Operating Instructions inEos Blue
Table of contents
1Dear Customer, ........................................................................................................ 5
2General information.................................................................................................. 6
2.1 Structure of the document............................................................................. 6
2.1.1 Identification of the danger levels....................................................... 6
2.1.2 Formats and symbols used................................................................ 7
2.2 Legend .......................................................................................................... 7
2.3 Warranty........................................................................................................ 7
3Safety ....................................................................................................................... 8
3.1 Basic safety information ................................................................................ 8
3.1.1 Prerequisites ...................................................................................... 8
3.1.2 Connecting the unit ............................................................................ 8
3.1.2.1 Connection to the CEREC 3 acquisition unit.................................... 8
3.1.2.2 Connection to the CEREC AC acquisition unit................................. 9
3.1.3 Connection of external equipment ..................................................... 10
3.1.4 Maintenance and repair ..................................................................... 10
3.1.5 Changes to the product...................................................................... 10
3.1.6 Accessories........................................................................................ 10
3.1.7 In case of damage.............................................................................. 10
3.2 Intended use ................................................................................................. 11
3.3 Certification ................................................................................................... 12
4Technical description................................................................................................ 13
4.1 Design and function ...................................................................................... 13
4.2 System components...................................................................................... 13
4.2.1 Components of the inEos Blue........................................................... 14
4.2.2 Accessories (included in the scope of supply) ................................... 15
4.2.3 Calibration parts (included in scope of supply) .................................. 16
4.2.3.1 3D acquisition system ...................................................................... 16
4.2.3.2 XYZ axis alignment .......................................................................... 16
4.3 Technical data............................................................................................... 17
4.4 System requirements .................................................................................... 17
5Transport to the installation site ............................................................................... 18
5.1 Transport and unpacking .............................................................................. 18

62 99 205 D3446
D3446.201.02.04.02 03.2012 3
Sirona Dental Systems GmbH
Operating Instructions inEos Blue
båÖäáëÜ
5.2 Installation site.............................................................................................. 18
6Initial installation of the software.............................................................................. 19
6.1 Installing the software................................................................................... 19
6.2 Connecting the unit to the PC ...................................................................... 19
6.3 Switching the unit on and starting the software............................................ 20
6.4 Adding inEos Blue to the device configuration ............................................. 20
7Connecting the unit.................................................................................................. 21
7.1 Connecting the inEos Blue to the PC........................................................... 21
7.2 Connecting the inEos Blue to the power supply........................................... 22
7.3 Connecting the rotation mouse .................................................................... 23
7.4 Connecting the foot switch ........................................................................... 23
8Controls and displays .............................................................................................. 24
8.1 Controls on the inEos Blue........................................................................... 24
8.2 Ready LED ................................................................................................... 25
9Switching the system on.......................................................................................... 26
10 Calibration ............................................................................................................... 27
10.1 General information on calibration ............................................................... 27
10.2 Calibrating the 3D camera system ............................................................... 27
10.3 Calibrating the XYZ axis alignment .............................................................. 28
11 Performing 3D acquisition ....................................................................................... 32
11.1 General......................................................................................................... 32
11.2 Switching the inEos Blue on/off.................................................................... 34
11.3 Grid scans .................................................................................................... 34
11.4 Free scans.................................................................................................... 36
11.5 Rotational scans........................................................................................... 38
11.6 Acquiring a buccal registration ..................................................................... 39
11.7 Examples...................................................................................................... 39
11.7.1 Scan of a 3-unit bridge ...................................................................... 39
11.7.1.1Acquiring a preparation ................................................................... 39
11.7.1.2Rotational scans.............................................................................. 41
11.7.1.3Scanning the opposing jaw ............................................................. 42

62 99 205 D3446
4D3446.201.02.04.02 03.2012
Sirona Dental Systems GmbH
Operating Instructions inEos Blue
11.7.1.4Buccal registration............................................................................ 42
11.7.1.5Terminating acquisition .................................................................... 43
11.7.2 Acquisition of a crown cap ................................................................. 43
12 Cleaning and care .................................................................................................... 44
12.1 Care of the 3D camera system ..................................................................... 44
12.2 Cleaning the outer surface ............................................................................ 44
12.3 Protection against medicaments ................................................................... 44

62 99 205 D3446
D3446.201.02.04.02 03.2012 5
Sirona Dental Systems GmbH 1Dear Customer,
Operating Instructions inEos Blue
båÖäáëÜ
1Dear Customer,
Dear Customer,
Thank you for purchasing your inEos Blue® from Sirona.
Improper use and handling can create hazards and cause damage.
Please read and follow these operating instructions carefully and always
keep them within easy reach.
To prevent personal injury or material damage, it is important to observe
all safety information.
To safeguard your warranty claims, please complete the attached
Installation Report / Warranty Passport when the system is handed over
and send it to the indicated fax number.
Your
inEos Blue Team

62 99 205 D3446
6D3446.201.02.04.02 03.2012
2 General information Sirona Dental Systems GmbH
2.1 Structure of the document Operating Instructions inEos Blue
2General information
General information
Please read this document completely and follow the instructions exactly.
You should always keep it within reach.
Original language of the present document: German.
2.1 Structure of the document
Structur e of the d ocument
2.1.1 Identification of the danger levels
Identificatio n of the dan ger levels
To prevent personal injury and material damage, please observe the
warning and safety information provided in this document. Such
information is highlighted as follows:
Tip: Information on making work easier.
DANGER
An imminent danger that could result in serious bodily injury or death.
WARNING
A possibly dangerous situation that could result in serious bodily injury
or death.
CAUTION
A possibly dangerous situation that could result in slight bodily injury.
NOTICE
A possibly harmful situation which could lead to damage of the product
or an object in its environment.
IMPORTANT
Application instructions and other important information.

