Olympus OmniScan MX2 Manuale utente

OmniScan MX2 — Getting Started Guide 1
Intended Use
The OmniScan MX2 is designed to perform nondestructive inspections of industrial and commercial
materials. Do not use the OmniScan MX2 for any purpose other than its intended use.
Instruction Manual
Before use, thoroughly review the OmniScan MX and MX2 User’s Manual and the OmniScan MXU
Software User’s Manual, and use the product as instructed. The user’s manuals contain essential
information on how to use this Olympus product safely and effectively. Keep this guide and the user’s
manuals in a safe and accessible location. The PDF versions of the OmniScan MX2 User’s Manual and
OmniScan MXU Software User’s Manual are also available on the documentation USB drive that is
shipped with the OmniScan MX2, or can be downloaded at www.olympus-ims.com.
Safety Signal Words
Indicates an imminently hazardous situation calling attention to a procedure, practice, or the like,
which, if not correctly performed or adhered to, could result in death or serious personal injury
Indicates a potentially hazardous situation calling attention to a procedure, practice, or the like, which,
if not correctly performed or adhered to, could result in death or serious personal injury.
Indicates a potentially hazardous situation calling attention to a procedure, practice, or the like, which,
if not correctly performed or adhered to, could result in minor or moderate personal injury, material
damage, particularly to the product, destruction of part or all of the product, or loss of data.
OmniScan MX2
Getting Started Guide

2 OmniScan MX2 — Getting Started Guide
Connections
To reduce the risk of electric shock, avoid touching the inner conductor of the probe connectors. Up to
340 V can be present on the inner conductor of the UT connectors, and up to 115 V can be present on
the PA connector. The warning symbol near the PA and UT connectors signals this electric shock risk.
Alarm and I/O
connector
Scanner interface
SVGA video
output
Ethernet (RJ-45) port
Phased array probe
connector
DC power adaptor plug
OTG (On-the-Go USB port compatible with USB 2.0)
Serial port
USB high-speed (HS) port (compatible with USB 2.0)
USB full-speed (FS) port (compatible with USB 1.1)
SDHC memory card slot

OmniScan MX2 — Getting Started Guide 3
Warnings on Battery Use
Carefully read and observe the following warnings on battery use.
• Do not open, crush, or perforate batteries; doing so could cause injury.
• Do not incinerate batteries. Keep batteries away from fire and other sources of extreme heat.
Exposing batteries to extreme heat (over 80 °C) could result in an explosion or personal
injury.
• Do not drop, hit, or otherwise abuse a battery, as doing so could expose the cell contents,
which are corrosive and explosive.
• Do not short-circuit the battery terminals. A short circuit could cause injury and severe
damage to a battery, making it unusable.
• Do not expose a battery to moisture or rain; doing so could cause an electric shock.
• Only use the OmniScan MX2 or an external charger approved by Olympus to charge the
battery.
• Do not recharge a battery unless the indicators light up when the capacity check button on
the battery is pressed. Doing so could be dangerous.
• Do not store batteries that have less than 40 % remaining charge. Recharge batteries to
between 40 % and 80 % capacity before storing them.
• During storage, keep the battery charged to between 40 % and 80 %.
• Do not leave the battery in the OmniScan MX2 during instrument storage.
Replacing a Lithium-Ion Battery
To replace a lithium-ion battery
1. On the right side panel, turn the two quarter-turn latches counterclockwise to open the battery
compartment door.
2. Remove the battery by pulling on the strap.
3. Insert a new battery. Make sure that the groove on the battery is aligned with the small edge inside
the battery compartment.
4. Close the battery compartment door.
Battery compartment door
Lithium-ion battery (P/N:
OMNI-A-BATT
[U8760010])

4 OmniScan MX2 — Getting Started Guide
Figure 1 Battery replacement
The OmniScan MX2 can operate using only one battery, making it possible to remove the other battery
while the unit is in operation (hot swap). When two batteries are used, the OmniScan MX2
automatically discharges both batteries at the same time, increasing the remaining operating time by
about 10 %. To discharge one battery at a time, simply remove the other battery.
Charging the Batteries
To charge the batteries
Connect the OmniScan MX2 to the supplied DC power adaptor.
• When the OmniScan MX2 is connected to a proper DC power source and is turned off, it
will automatically recharge any battery requiring charging, starting with the weakest
battery (if the recharge conditions are met).
Once the weakest battery has reached the level of the other battery, both batteries are
charged simultaneously.
The power indicator light indicates that batteries are currently being recharged when it
flashes orange. When the charge is complete, the power indicator light emits a steady
orange light. Each battery can take up to 3.5 hours to fully charge from a completely
discharged state (less than 5 % remaining charge).
• When the OmniScan MX2 is running and connected to a proper DC power source, it will
automatically recharge any battery requiring charging, starting with the weakest battery (if
the recharge conditions are met). The indicator for the battery being recharged flashes
yellow.
Once the weakest battery has reached the same level as the other battery, both batteries are
charged simultaneously.
Because the OmniScan MX2 is running, less power is available from the DC source to
recharge the battery or batteries. As such, it could take up to 8 hours to charge each
completely discharged battery. For more information about battery charge status, refer to
the OmniScan MX and MX2 User’s Manual.
The OmniScan MX2 DC power adaptor (P/N: OMNI-A-AC [U8767093]) is designed to charge
OmniScan MX2 batteries only (P/N: OMNI-A-BATT [U8760010]). Do not attempt to charge any other
type of battery, or use any other chargers/adaptors. Doing so could cause an explosion or injury. Do
not attempt to power or charge other electronic equipment with the OMNI-A-AC DC power adaptor,
as doing so could cause an explosion resulting in death or serious personal injury.
Battery Status Indicators
When the OmniScan MX2 is running, the battery status is shown by indicators located on the upper-
left corner of the touch-screen display. These indicators display the amount of power remaining in
each battery using one of the two following methods:
• The remaining operational time is displayed inside the battery status indicator. This
information can only be displayed accurately after the OmniScan MX2 has been operating
for approximately 15 minutes.

