Nuvo V1360A Manuale utente

Nuvo Video Display Systems – User Manual
Nuvo Surgical
Video Display System
Model V1360A
USER MANUAL
For Parts or Technical Assistance
USA and CANADA (800) 663-1152
INTERNATIONAL (814) 899-422
M2Z00167 Page 1

Nuvo Video Display Systems – User Manual
Table of Contents
Intended Use 3
Product Overview 4
Surgical Monitor Positioning 5
Instructions for Use 6
To Turn the Monitor On/Off 6
To Connect a Video Source to the Video Display System 7
To Connect the Video Display System to a Display or Recording Device 8
To Select a Video Signal for Display at a Monitor Using the Wall Control 9
To Select a Video Signal for Display at a Monitor Using the Monitor Handle 10
To Remove the Sterile Handle and Reinstall 11
To Adjust the Monitor Image 11
Maintenance 12
Cleaning 12
Sterilization 12
Sterilizable Handle 13
Using Optional Sterile Handle Covers 13
Preventive Maintenance 14
Consumable Parts 14
Troubleshooting 15
Nuvo Video Display System Specifications 16
Dimensions 16
Monitor Specifications 16
Environmental Conditions for Transport and Storage 17
Environmental Conditions for Use 17
Mains Power Requirements 17
Vertical Load Characteristics 17
Moment Load Characteristics 17
M2Z00167 Page 2

Nuvo Video Display Systems – User Manual
Intended Use
The Nuvo Video Display System
The Nuvo Surgical Monitor is intended to be used primarily by a surgeon to view images
generated by an endoscopic camera.
Introduction
This manual provides the information required for normal operation of the Nuvo Video
Display System. Before operating the Nuvo Surgical Monitor, be sure that you have read
and understand the contents of this manual. It is important that you read and strictly
adhere to the aspects of safety contained in these manuals.
Where applicable, this manual is intended to supplement the Nuvo Surgical Light User
Manual S2Z00125.
WARNING:
It is recommended that a complete back-up display system be available at all times
in case of a total system failure.
CAUTION:
It is strongly recommended the user verify the video system operation before
beginning any procedure.
WARNING:
Do not use the Nuvo Video Display System for radiology diagnosis. Radiographic
images are displayed for reference only. Possible misdiagnosis of the patient could
occur.
WARNING:
Equipment which is not medical grade, but connected to the system, should be kept
out of the patient vicinity. Failure to do so could result in personal injury or
equipment damage.
M2Z00167 Page 3

Nuvo Video Display Systems – User Manual
Nuvo Surgical Monitor Product Overview
Positioning
•Vertical axis rotates 300°
•Monitor rotates 300°
•Drift-resistant positioning
•Minimal airflow disturbance
•Stows out of the way
•Single or dual monitor mounts
Monitor Features
•19" flat-screen medical grade Liquid Crystal Display
•Up to four input channels (with optional switching
system)
•Wall-mounted input/output selector (with optional
switching system)
Maintenance Features
•Easily sterilized positioning handle
•Wires and connectors concealed for cleaning ease
Monitor
Assembly
Sterilizable
Handle
Counterbalance
Arm
Suspension
Arm
M2Z00167 Page 4

Nuvo Video Display Systems – User Manual
Surgical Monitor Positioning
The Nuvo Surgical Monitor system provides the surgeon with a suspension arrangement that is
well balanced and can be guided easily into position.
The suspension arms rotate 300° around the ceiling mount, and are positioned 180°apart to avoid
interfering with each other. The counterbalance arm rotates 300° around its vertical extension.
These extension pieces are custom-sized according to the ceiling height, assuring adequate
clearance.
The monitor arm turns 300° around the end of the counterbalance arm. The monitor pivots 300°
inside the arm. It can be moved completely out of the way when not in use, without affecting the
light.
The combination of these individual elements allows medical personnel to freely move the
monitor into a variety of positions using only slight pressure on the sterilizable positioning handle.
Once positioned the system remains stable, for optimal viewing of almost any procedure.
If performed by a sterile staff member, grasp the sterilized handle, and position the monitor
assembly in the desired position.
If performed by a non-sterile staff member, grasp the lower edge or side of the monitor (avoid
touching the monitor screen) and position the monitor assembly in the desired position.
190°
M2Z00167 Page 5

