Therm-App MD Manuale utente

Quick Start Guide
Congratulations on purchasing Therm-App®MD for Thermography —
The thermography camera that is changing the way thermal imaging
technology works.

Your Therm-App®MD comes with the following items:
• USB-OTG cables
• Quick Start Guide and Warranty
• Single Board Android Computer
Read This First
• The Therm-App MD device should be connected to the single board
computer (SBC) through the USB cable. The application that runs the
Therm-App MD is preinstalled on the SBC.
• Connect the SBC to a monitor through the HDMI plug and a mouse to a
USB plug and run the ThermApp aplication.
• Attach the Therm-App MD a tripod or any other fixed stationary point by a
¼ inch screw.
• Attach the SBC to a fixed location.

Take Note
• Therm-App MD can work as well with a standard Android device by downloading
the ThermApp application from Google Play and running the application. The
connection to the phone is via the USB OTG cable.
• Connect all the cabling before you connect the SBC to the power supply.
• In case the application is not installed on the SBC please contact Opgal and
the application would be sent to you.
Compatibility
• The Therm-App calibration tables are pre-installed on the SBC, if you
change the SBC, connect the SBC to an ethernet connection, plug-in the
Therm-App, and open the app to download the correct calibration tables.
• Therm-App MD was designed to work with the SBC and the preinstalled
software.
• Therm-App MD will only operate with mobile devices supporting USB OTG
(also called USB host).
• Therm-App MD requires a USB OTG cable supplied in the box.
• Therm-App MD will operate on Android version 6 (Marshmallow) and up.

Connecting the Android Single Board Computer (SBC)
• Connect the power cable to the SBC (1).
• Attach a mouse to the USB port (2).
• Attach a screen to the HDMI port (3).
• Turn on the SBC.
• With the cable supplied, connect the Therm-App USB OTG micro-USB
connection to the Therm-App device (5) and the USB to the SBC (4).
• The Therm-App application should automatically open (if not, open the
app manually) when the cable is connected from the Therm-App to the
computer.
1
2
3
4
5

Application Overview
Settings Menu
Quick Access Menu
Temperature Range & Lock Buttons
Temperature Scale
Gallery
Image Recording
Video Recording

Quick Tips
• To change the color palette, swipe the image right or left.
• To set the temperature threshold to “hot-red” and “cold-blue” palettes,
press the temperature reading in the middle of the temperature scale.
• To change the temperature from Celcius to Fahrenheit, click on the settings
button , click on Theromography Settings and then on Units.
• Pinch the touch screen for digital zoom.
• To use the automatic temperature range, press the current temperature
and then press the button on the dial.
• For best performance, the ambient room temperature should be 25° C.
• For optimal performance wait between 8-10 minutes from start-up, to allow
the Therm-App to reach an initial steady temperature state.
• In order to establish a baseline temperature, have 3 healthy people with
regular temperatures pass in front of the camera. Calculate from the
highest temperature of the 3 people and rasie the temperature dial 1.5° C
(see Configuring the Alert for instructions). This will be the threshold for
the alert for an individual with a fever.

• If you are using a 6.8mm lens, a person should stand 3-5m from the camera.
• If you are using a 19mm lens, a person should stand 6-10m from the
camera.
• The temperature indicator will give the surface temperature of the individual
with the hottest temperature in the field of view, or the configured region
of interest.
• When you receive an alert for an individual, check more thoroughly with a
thermometer.
• The Therm-App is affected by the ambient temperature of its surroundings
and it provides relative temperature readings, as opposed to a thermometer
which gives absolute temperature readings.
• While the baseline may change, the hightened temperature reading of a
feverish individual will give an alert.
• Note: The skin temperature of a person may be affected by; extreme
temperatures such as extreme cold or standing near a heat source,
drinking significant quantities of alcolhol, taking temperature altering drugs,
or strenuous exercise.

Configuring the Therm-App
• Click on Settings menu on the top left
of the Therm-App screen.
• This will open the settings menu.
• Click on General Settings.
• Ensure Time lapse is on Disable.
• The orientation of the video or
snapshot can be also be amended
from the General Settings menu,
by clicking on Video/Snapshot
orientation.
• Click on Alert mode settings (see
next page for instructions).

• Go to Advanced Parameters in
the Alert mode settings menu and
ensure the settings are as follows: -
• Check Play alert for the alert sound.
• - Detection duration is 0(seconds).
• - Alert delay is 0(seconds).
• - Recording time after detection
is 1(minute).
Configuring the Alert Mode
• These parameters
can be changed by
clicking on the name
of the setting and
clicking the -or +
symbol in the box
shown to the right.

Configuring the Thermography
• Click on the Quick Access Menu
• Click on Mode, then Thermography
Configuring the Emissivity
• Click on the Quick Access Menu
• Click on the Ruler and then on
Emissivity.
• Using the - or + symbols change the
Emissivity value to 0.98.
• This can also be done using the on
screen keyboard by typing the value
0.98 using your mouse pointer.