SUPMEA Turbidity/TSS/MLSS Manuale utente

User's Manual
U-DO2.1-MYEN1
Turbidity/TSS/MLSS Controller
Supmea Automation Co.,Ltd.
www.supmea.com
Headquarters
5th floor,Building 4,Singapore Hangzhou Science Technology Park,No. 6 street,
Hangzhou Economic Development Area,Hangzhou 310018,China
Singapore
2 Venture Drive #11-30 Vision Exchange Singapore
Philippines
Majestic Subdivision, Lot 1, 1800 Rainbow St, Marikina, 1811 Metro Manila, Philippines

I
Preface
Thank you for purchasing Turbidity/TSS/MLSS controller. Please
read this manual carefully before operating and using it correctly to
avoid unnecessary losses caused by false operation.
Note
Modification of this manual’s contents will not be notified as
a result of some factors, such as function upgrading.
We try our best to guarantee that the manual content is
accurate, if you find something wrong or incorrect, please contact
us.
Version
U-DO2.1-MYEN1

II
Safety Precautions
In order to use this product safely, be sure to follow the safety
precautions described here.
About this manual
Please submit this manual to the operator for reading.
Please read the operation manual carefully before applying
the instrument. On the precondition of full understanding.
This manual only describes the functions of the product. The
company does not guarantee that the product will be suitable for a
particular use by the user.
Precautions for protection, safety and modification of
this product
To ensure safe use of this product and the systems it controls,
please read carefully the operation manual and understand the
correct application methods before putting into operation, to avoid
unnecessary losses due to operation mistakes. If the instrument is
operated in other ways not described in the manual, the protections
that the instrument give may be destroyed, and the failures and
accidents incurred due to violation of precautions shall not be borne
by our company.
When installing lightning protection devices for this product
and its control system, or designing and installing separate safety
protection circuits for this product and its control system, it needs to
be implemented by other devices.

III
If you need to replace parts of the product, please use the
model specifications specified by the company.
This product is not intended for use in systems that are directly
related to personal safety. Such as nuclear power equipment,
equipment using radioactivity, railway systems, aviation equipment,
marine equipment, aviation equipment and medical equipment. If
applied, it is the responsibility of the user to use additional
equipment or systems to ensure personal safety.
Do not modify this product.
The following safety signs are used in this manual:
Hazard, if not taken with appropriate precautions, will result in
serious personal injury, product damage or major property damage.
Warning: Pay special attention to the important information
linked to product or particular part in the operation manual.
Confirm if the supply voltage is in consistent with the rated
voltage before operation.
Do not use the instrument in a flammable and combustible
or steam area.
To prevent from electric shock, operation mistake, a good
grounding protection must be made.

IV
Thunder prevention engineering facilities must be well
managed: the shared grounding network shall be grounded at
is-electric level, shielded, wires shall be located rationally,
SPD surge protector shall be applied properly.
Some inner parts may carry high voltage. Do not open the
square panel in the front except our company personnel or
maintenance personnel acknowledged by our company, to
avoid electric shock.
Cut off electric powers before making any checks, to avoid
electric shock.
Check the condition of the terminal screws regularly. If it is
loose, please tighten it before use.
It is not allowed to disassemble, process, modify or repair
the product without authorization, otherwise it may cause
abnormal operation, electric shock or fire accident.
Wipe the product with a dry cotton cloth. Do not use
alcohol, benzine or other organic solvents. Prevent all kinds
of liquid from splashing on the product. If the product falls
into the water, please cut off the power immediately,
otherwise there will be leakage, electric shock or even a fire
accident.
Please check the grounding protection status regularly. Do
not operate if you think that the protection measures such as
grounding protection and fuses are not perfect.
Ventilation holes on the product housing must be kept clear
to avoid malfunctions due to high temperatures, abnormal
operation, shortened life and fire.

