Thermo Scientific AerosolSense 2900 Manuale utente

AerosolSense 2900 Sampler
Manual
In-air pathogen surveillance solution
Part Number 121561-00
18March2021


Thermo Scientific AerosolSense Manual iii
Contents
Introduction........................................................................................................ 1-1
Features............................................................................................... 1-1
Product Description............................................................................ 1-2
Specifications ...................................................................................... 1-5
Basic Operation ................................................................................................ 2-1
Setting up the Sampler........................................................................ 2-1
Sample Cartridge Parts........................................................................ 2-1
Installing the Sample Cartridge ........................................................... 2-2
Removing the Sample Cartridge.......................................................... 2-5
Testing Services................................................................................... 2-7
Self-Testing On-site Lab .................................................................. 2-7
Third-Party Testing Lab................................................................... 2-7
Thermo Fisher Scientific Testing Services ........................................ 2-7
Cleaning Instructions.......................................................................... 2-8
Safety, Warranty, and WEEE.......................................................................... A-1
Safety ..................................................................................................A-1
Safety and Equipment Damage Alerts ..............................................A-1
Warranty.............................................................................................A-2
WEEE Compliance.............................................................................A-4
WEEE Symbol.................................................................................A-4
Quick Reference .............................................................................................. B-1
List of Figures .....................................................................................B-1
List of Tables.......................................................................................B-1
Chapter 1
Chapter 2
Appendix A
Appendix B


Thermo Scientific AerosolSense Manual 1-1
Chapter 1
Introduction
The Thermo Scientific™ AerosolSense™ Sampler, a new pathogen
surveillance solution, is designed to deliver timely and highly reliable
insight into in-air pathogen presence for monitoring and improving facility
safety protocols. Confidently help protect employees and communities
while maintaining business and operational continuity.
Figure 1–1. AerosolSense 2900 Sampler
The AerosolSense Sampler has the following features:
●Continuous sampling
●Small, easy to use design
●Highly sensitive and stable aerosol collection and separation technology
●Validated specifically for the SARS-CoV-2 pathogen, the sampler
design is intended to capture a wide variety of in-air pathogens
Features

Introduction
Product Description
1-2 AerosolSense Manual Thermo Scientific
The sampler collects air samples through an omnidirectional inlet. A
sample cartridge installed into the sampler contains the collection substrate.
The air sample is directed toward the collection substrate through an
accelerating slit impactor. Particles are trapped on the collection substrate
as the air is drawn through the sampler. After the sampling cycle, the
sample cartridge is removed, placed into a biohazard bag, and sent to a
laboratory for PCR testing. There are three laboratory testing options: (1)
self-testing at on-site lab, (2) third party lab testing, or (3) Thermo Fisher
Scientific testing services.
Figure 1–2. Sampler LED Guide
Required Materials Not Included with the Device:
●Personal Protective Equipment (PPE)
●Applicable EPA-approved disinfectant
Warning Use CDC standard precaution guidelines, including hand
hygiene and personal protective equipment (PPE) when handling the
sampler and sample cartridge. ▲
Equipment Damage Do not attempt to lift the sampler with the inlet
cap. ▲
Product
Description
Icon Here

Introduction
Product Description
Thermo Scientific AerosolSense Manual 1-3
Figure 1–3. AerosolSense Sampler Front View
Figure 1–4. AerosolSense Sampler Inside View

Introduction
Product Description
1-4 AerosolSense Manual Thermo Scientific
Figure 1–5. AerosolSense Sampler Back View

Introduction
Specifications
Thermo Scientific AerosolSense Manual 1-5
Table 1–1 lists the specifications for the AerosolSense Sampler.
Table 1–1. AerosolSense Sampler Specifications
Dimensions 14.5” (H) x 14” (W) x 13” (L)
Weight 26 lbs
Sample collection holder and media Single-use sampling cartridge containing (2)
1-inch collection substrates
Safety and electrical designations TUV-SUD, CE, UKCA, ROHS
Sample flow control 200 L per minute
Operating temperature 5–45 °C
Status indicator 3 color LED
Power consumption 100-240 VAC 50-60 Hz 125 Watts
Dust interference (PM1-PM2.5-PM10) Minimum impact (80-1050 µg/m3)
Humidity interference Minimum impact (over 10-90% RH)
Specifications

Indice
Manuali Campionatore popolari di altre marche

Akai
Akai MPCX Manuale utente

KAM
KAM Simple Precision IAS Manuale utente

YZ Systems
YZ Systems DynaPak DP-2010FU Manuale elenco delle parti

YZ Systems
YZ Systems DynaPak 2010N Series Manuale utente

YZ Systems
YZ Systems DynaPak DP-2010UX Manuale elenco delle parti

Novation
Novation Circuit Rhythm Manuale utente