MRC CWN-1.6KW Istruzioni operative

User Operating Manual
Recirculating Coolers, Chiller
Original Instructions
3, Hagavish st. Israel 58817 Tel: 972 35595252, Fax: 972 3 5594529
MRC. 9.22

Congratulations!
Thanks for choosing our product.
We are confident it will serve you for a long time!
First of all, we thanks for the trust to us. This operating manual has been
designed to help you understanding the principles of operation and possibilities
of chillers. For optimum utilization of all functions, we recommend that study this
manual before operation.

Table of Contents
Section I General Information.................................................................................... 1
1.1 Safety Warnings................................................................................................. 1
1.2 System Introduction........................................................................................... 1
1.3 Symbol Instructions............................................................................................2
1.4 Warning Labels...................................................................................................2
Section II Transportation & Installation...................................................................5
2.1 Transportation Requirements...........................................................................5
2.2 Unpacking............................................................................................................5
2.3 Package Contents..............................................................................................5
2.4 Installation Instruction........................................................................................5
Section III Setting Guide.............................................................................................. 8
3.1 Preparations before operation..........................................................................8
3.2 Controller Panel................................................................................................10
3.3 Advanced Application Mode...........................................................................12
3.4 Common Alarms.............................................................................................. 14
3.5 Hardware interface (Optional)........................................................................19
Section IV Operation guide.......................................................................................19
4.1 Switching on...................................................................................................... 20
4.2 Temperature Setting........................................................................................ 21
4.2.3 Tracking control mode.................................................................................. 23
4.3 Start/Stop compressor..................................................................................... 23
4.4 Start/Stop pump................................................................................................ 24
4.5 PID Adjustment................................................................................................. 24
4.6 Switch off............................................................................................................25
Section V Trouble Shooting and Maintenance.................................................... 26
5.1 Trouble Shooting..............................................................................................26
5.2 Daily Maintenance........................................................................................... 27
Appendix I Parameter Database........................................................................... 28
Appendix II Warning Information......................................................................... 33

Recirculating cooler(Chiller)
1
Section I General Information
1.1 Safety Warnings
1) Make sure chiller is operated only by qualified persons.
2) Make sure you have read and understood all instructions and safety
precautions listed in this manual before installing or operating your chiller. If
you have any questions, contact the supplier.
3) Performance of installation, operation, or maintenance procedures other than
those described in this manual may result in a hazardous situation and may
void the manufacturer's warranty.
4) Transport the chiller with carefully. Sudden jolts or drops may cause internal
damages.
5) Observe all warning labels.
6) Never remove warning labels.
7) Never operate damaged or leaking equipment.
8) Never operate the chiller without bath fluid in the bath.
9) Always turn off the chiller and disconnect the mains cable from the power
supply before performing any service or maintenance procedures, or before
moving the chiller.
10) Always empty the bath before moving the chiller.
11) Never operate the chiller with damaged mains power cables.
12) Refer service and repairs to a qualified technician.
13) As the upgrade of productions, the operating manual may have some
differences from the newest production you buy, please contact the
on-the-spot service people or local distributor at the first time. In your
afterwards operation, please contact us in time if there is any question, we will
serve you sincerely.
1.2 System Introduction
The recirculating cooler, also called as chiller, draws heat from bath liquid via the
cooling system for instrument or equipment that need to be cooled (Water is a
common bath liquid, other liquid, such as glycol is used for lower temperature) to
ensure instrument or equipment to work properly under the normal temperature.

Recirculating cooler(Chiller)
2
Circulation is looped between the chiller and the instrument depending on the
pump’s pressure inside the chiller, and the temperature sensor detects the
medium temperature to realize the control of the cooler.
The recirculating cooler is applied vastly on laboratory test equipment such as
rotary evaporators, X-ray diffractometer, ICP emission spectrometer, ICP-MS,
graphite furnace AAS, ED-XRF, electron microscope, medical CT, medical
physiotherapy chiller, production test chiller, precision grinder, computer
numerical control machine, machining center, laser equipment etc.
This chiller is an integral product with parts such as condenser, compressor,
evaporator and pump inside the same case.
1.3 Symbol Instructions
Very dangerous!Denotes an imminent hazardous situation
that will result in death or serious injury.
Denotes a generally hazardous situation that can result in
death or serious injury.
Denotes a hazardous situation that can result in injury.
Denotes a situation that can result in property material
d damage.
1.4 Warning Labels
Don't operate
Authorized people only.

Recirculating cooler(Chiller)
3
Electric Shock
This device only operated by professionals.
WARNING
HIGH TEMP
WARNING
HANDS PINCHING
Hazardous moving parts
Keep fingers and other Body parts away

Recirculating cooler(Chiller)
4
Emergency stop button should be included on the
customers equipment which chillers supporting for
to ensure the operator's safety.

Recirculating cooler(Chiller)
5
Section II Transportation & Installation
2.1 Transportation Requirements
Please take appropriate safety measures for the transportation of the goods and
shall pack the goods according to the general way.
1) The cargo shall be fixed and protected against shock and violent vibration.
2) Necessary measures should be taken to prevent moisture, rain and theft.
3) Inverted, inclined and horizontal are prohibited.
2.2 Unpacking
Your chiller is shipped in a special carton. Retain the carton and all packing
materials until the unit is completely assembled and working properly. Set up and
run the unit immediately to confirm proper operation. Beyond one week, your unit
may be warranty repaired, but not replaced. If the unit is damaged or does not
operate properly, contact the transportation company, file a damage claim, then
contact the company where your unit was purchased.
2.3 Package Contents
1) Chiller
2) User Operating Manual
3) Power Core
4) Certificate of Quality and Packing List
5) Standard Output and Accessories
2.4 Installation Instruction
The unit shall be installed indoor with good environment. The principles are as
following:

Recirculating cooler(Chiller)
6
Location Principle
1
Firm ground or experiment table in a dust-free and cool environment with
good ventilation
2
Far from droplets or steam
3
Far from the location that influx, viscous flow, leak of fuel gas may happen
4
Far from high frequency devices such as high frequency welding machine
5
Far from the location where acidic solutions used frequently
6
Far from the location where special sprayer (sulfide contained) used
frequently
7
Ambient temperature 5 ° to 30 ° C, relative humidity less than 80%
(non-condensing)
As for convenience and safety of maintenance and installation, keep certain
distance between the indoor unit and obstacles.
Ensure the measurements during indoor installation:
1)The distance between the bath top and obstacles shall be at least 300mm;
2)The distance between two bath sides and obstacles shall be at least 500mm;
3)The distance between bath back and obstacles shall be at least 500mm;
4)The distance between bath front and obstacles shall be at least 1000mm;
Such as following picture:(unit: mm)

Recirculating cooler(Chiller)
7
1-2 times enlarge the distance if large size unit installed.
Unit installation:
Unpack the unit and check the packing list, if there is a shortage contact your
local distributor for solutions .
Move the unit from the packing base to spacious position and check for damages
incurred during transit. Install the unit in the proper position and ensure it is stable
and safe, then install the external circulating loop pipelines.
Please ensure correct collection of the external circulating water loop (That is,
the water outlet of the chiller should be connected to the water inlet pipe of the
instrument needs to be cooled; the return pipe of the instrument should be
connected to the water inlet ).
Please pay attention to joint leakage after injection of the bath liquid. If
leakage happens, please fix it in time.