V
Disclaimer
The company does not make any guarantees for the terms
outside the scope of this product warranty.
Please strictly follow the instructions in this manual,
otherwise the product's protective device may be damaged.
Do not use the instrument if it is found damaged or
deformed at opening of package.
Prevent dust, wire end, iron fines or other objects from
entering the instrument during installation, otherwise, it will
cause abnormal movement or failure.
During operation, to modify configuration, signal output,
startup, stop, operation safety shall be fully considered.
Operation mistakes may lead to failure and even destruction
of the instrument and controlled equipment.
Each part of the instrument has a certain lifetime, which
must be maintained and repaired on a regular basis for
long-time use.
The product shall be scrapped as industrial wastes, to
prevent
environment
pollution.
When not using this product, be sure to turn off the power
switch.
If you find smoke from the product, smell odor, abnormal
noise, etc., please turn off the power switch immediately and
contact the company in time.

VI
The company is not responsible for damage to the instrument
or loss of parts or unpredictable damage caused directly or
indirectly by improper operation of the user.
Package contents
After opening the box, please confirm the package contents
before starting the operation. If you find that the model and quantity
are incorrect or there is physical damage in appearance, please
contact us.
Chart 1 Product contents
No. Item Quantity Remarks
1 Turbidity/TSS/MLSS sensor 1
3 Turbidity/TSS/MLSS controller 1
4 Manual 1
5 Certificate 1

VII
Contents
Chapter I Introduction................................................................................. 1
1.1.Turbidity/TSS/MLSS sensor................................................................1
1.2.Turbidity/TSS/MLSS controller.......................................................... 2
Chapter II Installation..................................................................................4
2.1.Instrument installation......................................................................... 4
2.2.Sensor installation................................................................................6
2.3.Instrument wiring.................................................................................7
Chapter III Key Operation...........................................................................9
Chapter IV System menu...........................................................................11
Chapter V Communication........................................................................17
Chapter VI Calibration of MLSS/TSS...................................................... 20
Chapter VII Maintenance.......................................................................... 22
Chapter VIII Troubleshooting................................................................... 23

Chapter I Introduction
1
Chapter I Introduction
1.1.Turbidity/TSS/MLSS sensor
The sensor is based on the combined infrared absorption and scattering
light method, and the ISO7027 method can be used to accurately determine
the turbidity, TSS and MLSS. The main parameters of the sensor showed as
table 2:
Table 2 the main parameters of the Turbidity/TSS/MLSS sensor
Dimension 60mm*256mm (D*L)
Weight 1.65KG
Material
Body: SUS316L + PVC, Ti alloy (For seawater)
O-ring: Fluorine rubber
Cable: PVC
Ingress protection IP68/NEMA6P
Measure range
Turbidity: 0.00 ~ 4000NTU
MLSS/TSS : 0.1 ~ 20000 mg/L, 0.1 ~ 45000 mg/L,
0.1 ~ 120000 mg/L
Accuracy
Turbidity: Less than ±2% of the measured value or
±0.1NTU, whichever is larger
MLSS/TSS: less than ±5% of the measured value
(depending on the homogeneity of the MLSS/TSS)
Flow rate ≤2.5m/s, 8.2ft/s
Pressure range ≤0.4Mpa
Operating
temperature
0 ~ 40℃

Chapter I Introduction
2
Storage
conditions
Temperature(-15 ~ 65℃)
Humidity (5 ~ 95)%RH(no condensation)
Altitude: <2000m
Calibration Buffer solution calibration, Slope calibration
Cable length 10m standard cable, can be extended up to 100m
Sensor dimension
1.2.Turbidity/TSS/MLSS controller
This Turbidity/TSS/MLSS controller is a smart online chemical analyzer,
widely used in Thermal power, chemical fertilizer, metallurgy, environmental
protection, pharmaceutical, biochemical, food, tap water and other industries.
The main parameters of the controller showed as table 3:
Table 3 main parameters of the Turbidity/TSS/MLSS controller
Display 2.8-inch
Dimension Overall dimension:100mm * 100mm * 150mm
Cutout dimension:92.5mm*92.5mm
Weight 0.65Kg
Ingress
protection IP54
Indice