ABB WaterMaster FET100 Manuale utente

User Guide OI/FET100–EN Rev. A
WaterMaster FET100
Electromagnetic flowmeter
Transmitter
The perfect fit for all water
industry applications
Introduction
WaterMaster™ is a range of high performance
electromagnetic flowmeters for the measurement of
electrically-conductive fluids and systems are normally
supplied factory-configured and calibrated.
This User Guide provides installation, connection,
security, start-up and basic setup details. For
programming and configuration information refer to the
Programming Guide – IM/WMP.
This User Guide should be used in conjunction with the
following publications:
– Programming Guide – IM/WMP
– User Guide Supplement, PROFIBUS RS485 Physical
Layer (FEX100–DP) – IM/WMPBS–EN
– User Guide Supplement, PROFIBUS FEX100-DP
Parameter Tables – IM/WMPBST–EN
– User Guide Supplement, MODBUS RS485 Physical
Layer (FEX100–MB) – COI/FEX100/MOD–EN
– User Guide Supplement, MODBUS Tables –
COI/FEX100/MOD/TBL–EN

The Company
We are an established world force in the design and manufacture of instrumentation for industrial process control,
flow measurement, gas and liquid analysis and environmental applications.
As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service
and support worldwide.
We are committed to teamwork, high quality manufacturing, advanced technology and unrivalled service and
support.
The quality, accuracy and performance of the Company’s products result from over 100 years experience, combined
with a continuous program of innovative design and development to incorporate the latest technology.
Quality Control
The UKAS Calibration Laboratory No. 0255 is just one of the ten flow calibration plants operated by the Company
and is indicative of our dedication to quality and accuracy.
UKAS Calibration
Laboratory No. 0255
0255

WaterMaster FET100
Electromagnetic floweter | Transmitter
OI/FET100–EN Rev. A 1
1 Safety ............................................................................................................................................... 2
1.1 Electrical Safety ............................................................................................................................................. 2
1.2 Symbols ........................................................................................................................................................ 2
1.3 Health & Safety ............................................................................................................................................. 3
2 Mechanical Installation .................................................................................................................... 4
2.1 Installation Conditions ................................................................................................................................... 4
2.2 Dimensions ................................................................................................................................................... 6
3 Electrical Installation ....................................................................................................................... 8
3.1 Remote Transmitter / Sensor Arrangement ................................................................................................... 8
3.2 Transmitter Terminal Connections ............................................................................................................... 10
3.3 Cable Preparation (Remote Systems Only) .................................................................................................. 11
3.4 Transmitter / Sensor Cable Connections ..................................................................................................... 12
3.4.1 Sensor Cable Terminal Connections and Recommended Cable Lengths ...................................... 12
3.5 Output Connections .................................................................................................................................... 13
3.5.1 Frequency Outputs ....................................................................................................................... 13
3.5.2 Alarm Outputs .............................................................................................................................. 14
3.5.3 Current Output (4 to 20 mA) – HART (FEX100) Variant .................................................................. 14
3.5.4 RS485 Communications – PROFIBUS (FEX100-DP)
and MODBUS (FEX100-MB) Variants ........................................................................................... 15
3.5.5 Test Point Access ......................................................................................................................... 15
3.6 Power Supply Connections ......................................................................................................................... 16
3.6.1 AC Power Supply ......................................................................................................................... 16
3.6.2 DC (and Low Voltage AC) Power Supply ......................................................................................17
3.6.3 Configuration DIP Switches .......................................................................................................... 18
3.7 Refitting the Cartridge and Cover ................................................................................................................ 18
4 Start-up and Operation ................................................................................................................. 20
4.1 Navigating the Menus and Parameters ........................................................................................................ 20
4.2 Start-up Screens ......................................................................................................................................... 21
4.3 Security Levels and Password Access ........................................................................................................ 23
4.3.1 Default Passwords ....................................................................................................................... 24
4.3.2 Entering Passwords ...................................................................................................................... 24
4.4 Easy Setup ................................................................................................................................................. 25
5 Specification .................................................................................................................................. 27

WaterMaster FET100
Electromagnetic floweter | Transmitter 1 Safety
2OI/FET100–EN Rev. A
1 Safety
Information in this manual is intended only to assist our customers in the efficient operation of our
equipment. Use of this manual for any other purpose is specifically prohibited and its contents are not to be
reproduced in full or part without prior approval of the Technical Publications Department.
1.1 Electrical Safety
This equipment follows, obeys the requirements of CEI/IEC 61010-1:2001-2 'Safety Requirements for
Electrical Equipment for Measurement, Control and Laboratory Use' and follows, obeys NIST and OSHA.
If the equipment is used in a manner NOT specified by the Company, the protection provided by the
equipment may be impaired.
1.2 Symbols
One or more of the following symbols may appear on the equipment labeling:
Warning – Refer to the manual for instructions Direct current supply only
Caution – Risk of electric shock Alternating current supply only
Protective earth (ground) terminal Both direct and alternating current supply
Earth (ground) terminal The equipment is protected
through double insulation

WaterMaster FET100
Electromagnetic floweter | Transmitter 1 Safety
OI/FET100–EN Rev. A 3
1.3 Health & Safety
Health and Safety
To ensure that our products are safe and without risk to health, the following points must be noted:
The safety requirements of this equipment, any associated equipment and the local
environment must be taken into consideration during installation.
Install and use this equipment and any associated equipment in accordance with the relevant
national and local standards.
The relevant sections of these instructions must be read carefully before proceeding.
Warning labels on containers and packages must be observed.
Installation, operation, maintenance and servicing must only be carried out by suitably trained
personnel and in accordance with the information given.
Normal safety precautions must be taken to avoid the possibility of an accident occurring when
operating in conditions of high pressure and / or temperature.
Chemicals must be stored away from heat, protected from temperature extremes and powders
kept dry. Normal safe handling procedures must be used.
When disposing of chemicals ensure that no two chemicals are mixed.
Product liability – advice and assistance provided without charge is given in good faith but
without liability.
Safety advice concerning the use of the equipment described in this manual or any relevant hazard
data sheets (where applicable) may be obtained from the Company address on the back cover,
together with servicing and spares information.
Warning.
System configuration must be carried out only by users or personnel with approved access rights
(user privileges).
Read all relevant sections of this guide before configuring the system or modifying system
parameters.
Install and use associated equipment in accordance with the relevant national and local
standards.

WaterMaster FET100
Electromagnetic floweter | Transmitter 2 Mechanical Installation
4OI/FET100–EN Rev. A
2 Mechanical Installation
2.1 Installation Conditions
Fig. 2.1 Siting
Fig. 2.2 Within Temperature Limits
Fig. 2.3 Shade
Allow room
to read display
60 °C
(140 °F) Max.
–20 °C
(–40 °F)
Min.

WaterMaster FET100
Electromagnetic floweter | Transmitter 2 Mechanical Installation
OI/FET100–EN Rev. A 5
Fig. 2.4 Vibration
Fig. 2.5 Spillage
Fig. 2.6 Within Environmental Rating
Fig. 2.7 Access to Transmitter
IP67 (NEMA 4X)
Transmitter Submersion:
1 m (3.3 ft) <12 hours
Accrued time
Position to enable
access to display and
communication connector

WaterMaster FET100
Electromagnetic floweter | Transmitter 2 Mechanical Installation
6OI/FET100–EN Rev. A
2.2 Dimensions
Dimensions in mm (in).
Fig. 2.8 Integral Transmitter Dimensions (Standard Gland Shown)
118 (4.6)
98 (3.9) 202.5 (8) – Standard Gland
250 (9.8) Armored Gland
201 (7.9)

WaterMaster FET100
Electromagnetic floweter | Transmitter 2 Mechanical Installation
OI/FET100–EN Rev. A 7
Fig. 2.9 Remote Transmitter Dimensions (Standard Gland Shown)
Note. Fix remote transmitter to a secure surface using 3 x M5 screws (not supplied).
202.5 (8) – Standard Gland
250 (9.8) Armored Gland
98 (3.9) R3.2 (0.1 )
71 (2.8)
179 (7.0)

WaterMaster FET100
Electromagnetic floweter | Transmitter 3 Electrical Installation
8OI/FET100–EN Rev. A
3 Electrical Installation
3.1 Remote Transmitter / Sensor Arrangement
Note. For bonding connections use 4mm2(< 10AWG) cable.
Fig. 3.1 Remote Transmitter in Roadside Cabinet – Flanged Sensor
Power Supply
Indice
Altri manuali ABB Trasmettitore

ABB
ABB TTF300 Manuale di formazione

ABB
ABB LWT300 series Manuale utente

ABB
ABB 3299E-A21908 Manuale utente

ABB
ABB 2600T Series Guida

ABB
ABB LMT Series Manuale utente

ABB
ABB TB82EC Istruzioni operative

ABB
ABB 266 MODBUS Manuale utente

ABB
ABB AWT210 Manuale utente

ABB
ABB 266 DDS Manuale utente

ABB
ABB 265DS Manuale utente