③ Sash monitoring warning display
④ Status display (green, yellow, red) with text HIGH and LOW
⑤ Connection socket for service and commissioning
⑥ Acoustic alarm acknowledgement
⑦ Manual control
⑧ Selection of operating mode
⑨ Lighting or other equipment (RMF)
⑩ Automatic sash device (FH), sun protection, e.g. blinds (RMF)
Base plate
Connection socket for EASYLAB controller
Supply voltage 24 V DC from the controller TCU3 or adapter module TAM
Connecting cable Approx. 5 m standard network patch cable, SF-UTP, extendable to 40 m
Operating temperature 10 – 50 °C
IEC protection class III (protective extra-low voltage)
Protection level IP 20
EC conformity EMC according to 2004/108/EC
Dimensions 145 × 111 × 23 mm
Weight 0.280 kg
Control panels with adjustable status display for the EASYLAB system, used to display and control the most critical aerodynamic and safety-related functions of a
fume cupboard or a room. The control panel includes a backlit 40-character display, function buttons, a three-colour status display, an alarm sounder, and a service
socket.
Mounting on side frame or wall
Functions
Configurable, optical and acoustic signalling of the safety-related functions
Text displays in several languages
Large three-colour status display; green, yellow or red with text LOW or HIGH (red: either blinking or permanent)
Continuous self monitoring of the communication between controller and control panel
Integral service socket for configuration and diagnosis
Wireless communication with Bluetooth module BlueCON
Two control panels can be connected simultaneously to a controller
Configurable functions
Available (configured) function buttons are visible
Unavailable function buttons are not visible
Control panel for fume cupboard controller
Function buttons that can be activated
Warning display that the 'maximum operational sash opening' has been exceeded
Increased volume flow
Reduced volume flow
Shut-off mode
Automatic sash device
Fume cupboard lighting
Manual mode (override operating mode default setting)
Text display