Miris Ultrasonic Manuale utente

Miris Ultrasonic Processor
USER MANUAL
English


Miris Ultrasonic Processor User Manual | 3
READ THE MANUAL BEFORE USING THE INSTRUMENT
PREFACE
Thank you for choosing the Miris Ultrasonic Processor.
Please read this manual carefully before starting to use the
instrument.
Miris Ultrasonic Processor is used to prepare milk samples for
macronutrient analysis using Miris HMA™ - Human Milk Analyzer.
By the application of high-frequency ultrasonic waves, the milk
sample will be homogenised in a simple and convenient manner.
© Copyright 2018 Miris AB, all rights reserved. This manual is
protected by copyright. Distribution and copying in any form of
the content is prohibited without the explicit and written consent
of Miris AB.
name
ID
revision
issue date
article number
name
article number
standard
Miris Ultrasonic Processor User Manual
M&S003-22a
2
2018-04-03
36-06-201
Miris Ultrasonic Processor
08-04-201
2014/30/EU Electromagnetic
Compatibility Directive
2014/35/EU Low Voltage Directive
2011/65/EU RoHS 2 Directive
Document information
Product information
address
web
e-mail
telephone
MIRIS AB
Kungsgatan 115
753 18 Uppsala
Sweden
www.MirisSolutions.com
+46 18 14 69 07
Contact information
address SONICS & MATERIALS INC.
53 Church Hill Road
Newtown, CT 06470
USA
Manufacturer information

4 | Miris Ultrasonic Processor User Manual
TABLE OF CONTENTS
PREFACE.......................................................................................................................................... 3
TABLE OF CONTENT....................................................................................................................... 4
IMPORTANT USER INFORMATION ............................................................................................... 5
INTENDED USE OF THE MIRIS ULTRASONIC PROCESSOR ................................................... 5
WARRANTIES AND DUTIES ..................................................................................................... 5
SAFETY INFORMATION ........................................................................................................... 6
MIRIS ULTRASONIC PROCESSOR USER GUIDE........................................................................... 7
Chapter 1 INSTALLATION ........................................................................................................ 7
Chapter 2 SETTINGS ................................................................................................................ 9
Chapter 3 OPERATING THE MIRIS ULTRASONIC PROCESSOR............................................. 10
Chapter 4 CLEANING ROUTINES............................................................................................. 12
Chapter 5 TROUBLESHOOTING AND PROBLEM SOLVING.................................................... 14
ABOUT MIRIS ULTRASONIC PROCESSOR ................................................................................... 15
TECHNICAL SPECIFICATIONS................................................................................................. 15
WORKING PRINCIPLES OF THE INSTRUMENT ...................................................................... 15

Miris Ultrasonic Processor User Manual | 5
IMPORTANT USER INFORMATION
WARRANTIES AND DUTIES
By operating this Miris Ultrasonic Processor, the USER and MIRIS agree to the following
responsibilities, which constitute contractual warranties and conditions between MIRIS and
USER for the maximum benet and usefulness of the Miris Ultrasonic Processor.
MIRIS AB WARRANTS THAT IT:
• Knows of no defects in the construction of the Miris Ultrasonic Processor or its
materials used
• Will replace or repair the Miris Ultrasonic Processor according to the guarantee in the
product warranty
USER WARRANTS THAT:
• The Miris Ultrasonic Processor will be used according to the instructions given in the
user manual
• The Miris Ultrasonic Processor will not be altered without written approval of MIRIS AB
• MIRIS will be notied immediately if any injury occurs in any association with the Miris
Ultrasonic Processor, and will be allowed prompt and thorough examination of the Miris
Ultrasonic Processor in question
• MIRIS will not be held responsible in cases of injury arising from use of the Miris
Ultrasonic Processor:
a. When the Miris Ultrasonic Processor is not used according to the
instructions in this manual
b. When MIRIS is not notied within 5 days of said injury
INTENDED USE OF THE MIRIS ULTRASONIC PROCESSOR
Miris Ultrasonic Processor reduces the particle size distribution in milk. This device is intended for use in
laboratories by laboratory personnel.

6 | Miris Ultrasonic Processor User Manual
SAFETY INFORMATION
When mounting the probe, always clamp the converter housing.
Never clamp the probe.
Make sure the Ultrasonic Processor is properly grounded via a
3-prong outlet.
High voltage is present in the power supply. Do not remove the
cover. Service can only be performed by an authorized service
technician. Ask your local distributor or contact Miris.
Never operate the power supply unless it is connected to the
converter.
Never secure anything to the probe.
Never touch a vibrating probe.
Never allow a probe to vibrate in air for more than 10 seconds.
Always use input voltage 110/240 V , 50/60 Hz.

