2.1 Electrical Safety and Preparation for Use
Page 2
Voltages capable of causing injury or death are present in this instrument.
Use extreme caution whenever the instrument cover is removed. This instrument
was designed for indoor use only.
Line Voltage
This instrument is designed to operate with a 100 to 240 VAC, 47 to 63 Hz power source.
DC operation with +12 to +36 VDC, + 2 Amperes is also possible.
Fuse
A 1.0 Ampere 250V slow-blow fuse is used for 100-240 VAC operation.
A 2.5 Ampere 250V slow-blow fuse is used for +12 to +36 VDC operation.
Only replace fuses with the same type and specifications.
AC Power
The instrument has a detachable three wire power cord for connection to a grounded AC
power source. The enclosure of the unit is directly connected to the outlet ground to protect
against electrical shock. Always use an outlet with a protective ground and do not disable
this safety mechanism. Detaching the AC power cord is the only option of disconnecting
the unit from the AC mains supply. Make sure you have access to the rear panel or provide
an external accessible AC disconnect means for your PD-15RMi-C.
DC Power
If the instrument was acquired with the DC power module, the DC connector on the back
panel has the following configuration:
Pin 1 NC
Pin 2 NC
Pin 3 NC
Pin 4 +12 to +36 VDC power return
Pin 5 +12 to +36 VDC power
Pin 6 Chassis GND /Earth GND
Verify that the connector from your DC power supply has the pin configuration mentioned
above. Do not apply AC voltage to the DC power connector. Failure to do so may cause
injury or death to personnel, cause damage to the instrument and void all warranties.
Please note that the power return (pin 4) is NOT connected to the instrument case ground
internally, however both ground connections pin 4) and pin 6) are available at the DC
power connector and may be connected together at this point.