocationsthatareinconvenientordifficultto
accessona regular basis, suchasrefineryoverheads,andflue-gasstacks.
TheMS3500E/3510Eiscompletelyself-containedwiththeonboardbatterysystemsupplyingthetotal
powerrequirementfor operation. It canbeusedwithanyof Metal Samples’ ERprobes,orthoseofall
othermajormanufacturers.Onceinplace,itwillautomaticallyreadtheprobeatcustomer-selected
intervalsandstoretheresultingdataintheunit’sonboardmemory.Theonboardmemorywillcollectup
to3,100datapointsbeforedata download is required. The datacollectionintervalisprogrammablein
hourlyincrements.Evenattheminimumdatacollectioninterval(1hour),datadownloadingneedonlybe
performedeveryninetydays.Withmore practical data collection intervals (8 - 12 hours), the unit may
beunattendedfor as longas6-8months between downloading operations.
Anoptical,infrared,RS232communicationlink is provided for datadownloadtoeitheraPCorMetal
Samples'MS1500E portable ERdatalogger.Downloadeddata may beanalyzed,reviewed,or re-
portedbyconventionalspreadsheet,database, or mathematical software packages.
Theinfraredcommunicationlink isanintegralpartoftheintrinsicallysafedesignoftheMS3500E/
3510E.Opticalinsteadofelectrical,thisuniquefeaturepermitsdatadownloadingwithoutremovingthe
instrumentor“memorymodule” fromthehazardousarea.
AnotheruniquefeatureoftheMS3500E/3510Eis the highlevelofonboardintelligence.Thetwo-line,
20-characterLCDallowsvisualreviewofallhistoricaldatainmemoryandreads directly as “total metal
loss”and/or“averagecorrosionrate”. Thisisincontrasttoothercommerciallyavailableunits that store
anddisplayonlytheraw “resistanceratio”thatrequiresadditionalmanipulationtoprovideusefuland
intelligibledata.
TheLCDscreen,togetherwiththe2-keymembranekeypad,provides a user-friendly,interactive,
promptingsystemthatis used for bothsystemsetupanddatareview.ThismakestheMS3500Ethe
mostadvancedunitofits type on the market.
Byaddinga4-20mAcommunication facility, the MS3500E becomes an MS3510E.Thisallows
continuoustransmissionofdata, via a 4-20mAloop,to a plant computer orcentraldatalogger, for
integrationwithotherreal-timeprocessparameters.This data transmissioncanbeaccomplishedwithout
disruptionoftheunit’s basic logginganddatastorageoperations.The 4-20mAloopextends the capa-
bilitiesoftheunittoincludeconventional,in-plant,real-timedatacommunication.
TheMS3500E/3510Eunitusesastainlesssteel NEMA 4X (IP 66) enclosure, makingitsuitableforuse
inthemostextremeofoutdoorconditions.