www.Siargo.com
FS4100UserManual
9
|
Page
4. Installation
Donotopenoralteranypartoftheproductwhichwouldleadtomalfunctionandirrecoverable
damage.Itwillalsoforfeitthetermsofthewarrantyandcauseliability.
Theproductatthetimeofshipmentisfullyinspectedforitsqualityandmeetsallsafetyrequirements.
Additionalsafetymeasuresduringtheinstallationshouldbeapplied.Thisincludes,butisnotlimited
totheleakageverificationprocedures,standardEDS(electrostaticdischarge)precautions,DC
voltageprecautions.Othertaskssuchascalibration,partreplacement,repair,andmaintenance
mustonlybeperformedbytrainedpersonnel.Uponrequest,themanufacturerwillprovide
necessarytechnicalsupportand/ortrainingofthepersonnel.
Therearenopreferredspacedirectionsfortheinstallation.Flowdirectionshouldbealignedwiththe
arrowmarkonthemeterbody.Iftheflowingfluidmayhaveparticlesordebris,afilterisstrongly
recommendedtobeinstalledupstreamofthemeter.
Pleasefollowthefollowingstepstocompletetheinstallation:
a)Uponopeningthepackage,theproduct'sphysicalintegrityshouldbeinspectedtoensureno
visualdamage.
b)Beforeinstallationoftheproduct,pleaseensurethatthepipedebrisorparticlesoranyother
foreignmaterialsarecompletedremoved.
c)Closetheupstreamvalve,ifany,completely.
d)Duringinstallation,pleasemakesurenoforeignmaterials(suchaswater,oil,dirt,particles,etc.)
enteringintotheinstallationpipeline.
e)ConnectelectricalwiresperthewiredefinitioninTable3.1.Pleasebesureofthepowersupply
range(i.e.,8~24VDC)andpowersupplypolarization.Ifanadapterisused,makesurethe
adaptermeetsindustrialstandardsandhasallsafetycertifications.Alternatively,thisproduct
canalsobepoweredbya9Vdcbattery.
f)Forthedatacommunicationwireconnection,pleasefollowthedescriptioninTable3.1andmake
surethatthewiresarecorrectlyconnectedtotheproperportsonyourdatadevice/equipment.
Pleasemakesurethedatacablemeetsindustrialstandardswithpropershielding.
g)Slowlyopenthevalve(s)ofthegassupplyifany,upstreamordownstream,orbothofthe
pipeline.theproductshouldthenstarttomeasuretheflowinthepipeline.
Note:becausethemeterhasalargedynamicalmeasurementrange,itcouldbenormalif
youseethesmallinstantflowrateevenifthereis“no‐flow”inthepipeline.Ifthe
valueconsistentlypresent,double‐checkthepipeleakageandthenresettheoffsetif
youaresurethereisnoleakageorflow.
h)Thiswillconcludetheinstallation.