SUTO S201 Manuale

English
Instruction and operation manual
S201
Dew point sensor

Dear Customer,
Thank you for choosing our product.
The operating instructions must be read in full and carefully observed
before you start up the device. The manufacturer cannot be held liable
for any damage that occurs as a result of non-observance or non-
compliance with this manual.
Should the device be tampered with in any manner other than a
procedure that is described and specified in the manual, the warranty is
cancelled and the manufacturer is exempt from liability.
The device is destined exclusively for the described application.
S TO offers no guarantee for the suitability for any other purpose.
S TO is also not liable for consequential damage resulting from the
delivery, capability or use of this device.
2 S201

Safety instructions
Table of contents
1 Safety instructions......................................................................4
2 Registered trademarks.................................................................6
3 Application.................................................................................6
4 Features....................................................................................7
5 Technical data.............................................................................8
5.1 General................................................................................8
5.2 Electrical data.......................................................................8
5.3 Output-signals......................................................................8
6 Dimensional drawing..................................................................10
7 Determination of the installation point.........................................11
8 Installation ..............................................................................11
8.1 Installation requirements......................................................11
8.2 Installation procedure .........................................................12
8.3 Electrical connection ...........................................................13
9 Signal outputs..........................................................................15
9.1 Analog output ....................................................................15
9.2 Relay output ......................................................................15
10 Sensor display ........................................................................15
10.1 Starting process................................................................16
10.2 Error messages.................................................................16
10.3 Configuration using the display............................................17
11 Optional extra accessories........................................................18
11.1 Measuring chambers .........................................................18
11.2 Service kit........................................................................18
12 Calibration..............................................................................19
13 Maintenance...........................................................................19
14 Disposal or waste....................................................................19
15 Warranty................................................................................20
S201 3

Safety instructions
1 Safety instructions
Please check if this instruction manual matches with the
product type.
Please observe all notes and instructions indicated in this
manual. It contains essential information which must be
observed before and during installation, operation and
maintenance. Therefore this instruction manual must be read carefully
by the technician as well as by the responsible user / qualified
personnel.
This instruction manual must be available at the operation site of the
dew point sensor at any time. In case of any obscurities or questions,
regarding this manual or the product, please contact the manufacturer.
W RNING!
Compressed air!
ny contact with quickly escaping air or bursting
parts of the compressed air system can lead to
serious injuries or even death!
•Do not exceed the maximum permitted pressure range (see
sensors label).
•Only use pressure tight installation material.
•Avoid that persons get hit escaping air or bursting parts of the
instrument.
•The system must be pressureless during maintenance work.
W RNING!
Voltage used for supply!
ny contact with energized parts of the product, may
lead to a electrical shock which can lead to serious
injuries or even death!
•Consider all regulations for electrical installations.
•The system must be disconnected from any power supply during
maintenance work.
•Any electrical work on the system is only allowed by authorized
qualified personal.
4 S201

Safety instructions
TTENTION!
Permitted operating parameters!
Observe the permitted operating parameters, any
operation exceeding this parameters can lead to
malfunctions and may lead to damage on the
instrument or the system.
•Do not exceed the permitted operating parameters.
•Make sure the product is operated in its permitted limitations.
•Do not exceed or undercut the permitted storage and operation
temperature and pressure.
General safety instructions
•It is not allowed to use the product in explosive areas.
•Please observe the national regulations before/during installation
and operation.
Remarks
•It is not allowed to disassemble the product.
•Always use spanner to mount the product properly.
TTENTION!
Measurement values can be affected by malfunction!
The product must be installed properly and
frequently maintained, otherwise it may lead to
wrong measurement values, which can lead to
wrong results.
Storage and transportation
•Make sure that the transportation temperature of the sensor
without display is between -30 ... +70°C and with display between
-10 ... +60°C.
•For transportation it is recommended to use the packaging which
comes with the sensor.
•Please make sure that the storage temperature of the sensor is
between -10 ... +50°C.
•Avoid direct V and solar radiation during storage.
•For the storage the humidity must be <95% rH, no condensation.
S201 5

Registered trademarks
2 Registered trademarks
S TO®
Registered trademark of S TO iTEC
MODB S®
Registered trademark of the Modbus Organization, Hopkinton, SA
HART®
Registered trademark of the HART Communication Foundation, Austin,
SA
PROFIB S®
Registered trademark of the PROFIB S ser Organization, Karlsruhe,
Germany
3 pplication
The S201 is a dew point sensor which is designed to monitor the dew
point in industrial application within the permissible operating
parameters. These parameters can be found in the technical data
section.
The S201 can measure the following values:
•Temperature of the compressed air or gas.
•Humidity of the compressed air or gas.
•Dew point of the compressed air or gas.
The default factory settings are: temperature in °C, Humidity in % rH
and dew point in °C Td. Other units can be programmed by the optional
display or the service kit.
The S201 dew point sensor is not developed to be used in explosive
areas. For the use in explosive areas please contact the manufacturer.
The S201 dew point sensor is mainly used in compressed air systems in
industrial environment.
6 S201

Features
4 Features
•Dew point sensor for low dew point applications down to -60°C Td.
•Long term stability.
•Graphic display.
•Relay output.
•IP65 casing provides robust protection in rough industrial
environment.
•Very fast response time ensures safe and reliable indication
whenever dew points are out of valid ranges.
•Can be installed directly into dryers through G ½” thread.
•High accuracy of ± 2°C Td dew point.
S201 7

Technical data
5 Technical data
5.1 General
Parameters Standard unit dew point: °C Td
Standard unit temperature: °C
other units: °F, K
Standard unit humidity: % rH
Principle of measurement Capacitive method
Sensor Polymer
Measuring medium Non-corrosive gases
Measuring range Dew point: -60 ... +20°C Td
Relative humidity: 0 ... 90% rH
Temperature: -30 ... +70°C
Ambient temperature -20 ... +50°C
Ambient humidity: 0 ... 100% rH
Operating pressure -0.1 ... 5.0 MPa
Housing material PC + ABS
Material of the sensor
head and the screwing
Stainless steel sinter filter pore size < 30 µm
Protection class IP65
Dimensions See dimensional drawing on the page 10
Display 128 x 64 pixel, with backlight
Screwing thread G ½” thread (ISO 228/1);
NF 5/8” thread (on request)
Weight 226 g
5.2 Electrical data
Power supply 12 ... 30 VDC / 100 mA
5.3 Output-signals
Analog output Signal: 4 ... 20 mA, 3 wire
Analog output scaling 4 mA = -60°C Td
8 S201