Nuvo Video Display Systems – User Manual
Instructions for Use of the Nuvo Video Display System
The Nuvo Video Display System is designed to view images from multiple sources, such as
endoscopic cameras, remote video cameras, radiology, pathology, PACS, etc.
Systems may handle up to four separate input signals via an optional video switching board,
although basic systems only handle one input.
Each source can be easily selected and viewed by pressing a button on the wall controls or on
the sterilizable handle at the monitor.
To Turn the Monitor On/Off:
Turn the monitor system on by using the on/off switch located on the back of the
monitor.
CAUTION:
Ensure the monitor is turned off when not in use. Failure to do so will decrease the life of
the monitor.
CAUTION
Some video sources are unstable during their power on sequence – it may be necessary to
turn the monitor power off, then back on for the image to be displayed properly.
CAUTION
The optional video display system requires the monitors are powered on before input
selections are made.
Monitor
On/Off
Switch
M2Z00167 Page 6

Nuvo Video Display Systems – User Manual
To Connect a Video Source to the Video Display System
CAUTION:
Take care to align the pins on the connector with the holes in the plug on the
input panel. Do not force the connector into the plug as this may damage the
connector.
Plug the video cable (supplied by others) into a Video Input Panel having the same video format.
Composite Video S-Video RGB Video
NOTE:
The composite video cable must be twisted clockwise to ensure that the cable is securely locked.
The s-video connector can only be inserted in one orientation and does not have any mechanical
retention, other than the friction fit of the pins.
The RGB video cable must be secured with the two small jack screws on either side of the
connector.
M2Z00167 Page 7

Nuvo Video Display Systems – User Manual
To Connect the Video Display System to a Display
CAUTION:
Take care to align the pins on the connector with the holes in the plug on the
output panel. Do not force the connector into the plug as this may damage the
connector.
Plug the video cable (supplied by others) into the nearest appropriate Video Output Panel.
NOTE:
Output panels are provided with sockets for each video format available for input signals, with the
exception that Composite video signals are converted to S-Video signals for display and output.
Always ensure that the output signal is compatible with
(a) the signal to be recorded or displayed and
(b) the device to which it will be connected.
NOTE:
Output Panels are labeled Output A and Output B. The signal delivered to Output A duplicates
the signal displayed on Monitor A and, if a second monitor (Monitor B) is provided, then its
display image is duplicated at the Output B panel.
NOTE:
The s-video connector can only be inserted in one orientation and does not have any mechanical
retention, other than the friction fit of the pins.
The RGB video cable must be secured with the two small jack screws on either side of the
connector.
The RS-232 cable must be secured with the two small jack screws on either side of the connector.
This RS-232 port is only configured for use with National Display Systems monitors.
M2Z00167 Page 8

Nuvo Video Display Systems – User Manual
To Select a Video Signal for Display at a Monitor Using the Optional Wall
Control
NOTE:
This section applies only to systems with the optional video switching.
NOTE:
The Wall Control Panel is an optional feature and may not be included in the system. Identical
functionality is available from the button, if provided, on the monitor’s sterilizable handle.
1. Determine whether Monitor A or Monitor B is to be switched.
2. Go to the Wall Control, where the bottom panel is dedicated to video switching. Press the
associated button (to the left of the panel for Monitor A, to the right for Monitor B)
repeatedly until the desired image appears on the monitor.
NOTE:
For systems which have a single video source, there is no need for switching and the wall control
panel is not provided.
NOTE:
The video channel selected for display is indicated by one of the four blue LED indicators
associated with that monitor on the panel. If fewer than 4 video inputs are provided, or if
any monitor does not have a video source connected to it, the monitor will show a blank
image when the associated channel is selected.
M2Z00167 Page 9

Nuvo Video Display Systems – User Manual
To Select a Video Signal for Display at a Monitor Using the Monitor Handle
NOTE:
This section applies only to systems with the optional video switching.
Video Display Systems with switching capability may be provided with a button at the bottom of
the sterilizable handle adjacent to each monitor. Channel selection is achieved by repeatedly
pressing this button until the desired image is displayed.
NOTE:
The selection button on the handle performs the same function as the selection button at the
optional wall control panel.
NOTE:
This operation should only be performed by a surgeon or other sterile staff member. Operation by
a non-sterile member of staff will contaminate the handle, which should be immediately replaced
with a sterile handle or covered by a sterile disposable cover.
NOTE:
The video channel selected for display is indicated at the wall control panel by one of the four blue
LED indicators associated with that monitor. If fewer than 4 video inputs are provided, or if any
monitor does not have a video source connected to it, the monitor will show a blank image when
the associated channel is selected.
M2Z00167 Page 10
Indice