Vesticam 2 Manuale utente

h
f
g
i
j
edc
b
a
j
i
h
dc
b
g
a
silicone head band
quick release side buons
mounng plate - adjusts for interpupillary variaon
locking ring
eye patch cylinder
exible goggles frame
silicone eye cups - removeable and reusable
camera cylinder - with infra-red USB camera and
xaon light
joysck - for focus and ne adjustment
xaon light buon
a
a
b
b
c
c
d
d
e
f
f
g
h
h
i
i
j
j
b
g
Squeeze buon toward
flange to release
b
1. Start with the camera cylinder and eye patch cylinder o…not
aached to the goggles
2. Follow your normal paent hygiene pathways when applying
the goggles. Use eecve alcohol/an-microbial wipes to
. Do not use on paents with
infecve disorders. (see also 3.1.2 “Cleaning and avoiding
contaminaon” and Appendix 1 full instrucon manual)
If you cannot see
your paent’s eye
completely on the
viewing screen,
simply loosen the
locking ring, realign
the camera, and
reghten the
locking ring.
The goggles are
comfortable, high quality
and exible enough to t
most face shapes and
sizes.
by squeezing the
and the goggles ange
together
3. Help your paent to .
Most paents can do this themselves, like swimming goggles.
4. -
, loosen the
locking rings at the front
of the mounng plate by
turning an-clockwise.
Allow the mounng
plates to slide into a
comfortable posion on
your paent’s face.
Your paent should be
able to see out of both
viewing holes.
Locking ring
anti-clockwise to release and slide
clockwise to lock into correct position
slide
slide
5. Pull rubber tags at each side of goggles gently
by
rmly ghtening the locking rings in a clockwise direcon.
6. Hold the camera cylinder so that the lead is exing the
cylinder at “5 o’clock”.I
, and rotate it clockwise unl the camera
cylinder “clicks” and locks into place (lead should be at
“6 o’clock” when locked).
Camera cylinder and eye patch cylinder
anti-clockwise to unlock (”click”)
clockwise to lock (”click”)
7. use one hand to
stabilize the camera cylinder and the other hand to
. Firmly
. The camera cannot be pushed into the eye as
there is an end stop. It is important to focus the eye
to ensure that possible torsion is visible.
8. To deny vision, slip the eye patch cylinder into the
remaining mounng plate. Twist the cylinder clockwise
and “click” to lock into place.
9. Press the small buon at the end of the joysck to
acvate the xaon light. When the buon is , the
xaon light is. When the buon is , the xaon
light is .
: the xaon light will only work when the camera is
plugged into the external recording device (phone, tablet
or computer).
Vescam2 goggles are designed to be light ght.
Adjust the goggles with your paent so they are
comfortable, secure, light ght, and rm.
If there is any light leaking into the goggles at this point,
check that the camera cylinder and eye patch cylinder are
correctly in place. If there is sll light leakage, gently move
the goggles on your paent’s face to eliminate any gaps
between face and goggles.
Joy-sck
focus, in and out
fine adjustment
up/down, side/side
e
1
2
3
4
5
6
7
10
9
8

Follow your normal paent infecon control
pathways when applying Vescam goggles.
Use of PPE and general environmental cleaning
is the responsibility of the user.
-
Remove the camera cylinder prior to cleaning.
Do not autoclave. Do not use hard or pointed
objects to clean any part of the Vescam unit.
Do not allow liquid to enter any part of the
goggles camera unit. The
(visible when looking into the camera cylinder)
should be checked regularly for stains and dust,
and should be cleaned as required using a
standard spectacle lens cloth.
The silicone eye cups can be removed, washed
and air dried separately and then placed back
onto the mounng plate if required.
Extra eye cups are available for purchase.
The goggles mounng plate, head band and
eye cups can be immersed in gentle cleaning
detergent, rinsed and allowed to air dry
completely if required.
If you have quesons, concerns or feedback,
please contact Vescam
p: 1300141488
e: support@vescam.com
w: www.vescam.com
Never leave your Vescam
unit in a hot car.
Avoid dropping your Vescam unit or recording
device.
.
As Vescam IR goggle units are not “therapeuc or
diagnosc tools” and do not provide a measurement
or trace, they do not require regular calibraon.
For Vescam2 servicing or repair contact
support@vescam.com
Take care when connecng and disconnecng the
camera or any adaptors to your recording device.
The USB ports can be damaged if not used correctly
Please follow your normal service and maintenance
pathways for maintenance of the charging
aachments and usb 240v wall socket charger
supplied with your recording device.
Lock your Vescam unit with your other medical
equipment/medical records aer use. Remember
that your recording device may contain videos of
your paents, and these should be treated as
medical records.
: it is the user’s responsibility to follow
their usual medical record pathways and
recommendaons for storing paent eye movement
videos. To secure paent data it is important to
regularly transfer paent videos into the usual or
required medical record storage system.