62 99 205 D3446
D3446.201.02.04.02 03.2012 7
Sirona Dental Systems GmbH 2General information
Operating Instructions inEos Blue 2.2 Legend
båÖäáëÜ
2.1.2 Formats and symbols used
Formats and symbols used
The formats and symbols used in this document have the following
meaning:
2.2 Legend
Legend
Year of manufacture
2.3 Warranty
Warranty
To safeguard your warranty claims, please complete the attached
Installation Report / Warranty Passport when the unit is handed over.
Then fax it to the specified fax no.
✔Prerequisite
1. First action step
2. Second action step
or
➢ Alternative action
Result
Prompts you to do something.
See "Formats and symbols
used [ → 7]“
Identifies a reference to another
text passage and specifies its page
number.
● List Designates a list.
"Command/menu item" Identifies commands, menu items
or quotations.
20XX

62 99 205 D3446
8D3446.201.02.04.02 03.2012
3 Safety Sirona Dental Systems GmbH
3.1 Basic safety information Operating Instructions inEos Blue
3Safety
Safety
3.1 Basic safety information
Basic safety in formation
3.1.1 Prerequisites
Prerequisites
3.1.2 Connecting the unit
Connecti ng the un it
Description
Perform connection by following the directions given in the present
operating instructions.
3.1.2.1 Connection to the CEREC 3 acquisition unit
Connecti on to the CEREC 3 acquisiti on unit
Warning: inEos connected to CEREC 3 acquisition unit
NOTICE
Important information on the building installation
The building installation must be performed by a qualified expert in
compliance with the national regulations. DIN VDE 0100-710 applies in
Germany.
NOTICE
Restrictions regarding installation site
The system is not intended for operation in areas subject to explosion
hazards.
NOTICE
Do not damage the unit!
The unit can be damaged if opened improperly.
It is expressly prohibited to open the unit with tools!
WARNING
No acquisitions in the patient's mouth
As long as an inEos Blue is connected to a CEREC 3 acquisition unit,
no acquisitions may be made in the patient's mouth using the connected
CEREC 3D camera.

62 99 205 D3446
D3446.201.02.04.02 03.2012 9
Sirona Dental Systems GmbH 3Safety
Operating Instructions inEos Blue 3.1 Basic safety information
båÖäáëÜ
3.1.2.2 Connection to the CEREC AC acquisition unit
Connection to the CEREC AC acquisition unit
Patient environment
Within the patient environment (A), direct contact is only permissible with
devices or system parts that are approved for use in the patient
environment (A).
This applies to all possible patient positions (B) during the examination or
treatment.
inEos outside of patient environment
WARNING
No acquisitions in the patient's mouth within the patient environment
As long as an inEos Blue is connected to a CEREC AC acquisition unit,
no acquisitions may be made in the patient's mouth using the connected
CEREC Bluecam if the inEos Blue is located outside of the patient
environment.
1500mm
59”
1500mm
59”
1500mm
59”
1500mm
59”
1500mm
59”
1500mm
59”
1500mm
59” 1500mm
59”
1500mm
59”
1500mm
59”
1500mm
59”

62 99 205 D3446
10 D3446.201.02.04.02 03.2012
3 Safety Sirona Dental Systems GmbH
3.1 Basic safety information Operating Instructions inEos Blue
3.1.3 Connection of external equipment
Connecti on of exte rnal equipment
If any equipment not approved by Sirona is connected, it must comply
with the applicable standards:
● EN 60 950 for information technology equipment, and
● EN 61 010-1 for laboratory equipment.
3.1.4 Maintenance and repair
Maintenance and repair
As manufacturers of dental instruments and laboratory equipment, we
can assume responsibility for the safety properties of the unit only if the
following points are observed:
● The maintenance and repair of this unit may be performed only by
Sirona or by agencies authorized by Sirona.
● Components which have failed and influence the safety of the unit
must be replaced with original (OEM) spare parts.
Please request a certificate whenever you have such work performed. It
should include:
● The type and scope of work.
● Any changes made in the rated parameters or working range.
● Date, name of company and signature.
3.1.5 Changes to the product
Changes to th e product
Modifications to this unit which may affect the safety of the operator,
patients or third parties are prohibited by law!
3.1.6 Accessories
Accessories
In order to ensure product safety, this product may be operated only with
original Sirona accessories or third-party accessories expressly approved
by Sirona. The user assumes the risk of using non-approved accessories.
3.1.7 In case of damage
In case of da mage
In case of noticeable malfunctions or damage, stop using the instrument
immediately and notify your dental depot or the manufacturer.