Miris Ultrasonic Processor User Manual INSTALLATION | 7
MIRIS ULTRASONIC PROCESSOR USER GUIDE
Chapter 1 INSTALLATION
1. Sound reducing cabinet
2. Display and keypad
Figure 1. The Miris Ultrasonic Processor. 1
2

8 | INSTALLATION Miris Ultrasonic Processor User Manual
THE INSTRUMENT
ELECTRICAL REQUIREMENTS
The Miris Ultrasonic Processor requires a fused, single phase 3-terminal grounding type elec-
trical outlet capable of supplying 50/60 Hz at 100 Volts, 115 Volts, 220 Volts, or 240 Volts,
depending on the voltage option selected. For power requirements, check the label on the back
of the unit.
Probe processing capacity
For efcient homogenisation, it is important to not exceed the processing capacity of the probe.
The 3 mm diameter probe is for milk volumes 3-10 ml.
3. Height adjustment knob
4. Convertor housing
5. Probe
6. Display with current settings
7. Probe output power indicator
8. Digit input 1-9
9. Clear
10. Enter/Review
11. Start/Stop
12. On
13. Off
14. Timer
15. Pulse
16. Amplitude
Figure 2. Inside the Sound reducing cabinet.
Figure 3. Display and keypad.
3
4
5
6
7
89 10
11 12
1314
15
16

Miris Ultrasonic Processor User Manual SETTINGS | 9
MIRIS ULTRASONIC PROCESSOR USER GUIDE
Chapter 2 SETTINGS
TIME
Processing time for human milk samples is 1.5 s/ml milk.
Calculate sample processing time in seconds by multiplying
sample volume in ml by 1.5
Set sample processing time by pressing [Timer]
Use the key pad to enter sample processing time given as hours,
minutes, seconds, in the format H:MM:SS
Press [Enter] to conrm
The set time is displayed
PULSE
Pulsating mode, Pulse, should be turned off.
Press [Pulse]
Press [Clear] to erase current setting
Press [Enter] to conrm
Pulse is now turned off, shown on the display as --
AMPLITUDE
Amplitude should be set to 75%.
Press [Ampl]
Use the key pad to enter 075
Press [Enter] to conrm
Amplitude is now set to 75%, shown on the display

10 | OPERATING THE MIRIS ULTRASONIC PROCESSOR Miris Ultrasonic Processor User Manual
MIRIS ULTRASONIC PROCESSOR USER GUIDE
Chapter 3 OPERATING THE MIRIS ULTRASONIC PROCESSOR
This chapter describes how to use the Miris Ultrasonic Processor.
See Chapter 1 for start-up procedure and Chapter 2 for settings.
For information on use and handling of human milk and the Miris HMA™,
refer to instructions in the Miris HMA™ User Manual.
STANDARD OPERATING PROCEDURE
Equipment Consumables
Miris Ultrasonic Processor Miris Cleaner™
Emery cloth
Sample container
Table 1. Standard Operating Procedure (SOP) for processing milk using Miris Ultrasonic Processor.
Set-up Prepare working solution of Miris Cleaner™
Turn on the Miris Ultrasonic Processor by pressing [I]
Instrument preparation
See Chapter 2 and Chapter 4
Check that amplitude is set to 75% and pulsating
mode is turned off
Set the processing time (1.5 s/ml) according to
sample volume
Make sure that the Miris Ultrasonic Processor probe is clean and that the tip is evenly polished
Sample processing (For instructions on human milk preparation and handling, see the Miris HMA™ User Manual)
Use a sample container which will give a sample depth of at least 2 cm, see Figure 4, allowing some extra
space for sample movement
If using conical tubes, a tube holder i necessary to keep the tube upright and stable. If using a at bottom or
skirted tube, a tube holder or tube rack may not be necessary.
Place the container with the sample in the Miris Ultrasonic Processor Sound reducing cabinet
Adjust the probe height using the Height adjustment knob
Immerse the probe into the sample
Note! If the probe is lowered to an insufcient depth the sample will foam and/or splash, which should be
avoided
Centre the probe in the container
Close the door of the Sound reducing cabinet
Press Start [Start/Stop]
The processing can be stopped at any time by pressing [Start/Stop] again
After the processing is nished, remove the sample
(The sample is now ready for analysis with Miris HMA™ Calibration 1 – Homogenized milk)
Human milk (ml) Time (s)
3 5
5 8
10 15
Indice
Manuali Hardware per computer popolari di altre marche

EMC2
EMC2 VNX Series Manuale del proprietario

Panasonic
Panasonic DV0PM20105 Manuale utente

Mitsubishi Electric
Mitsubishi Electric Q81BD-J61BT11 Manuale utente

Gigabyte
Gigabyte B660M DS3H AX DDR4 Manuale utente

Raidon
Raidon iT2300 Manuale utente

National Instruments
National Instruments PXI-8186 Manuale utente