OmniScan MX2 — Getting Started Guide 5
• The charge indicator (bar) inside the battery status indicator provides an approximation of
the amount of power remaining for that battery.
If you attempt to start up the OmniScan MX2 using batteries with a charge too low for operation, the
power indicator light rapidly flashes red for approximately three seconds. To operate the
OmniScan MX2, replace the batteries or plug in the DC power adaptor.
The battery currently in use is highlighted. When both batteries are used simultaneously, they will
both appear highlighted. The OmniScan MX2 software keeps the user informed of the battery status.
The table below describes some of the battery status indicators available:
For other status indications, refer to the OmniScan MX and MX2 User’s Manual.
Indicator Outline Fill Meaning
Dotted N/A There is no battery in that compartment.
Blue Blue The battery is functioning properly.
Yellow
(blinking)
Blue The battery is charging.
Red
(blinking)
Blue The battery charge is critically low (less
than 10 %). A beeping sound is emitted if
no other battery is available.
N/A Yellow The OmniScan MX2 is being powered
externally through the DC adaptor.

6 OmniScan MX2 — Getting Started Guide
Front Panel
The front panel of the OmniScan MX2 contains all the main controls and indicators. This panel is
divided into seven areas: the touch-screen display, the main control area, the function keys, the power
button, the increment keys, the Help key, and the indicator lights.
Main Control Area
You can control the OmniScan MX2 using the main control area alone.
Element Name Description
Scroll knob Used to navigate through selections without the
need for a keyboard, mouse, or touch screen.
Turning the scroll knob in a clockwise direction
shifts a selection to the left (horizontal list) or
upward (vertical list).
Accept key Used to confirm a selection.
Cancel key Used to cancel the current selection, or to go back
one level in the menu hierarchy.
Display touch
screen
Increment
keys
Power
indicator
light
Acquisition
indicator light
Help key
Alarm indicator
lights
Function keys
Power button
Function
keys
Main
control
area

OmniScan MX2 — Getting Started Guide 7
Turning On and Off the OmniScan MX2
To turn on the OmniScan MX2
Press the Power key.
The system starts up and performs a memory check.
If the system encounters a problem during its start-up phase, the power indicator light uses a color
code to identify the nature of the problem. Refer to the OmniScan MX2 User’s Manual for more
details.
To turn off the OmniScan MX2
1. Quickly press the Power key.
The “Select a command” message appears.
2. Select Shut Down.
A message asking you if you want to save your setup appears.
3. To save your setup, select Yes.
You can also turn off the OmniScan MX2 by pressing and holding the Power key for ten seconds.
However, your setup will NOT be saved.
Sleep Mode
The OmniScan MX2 can be put into sleep mode when not in use to conserve power.
To use the sleep mode
1. With the OmniScan MX2 turned on, quickly press the Power key.
The “Select a command” message appears.
2. Select Sleep.
3. To disable the sleep mode, quickly press the Power key.
The OmniScan MX2 returns to its previous state (inspection or analysis mode).
The Shut Down
button
The Sleep button

Olympus NDT, 48 Woerd Avenue, Waltham, MA 02453, USA
www.olympus-ims.com
Printed in Canada • © 2011, 2013 by Olympus. All rights reserved.
All brands are trademarks or registered trademarks of their respective owners and third party entities.
Part ID: U8778403 *U8778403* Printed on Rolland Hitech50,
which contains 50 % post-
consumer fiber.
DMTA-20019-01EN
Rev. B, February 2013 *DMTA-20019-01EN*
50%
OmniScan MX2 — Getting Started Guide
Adaptors
The figure below shows how the OmniScan MX2 can be integrated with your existing equipment.
Adaptors Phased array extension
cable (U8800635 or
similar)
DE15 to LEMO encoder
cable adaptor
(U8775201)
Scanner with DE15 encoder
cable
To phased
array probe
connector
To scanner interface
OR
Scanner with LEMO encoder
cable
OR
Altri manuali per OmniScan MX2
1
Indice
Altri manuali Olympus Sensore di sicurezza

Olympus
Olympus EPOCH 6LT Manuale utente

Olympus
Olympus Sonic 1200S/HR Manuale utente

Olympus
Olympus nortec 600 Manuale utente

Olympus
Olympus Panametrics Epoch 4 Plus Manuale utente

Olympus
Olympus EPOCH 6LT Manuale utente

Olympus
Olympus nortec 600 Manuale utente

Olympus
Olympus EPOCH 600 Istruzioni per il montaggio

Olympus
Olympus OmniScan X3 Manuale utente

Olympus
Olympus EPOCH 1000 Series Manuale utente

Olympus
Olympus EPOCH 6LT Manuale utente