6 Device LED Indicators
LED Pattern Red Blue Status
1Blinks for 20ms
duty is set by
server
Heart beat check-in
message for location
status
OFF
2Blinks for 1s
when Pendant
button is pressed
OFF Pendant sending
Alarm message to
Repeater
3Blinks for 20ms
after Pendant
button is pressed
Pendant has
received check-in
aknowledgement
from Repeater
OFF
4OFF Blinks for 50ms
after Pendant
button is pressed
Pendant has
received Alarm
aknowledgement
from Repeater
5Blinks for 2ms
every 3sec when
Pendant in Alarm
mode
OFF Pendant is in
Alarm mode and
waiting for staff to
cancel at source
7
6Blinks for 2ms
duty cycle set by
server
Pendant sends
Repeater Position
Wake status
Blinks for 2ms
duty cycle set by
server
Pendant enters
storage mode and
is in ultra low
battery mode
Pendant enters storage mode it
will flash Red 1sec - Blue 1sec -
Red 1sec
7 Specifications
Dimensions: 36mm x 41mm x 12mm.
Weight: 16g with battery inserted.
Operating temperature: 32 to 140°F (0 to 60°C).
Operating environment: water resistant to IP65 standard.
Estimated battery life: One year (assuming five alarm activations per day
and alarm clearance within ten minutes of alarm activation).
Battery type: Panasonic CR2450.
Power requirement: 3VDC, 15µA.
Storage requirement: Ambient conditions of 69 to 73°F (20.5 to 22.8°C).
UL certification for UL 2560
FCC RF Radiation Exposure Statement
This equipment complies with FCC radiation exposure limits set forth for an
uncontrolled environment. End users must follow the specific operating
instructions for satisfying RF exposure compliance. This transmitter must
not be co-located or operating in conjunction with any other antenna or
transmitter except in accordance with FCC multi-transmitter product
procedures.
FCC Part 15 and Industry Canada Compliance
This device complies with part 15 of the FCC Rules and Industry Canada
license-exempt RSS standard(s). Operation is subject to the following two
conditions: (1) this device may not cause interference, and (2) this device
must accept any interference, including interference that may cause
undesired operation of the device. Changes or modifications not expressly
approved by the party responsible for compliance could void the user's
authority to operate the equipment.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables
aux appareils radio exempts de licence. L'exploitation est autorisée aux
deux conditions suivantes : (1) l'appareil ne doit pas produire de brouillage,
et (2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique
subi, même si le brouillage est susceptible d'en compromettre le
fonctionnement.
Battery Handling
If you are installing CR2450 lithium metal coin cell batteries in multiple
devices, keep the batteries from coming into contact with each other to
prevent discharge.
Pendant Shipping
Both lithium-ion and lithium-metal battery types are subject to
transportation regulations such as International Air Transport Association
(IATA), International Maritime Dangerous Goods (IMDG), US Department
of Transportation (DOT) and UN Recommendation on the Transport of
Dangerous Goods.
To prevent accidental alarm activation due to air pressure changes,
miAlert recommends that you avoid shipping the devices by air with the
CR2450 lithium-metal coin cell battery installed in the device.
When disposing of this electronic device or depleted CR2450 lithium-metal
coin cell batteries, please do so in accordance with federal, state and local
regulations.
Note: Changes or modifications not expressly approved by the party
responsible for compliance could void the user's authority to operate the
equipment.
7.a Medical Device Interference
Under FDA and FCC guidelines, the responsibility for verifying continuous
and safe operation of medical devices such as pacemakers and implanted
defibrillators in the presence of electromagnetic interference (EMI) rests
with the manufacturer of the medical device. As such, miAlert does not
have the authority or specific device knowledge to conduct or interpret
formal tests on their behalf.
miAlert transmitter devices comply with Part 15 of the FCC Rules and
Industry Canada license-exempt RSS standards.
If you have any concerns about the use of miAlert transmitter devices in
the presence of medical devices used by a particular resident or patient, we
recommend that you consult with their physician. Another potential source
for answers is the medical device manufacturer, who can provide more
information as to their compliance with federal guidelines and how they
have addressed EMI risk.
7.b Radiation Exposure Limits
This equipment complies with ISED RSS-102 radiation exposure limits set
forth for an uncontrolled environment. End users must follow the specific
operating instructions for satisfying RF exposure compliance. This
transmitter must not be co-located or operating in conjunction with any
other antenna or transmitter.
Cet équipement est conforme avec ISED RSS-102 des limites d'exposition
aux rayonnements définies pour un environnement non contrôlé. Les
utilisateurs finaux doivent suivre le fonctionnement spécifiqueinstructions
pour satisfaire la conformité à l'exposition RF. Cet émetteur doitne pas être
colocalisé ou fonctionner conjointement avec une autre antenne ou
émetteur.
Note: IP65 certification was performed by an independent laboratory not
affiliated with UL.
Note: Users that have achieved certification and will install UL 2560
certified systems are responsible for labeling all fundamental devices with
the UL 2560 system certification mark. Use compatible Repeater and
appropriate Gateway for UL 2560 installations.
Note: miAlert supports recycling and reuse whenever possible. Please
recycle these parts using a certified electronics recycler.
Note: For UL 2560 installations, miAlert repeaters must have 20 minute
check-in times. miAlert transmitters must have a minimum of 60 minute
check-in times.
Note: In a UL 2560 installation the miAlert Pendants must be used to
complete emergency call systems for assisted living and independent
living facilities.
For UL 2560 certified system installations, the following miAlert devices
are approved for installation within maximum system configuration limits
defined in section 1.b of this document: