Zolix Instruments GLORIA-X500A Manuale utente

GLORIA-X500A
500W Xenon Light Source System
Operation Manual
(Version 1.0 2015-4)
Zolix Instruments Co., Ltd.

Thanks for using Zolix products !
Please read this introduction carefully before using. And kindly please contact Zolix office in Beijing or
local office if you have any questions and suggestions.
Directory
1Descriptions ............................................................................................................................................... 1
2Main technical specifications of light source ............................................................................................. 2
3Xenon lamp light source housing ............................................................................................................... 3
3.1Light path ................................................................................................................................................... 3
3.2Xenon lamp bulb installation method ........................................................................................................ 4
3.3Light source housing operation guide and safety precautions .................................................................... 6
3.3.1Adjustment before the xenon lamp is lighted ............................................................................................. 6
3.3.2Adjustment after xenon lamp was ignited .................................................................................................. 7
3.3.3Safety precautions ...................................................................................................................................... 8
4Xenon lamp Power Supply ......................................................................................................................... 9
4.1Specifications ............................................................................................................................................. 9
4.2Function Introductions ............................................................................................................................... 9
4.3Operation Guide ....................................................................................................................................... 12
5Unpacking Procedures ............................................................................................................................. 13
6Safety instructions for products ................................................................................................................ 14
7Product quality assurance ......................................................................................................................... 15
8Product warranty clause ........................................................................................................................... 16
9Packing List .............................................................................................................................................. 17

Descriptions
1
Zolix Instruments Co., Ltd. http://www.zolix.com.cn E-mail:[email protected].cn
1 Descriptions
There is 500W high- pressure short-arc spherical xenon lamp in the 500W Xenon light source,
which will form the arc discharge under the high-frequency and high- pressure. The
high-voltage short-arc spherical xenon lamp can output the point light source and strong and
stable continuous spectrum from UV to NIR. The visible light is approximate to the sun light.
This Xenon light source is the ideal light source for the IR detection, UV spectrophotometer,
fluorescence spectrophotometer and so on. And also used for solar simulation.
500W Xenon lamp power supply
500W Xenon lamp light source housing

Specification
2
Zolix Instruments Co., Ltd. http://www.zolix.com.cn E-mail:[email protected].cn
2 Main technical specifications of light source
MAX. power 500W
Parallel light spot dia. 44mm(theoretical value is 50mm from exit)
Divergence half angle <1.54°
Spectrum range 200-2500nm(250-1800nm)
Lamp adjustment 2D
Reflector adjustment 3D
Xenon lamp life time 2000 hours (import);
500 hours (made in china)
Optical axis height 158-193mm
Type of cooling FAN
Volume 218*257*354mm
Weight 5.8Kg

Xenon lamp Light Source Housing
3
Zolix Instruments Co., Ltd. http://www.zolix.com.cn E-mail:[email protected].cn
3 Xenon lamp light source housing
3.1 Light path
In the xenon lamp housing, the output light energy is enhanced by spheric reflect mirror behind
the xenon lamp.Collimating lens collects the light energy from the lamp and reflector.
A
node
Lens
Parallel light
Cathode
Spheric reflect mirror

Xenon lamp Light Source Housing
4
Zolix Instruments Co., Ltd. http://www.zolix.com.cn E-mail:[email protected].cn
3.2 Xenon lamp bulb installation method
Caution: You must make sure the power supply is turned off when you mounting
the xenon lamp!
Side plate
Remove the
side
p
late
Light cup
fixed frame
After removing the light
cup fixed frame

Xenon lamp Light Source Housing
5
Zolix Instruments Co., Ltd. http://www.zolix.com.cn E-mail:[email protected].cn
Anode heat sink
Cathode block
Anode fixed nut
Anode wire
Cathode
fastening screw
1. Getting the light source housing for the first time, you should remove the light cup fixed frame as
shown above and keep the light cup fixed frame properly for the next transport.
2. Adjust the green hand wheel as below, make the vertical distance between height adjusting base
and translational base is about 4mm; Adjust the orange hand wheel, make the bulb supporting seat
centered around relative to the light source housing;
3. Before installation, you’d better swab the lamp with alcohol for fear that the transmittance and
life-time will not be detracted by the sweat on hand, oil pollution or dusty covered on the lamp
envelope;
4. Make the 150W xenon lamp ready (The thin end is the cathode);
5. Install the heat sink to the anode of xenon lamp bulb and make the screw tight;
6 . Put the cathode of lamp down and inside the setting location, then rotate bulb and make subline
on the bulb shell towards the direction of any one of the handwheel, maintain the constant
direction and fasten the positive power supply with positive fixed nut. Adjust the height of lamp,
Height adjusting knob Horizontal adjusting knob
Height
adjusting
base
Translational
block
Bulb supporting
seat
Distance of
about 4 mm

