Caution:
•Do not operate the Revive M2 system during transit.
•Do not store or transport the Revive M2 system beyond the specified
temperature, humidity, and atmospheric pressure range
(see Chapter 8 - Maintenance and Storage).
•Do not use the Revive M2 system beyond the specified temperature
range: 10 to 30°C (50 to 86°F).
•Do not use the Revive M2 system beyond the specified humidity range:
30%rH–75%rH, non-condensing.
•Use the Revive M2 system up to 3000m above mean sea level.
•To prevent any damage to the Revive M2 system, keep it away from
dust, lint, and dirt. Keep away from sources of heat or moisture.
•Do not inflate the garment without wearing it over the intended body
area. Doing so can damage the garment.
•To prevent any damage to equipment, use only accessories,
detachable parts, and materials described in this User Manual.
•While in use, place the console on horizontal firm surface only.
Do not place the console on bed, blanket, mattress, pillow, or soft
furniture. Do not cover the console.