13 (Continued on page 14)
MENU continued from page 12
option. The SAVE/ZERO button (C) advances to the next item
are based on US time and date settings using a 24 hour clock.
After adjusting all numbers press and release the POWER/
MUTEbutton (A)to savethe adjustment.Use theSAVE/ZERO
button (C) at this time to scroll to another menu function in the
USER MENU. Pressing the POWER/MUTE button (A) will re-
turn the instrument to the working display.
SHOW CAL LOG
From the working display access the user menu by pressing
andholding theMENU button(B)until thetop lineof thedisplay
readsUSER MENU.Press andreleasethe SAVE/ZERObutton
(C) until the bottom line displays SHOW CAL LOG. Press the
MENU button (B) and the display will show the last gas cali-
bratedand calibration date.Pressing any buttonwill return the
display to the user menu. Pressing the POWER/MUTE button
(A) will return the instrument to the working display.
SHOW SES LOG
From the working display access the user menu by pressing
andholding theMENU button(B)until thetop lineof thedisplay
readsUSER MENU.Press andreleasethe SAVE/ZERObutton
(C) until the bottom line displays SHOW SES LOG. Press the
MENU button (B). Use the SAVE/ZERO button (C) to scroll to
the saved session you wish to review. SESSION 1 is the most
recent data saved. Pressing the MENU button (B) will display
the date and time of that session. Pressing the MENU button
(B)againwill displaythegas reading.PressthePOWER/MUTE
14
(Continued on page 15)
MENU continued from page 13
(B)againwill displaythe gasreading.Press thePOWER/MUTE
button (A) to return to SESSION(#) and pressing the SAVE/
ZERO button (C) will allow you to scroll all previously saved
SESSIONS. Pressing the POWER/MUTE button (A) 2 times
willreturnyouto theUSERMENU.PressingthePOWER/MUTE
button (A) once more returns you to the working display. The
number of stored session log saves is factory set at 6. It can
store up to 100 by changing a factory setting (contact SENSIT
TECHNOLOGIESfor instructions).
BUMPTEST
Fromtheworkingdisplayaccessthe menubypressingandhold-
ing the MENU button (B) until the top line of the display reads
USERMENU.PressandreleasetheSAVE/ZERObutton(C)until
thebottomlinedisplaysBUMPTEST.Prepare50%LELmethane
(orpentane)calibrationgasforapplication totheinstrument.
Applycalibration gastothe instrumentsensor. PresstheMENU
button(B).The readingmustread80% ofcalibratedvaluewithin
30 seconds. The reading is on the left side and the timer is on
the right side of the display.
If the instrument passes, the display will read BUMP TEST
PASSES, a beep will sound and the unit display will automati-
cally return to the user menu. If the instrument fails, the display
willreadBUMP TEST FAILED and abeepwillsound. Repeated
bump test failure indicates possible need for instrument repair.
Contact JAnd N for instructions.
At the end of any bump test press the POWER/MUTE button
(A) to return to the working display.