Xenon lamp Light Source Housing
6
Zolix Instruments Co., Ltd. http://www.zolix.com.cn E-mail:[email protected].cn
make the cathode of lamp locate the center of reflector and make the screw tight.
3.3 Light source housing operation guide and safety precautions
3.3.1 Adjustment before the xenon lamp is lighted
1) Adjust 3 reflector turret at the same time, make the
diatance between light cup underside and adjusting
bracket base is about 3mm
2) Reove collimating lens barrel, then adjust shift knob
and high knob of lamp, to make the electrode and the
electrode imaging as shown in the picture on the
right:
the diatance between cathode and cathode imaging is
about 2mm.
3) After adjustment, restore the collimating lens barrel
and side plate in place.
Electrode
Electrode
imaging
2mm

Xenon lamp Light Source Housing
7
Zolix Instruments Co., Ltd. http://www.zolix.com.cn E-mail:[email protected].cn
3.3.2 Adjustment after xenon lamp was ignited
Caution:
Please wear the protective glasses before the following operations!
1) Start the power supply and light up the bulb, adjust the collimating lens barrel knob just above the
light tube in order to the image of the lamp filament appears on the wall(The diatance between
screen and light outlet is about 200mm)
2) Adjust the horizontal adjusting knob, which can make the lamp bulb self-emission light spot moves
left and right relative to the center of light tube(solid line spot as shown above, figure 1),make the
light spot symmetrical relative to the light tube center.
3) Adjust the vertical adjusting knob, which can make the lamp bulb self-emission light spot moves
up and down relative to the center of light tube(solid line spot as shown above, figure 2), make the
light bulb glow center and the light tube center coincided.
4) Finish the bulb position adjustment, it’s time to adjust the position of the reflector lamp cup. knob
3、4、5 of lamp cup at the back of lamp source as shown above can adjust the size and location of
reflected light spot.
5) By adjusting the knob 3 as shown above, make the reflected light spot moving left and right; By
adjusting the knob 4 as shown above, make the reflected light spot moving up and down; By
1 light bulb horizontal
adjustment
5
4Reflector lamp
cup pitch
3 Reflector lamp
cup oscillation
2 light bulb vertical
ad
j
ustment
Light spot diagram
Reflector lamp cup collect
emer
g
ent li
g
ht s
p
ot
Lamp bulb self-
emission li
g
ht s
p
ot
Corresponding to the light
tube center
p
osition

Xenon lamp Light Source Housing
8
Zolix Instruments Co., Ltd. http://www.zolix.com.cn E-mail:[email protected].cn
adjusting red knob 5, make the reflected light spot moving at 45 degree angle; By adjusting knob
3、4、5, make reflected light spot size and location coincide with light spot.
6) Parallel light adjustment. Firstly, you need to put a blank paper in about 250mm from the light
outlet, then by moving the blank paper(to the direction of the light outlet), you can observe the size
of light spot on the blank paper. if the size of light spot change obviously,you need to slide lens
cone until the size of light spot don’t change again.
7) If it is a multi outlet type light source in the room, according to the above steps before shooting the
main light path adjustment, to meet the use requirements by adjusting the optical path
corresponding to the side lamp side light path can be.
Note: the beam quality of the auxiliary light path may not be consistent with the main light path. When the actual
effect has been affected, the main optical path should be adjusted repeatedly so as to meet the use requirements.
3.3.3 Safety precautions
1) The output spectra of xenon lamps contain high levels of short wavelength ultraviolet radiation,
which can cause permanent damage to unprotected skin and eyes. Therefore, never look directly
into the output beam, any diffuse reflection, or any specular (mirror) reflection even for a short
period of time without suitable protective ultraviolet blocking filters or glasses.
2) Xenon lamps are always under pressures above atmosphere, so use extreme care when handling,
installing or mounting them to avoid any mechanical strains on the lamp. Protective glasses, heavy
gloves and a mask should always be worn when handling xenon lamps. Mechanical strain can
cause the lamp to shatter violently!
3) All xenon lamps, other than ozone free lamps, allow ozone gas to be generated while the lamp is
in operation. Always provide for proper ventilation to remove this ozone. High concentrations of
ozone are dangerous and can constitute a serious health hazard!
4) While the lamp is in operation, the instantaneous voltage will be more than 33,000 V and the
current will be larger than 25 A. Please let the lamp anode and the pole connection on trigger will
be contacted tightly.
5) In order to get the better light source stability, the lamp needs 30min preparing time.
6) The lamp in operation is very hot. Therefore, never touch the lamp envelops when it is in operation
or be turned off within 20 minutes.
7) In addition, please not cover the lamp housing with anything, because the Xenon lamp will
produce much heat when working.
Indice